Evaluation of a fully automated high-performance liquid chromatography assay for hemoglobin A1c

被引:0
|
作者
Khuu, HM [1 ]
Robinson, CA [1 ]
Goolsby, K [1 ]
Hardy, RW [1 ]
Konrad, RJ [1 ]
机构
[1] Univ Alabama Birmingham, Dept Pathol, Div Lab Med, Birmingham, AL 35233 USA
关键词
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Introduction.-Measurement of hemoglobin A(1c) (HbA(1c)) is used as an objective measure of long-term blood glucose control in diabetic patients. Recent improvements in automation combined with new recommendations for precision and accuracy have caused us to reevaluate our methods for measuring HbA(1c). Objective,We evaluated a newly automated high-performance liquid chromatography (HPLC) instrument for measurement of HbA(1c) (Tosoh A(1c) 2.2 Plus Glycohemoglobin Analyzer, Tosoh Medics, Foster City, Calif) and compared the results obtained by HPLC to those obtained with an immunoassay (Hitachi 911, Boehringer Mannheim Corporation, Indianapolis, Ind). Results.-The Tosoh analyzer was found to be linear in a range of 5.3% to 17% and had a throughput of 20 samples per hour. HbA(1c) results for 102 patient samples by the 2 techniques showed good correlation, with a slope of 0.87 and an intercept at 1.27% +/- 0.15%. Both the total and within-run coefficients of variation were consistently lower for the HPLC method compared with the immunoassay method. The HPLC method produces a chromatogram that shows the different hemoglobin fractions, allowing identification of abnormal hemoglobin variants. in heterozygous individuals, HbA(1c) measurements are made with no interference from the hemoglobin variant. In the case of homozygous or doubly heterozygous hemoglobin variants, the Tosoh HPLC identifies the hemoglobin variants as such and correctly does not report a HbA(1c) value in the presence of a markedly decreased amount of hemoglobin A. Conclusions.-The Tosoh HPLC provides adequate throughput and improved precision, and the method is traceable to the Diabetes Control and Complications Trial.
引用
收藏
页码:763 / 767
页数:5
相关论文
共 50 条
  • [1] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY METHOD FOR HEMOGLOBIN A1C
    DAVIS, JE
    MCDONALD, JM
    JARETT, L
    DIABETES, 1978, 27 (02) : 102 - 107
  • [2] Evaluation of hemoglobin A1c measurement from filter paper using high-performance liquid chromatography and immunoturbidimetric assay
    Wu, Yonghua
    Yang, Xu
    Wang, Haining
    Li, Zhenrong
    Wang, Tiancheng
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2017, 77 (02): : 104 - 108
  • [3] Effect of hemoglobin F and A2 on hemoglobin A1c determined by cation exchange high-performance liquid chromatography
    Can, Murat
    Guven, Berrak
    Akca, Ayse Semra Demir
    JOURNAL OF LABORATORY MEDICINE, 2019, 43 (05) : 265 - 268
  • [4] QUANTITATION OF HEMOGLOBIN-A1A+B AND HEMOGLOBIN-A1C BY AUTOMATED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    GRUBER, CA
    KOETS, MD
    CLINICAL CHEMISTRY, 1979, 25 (11) : 1970 - 1971
  • [5] A new automated enzymatic assay for hemoglobin A1c
    Taniguchi, Y
    Nishio, T
    Saito, K
    Yago, H
    Ushizawa, K
    CLINICAL CHEMISTRY, 2005, 51 : A246 - A246
  • [6] Evaluation of an Enzymatic Hemoglobin A1C Assay
    Majeed, Numan
    Kanani, Fatima
    Zubairi, Adnan Mustafa
    ANNALS OF KING EDWARD MEDICAL UNIVERSITY LAHORE PAKISTAN, 2024, 30 (02): : 184 - 190
  • [7] Evaluation of an immunoturbidimetric assay for hemoglobin A1c
    Akan, P
    EUROMEDLAB 2003: PROCEEDINGS OF THE 15TH IFCC-FESCC EUROPEAN CONGRESS OF CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2003, : 187 - 193
  • [8] COMPARISON OF THREE HEMOGLOBIN A1C POINT OF CARE ANALYSERS WITH A HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) REFERENCE METHODS
    Fatas, M.
    Gasalla, J. M.
    Garcia, B.
    Janez, R.
    Hernando, C.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S678 - S678
  • [9] Comparison of hemoglobin A1c and total glycated hemoglobin using high performance liquid chromatography (BioRad variant).
    Duffy, CA
    Duncan, MR
    Schellhorn, S
    Downing, RM
    Plapp, FV
    Miles, J
    CLINICAL CHEMISTRY, 1997, 43 : 218 - 218
  • [10] Development of a fully automated hemoglobin A1c assay on the Boehringer Mannheim Elecsys® system.
    Shindelman, J
    Pham, D
    Mahal, J
    Rinne, S
    Chang, CC
    CLINICAL CHEMISTRY, 1998, 44 : A9 - A9