Initial Stability Measurements of Implants Using a New Magnetic Resonance Frequency Analyzer With Titanium Transducers: An Ex Vivo Study

被引:16
|
作者
Bural, Canan [1 ]
Dayan, Cagatay [2 ]
Geckili, Onur [1 ]
机构
[1] Istanbul Cerrahpasa Univ, Fac Dent, Dept Prosthodont, Istanbul, Turkey
[2] Istanbul Univ, Program Dent Technicians, Istanbul, Turkey
关键词
osseointegration; implantology; experimental design; implant stability; resonance frequency; dental implant; IN-VITRO; DENTAL IMPLANTS; DEVICES; DESIGNS;
D O I
10.1563/aaid-joi-D-19-00126
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The establishment of dental implant stability is mandatory for successful osseointegration. Resonance frequency analysis (RFA) is the most frequently used method for the clinical measurement of implant stability. The purpose of the present study was to evaluate the reliability of the recently developed RF analyzer Penguin RFA and to compare it with the traditional RF analyzer Osstell ISQ. Sixty implants were inserted into fresh steer vertebrae and pelvis. Implant stability was measured using Penguin RFA by its transducers (multipegs) and Osstell ISQ by its transducers (smartpegs). Additionally, stability was measured by multipegs with Osstell ISQ and by smartpegs with Penguin RFA. The intraobserver and interobserver reliability of Penguin RFA were estimated by the intraclass coefficient (ICC). Mean implant stability quotients (ISQs) measured with Osstell ISQ were higher than the ISQs measured with Penguin RFA (P < .05). The intra- and interobserver reliability of Penguin RFA were considered as excellent (ICC > 0.7). For Osstell ISQ, no significance in ISQs was detected between the readings by smartpegs and multipegs (P > .05), while for Penguin RFA ISQs by smartpegs were significantly higher than the ISQs by multipegs (P <.05). The recently developed Penguin RFA is reliable and can be used in clinical practice for the measurement of dental implant stability regardless of the bone type. The multipegs originally manufactured for the Penguin RFA is also compatible with Osstell ISQ.
引用
收藏
页码:35 / 40
页数:6
相关论文
共 37 条
  • [21] A new approach to study in vivo cellular metabolism using a modellistic analysis of magnetic resonance spectra
    Gastaldelli, Amalia
    Martini, Silvia
    Donati, Alessandro
    Marchettini, Nadia
    Rossi, Claudio
    Bastianoni, Simone
    MATHEMATICAL BIOSCIENCES, 2009, 222 (01) : 36 - 41
  • [22] A comparison of implant stability quotients measured using magnetic resonance frequency analysis from two directions: a prospective clinical study during the initial healing period
    Park, Jong-Chul
    Kim, Hyun-Duck
    Kim, Soung-Min
    Kim, Myung-Jin
    Lee, Jong-Ho
    CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (06) : 591 - 597
  • [23] Magnetic resonance characterization of the esophagus and small intestine using a flexible MRI catheter: An ex-vivo study in pigs.
    Hasharoni, K
    Sicron, N
    Beiln, S
    Golan, E
    Jacob, H
    GASTROENTEROLOGY, 2005, 128 (04) : A404 - A404
  • [24] New descriptors of trabecular bone microarchitecture: An in vivo longitudinal study using magnetic resonance imaging.
    Pothuaud, L
    Newitt, DC
    Chesnut, C
    Genant, HK
    MacDonald, B
    Majumdar, S
    JOURNAL OF BONE AND MINERAL RESEARCH, 2001, 16 : S461 - S461
  • [25] The active magnetic resonance imaging stent (AMRIS): Initial experimental in vivo results using angiography and flow measurements with local MR signal amplification
    Kivelitz, DE
    Wagner, S
    Hansel, J
    Schnorr, J
    Wetzler, R
    Hamm, BK
    RADIOLOGY, 2000, 217 : 422 - 423
  • [26] Is gingival biotype a critical determinant of secondary stability of orthodontic mini-implants - A prospective clinical study using resonance frequency analysis
    Pradhan, Sayali Prashant
    Nambiar, Supriya
    Shetty, Shravan
    Shetty, Siddarth
    Jose, Nidhin Philip
    APOS TRENDS IN ORTHODONTICS, 2020, 10 (04) : 245 - 252
  • [27] Detection and evaluation of cartilage defects in the canine stifle joint - an ex vivo study using high-field magnetic resonance imaging
    Flatz, K. M.
    Glaser, C.
    Flatz, W. H.
    Reiser, M. F.
    Matis, U.
    TIERAERZTLICHE PRAXIS AUSGABE KLEINTIERE HEIMTIERE, 2014, 42 (05): : 291 - 296
  • [28] Resonance frequency analysis (RFA) and insertional torque (IT) stability comparisons of implants placed using osteotomes versus drilling techniques: A preliminary case study
    Xing, Yi
    Khandelwal, Namita
    Petrov, Sofia
    Drew, Howard J.
    Mupparapu, Mel
    QUINTESSENCE INTERNATIONAL, 2015, 46 (09): : 789 - 798
  • [29] Effects of continual intermittent administration of parathyroid hormone on implant stability in the presence of osteoporosis: an in vivo study using resonance frequency analysis in a rabbit model
    Oki, Yoshifumi
    Doi, Kazuya
    Makihara, Yusuke
    Kobatake, Reiko
    Kubo, Takayasu
    Tsuga, Kazuhiro
    JOURNAL OF APPLIED ORAL SCIENCE, 2017, 25 (05) : 498 - 505
  • [30] Assessment of the primary stability of root analog zirconia implants designed using cone beam computed tomography software by means of the Periotest® device: An ex vivo study. A preliminary report
    Matys, Jacek
    Swider, Katarzyna
    Flieger, Rafal
    Dominiak, Marzena
    ADVANCES IN CLINICAL AND EXPERIMENTAL MEDICINE, 2017, 26 (05): : 803 - 809