Initial Stability Measurements of Implants Using a New Magnetic Resonance Frequency Analyzer With Titanium Transducers: An Ex Vivo Study

被引:16
|
作者
Bural, Canan [1 ]
Dayan, Cagatay [2 ]
Geckili, Onur [1 ]
机构
[1] Istanbul Cerrahpasa Univ, Fac Dent, Dept Prosthodont, Istanbul, Turkey
[2] Istanbul Univ, Program Dent Technicians, Istanbul, Turkey
关键词
osseointegration; implantology; experimental design; implant stability; resonance frequency; dental implant; IN-VITRO; DENTAL IMPLANTS; DEVICES; DESIGNS;
D O I
10.1563/aaid-joi-D-19-00126
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The establishment of dental implant stability is mandatory for successful osseointegration. Resonance frequency analysis (RFA) is the most frequently used method for the clinical measurement of implant stability. The purpose of the present study was to evaluate the reliability of the recently developed RF analyzer Penguin RFA and to compare it with the traditional RF analyzer Osstell ISQ. Sixty implants were inserted into fresh steer vertebrae and pelvis. Implant stability was measured using Penguin RFA by its transducers (multipegs) and Osstell ISQ by its transducers (smartpegs). Additionally, stability was measured by multipegs with Osstell ISQ and by smartpegs with Penguin RFA. The intraobserver and interobserver reliability of Penguin RFA were estimated by the intraclass coefficient (ICC). Mean implant stability quotients (ISQs) measured with Osstell ISQ were higher than the ISQs measured with Penguin RFA (P < .05). The intra- and interobserver reliability of Penguin RFA were considered as excellent (ICC > 0.7). For Osstell ISQ, no significance in ISQs was detected between the readings by smartpegs and multipegs (P > .05), while for Penguin RFA ISQs by smartpegs were significantly higher than the ISQs by multipegs (P <.05). The recently developed Penguin RFA is reliable and can be used in clinical practice for the measurement of dental implant stability regardless of the bone type. The multipegs originally manufactured for the Penguin RFA is also compatible with Osstell ISQ.
引用
收藏
页码:35 / 40
页数:6
相关论文
共 37 条
  • [31] Influence of cardiac motion on acquiring accurate measurements using Optical Frequency Domain Imaging (OFDI): Ex vivo study while heart beat and coronary flow being simulated
    Koyama, Kohei
    Kuwata, Shingo
    Mitarai, Takanobu
    Ishibashi, Yuuki
    Yoneyama, Kihei
    Kongoji, Ken
    Harada, Tomoo
    Akashi, Yoshihiro J.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 62 (18) : B196 - B196
  • [32] Brain energy metabolism in a sub-acute rat model of manganese neurotoxicity:: An ex vivo nuclear magnetic resonance study using [1-13C]glucose
    Zwingmann, C
    Leibfritz, D
    Hazell, AS
    NEUROTOXICOLOGY, 2004, 25 (04) : 573 - 587
  • [33] A new method for quantification of aortic stiffness in vivo using magnetic resonance elastography (MRE): a translational study from sequence design to implementation in patients
    Rachel Clough
    Ondrej Holub
    Henry Fok
    Nicholas Gaddum
    Jordi Alastruey
    Ralph Sinkus
    Journal of Cardiovascular Magnetic Resonance, 17 (Suppl 1)
  • [34] Study on High-Frequency 3-D Magnetization Precession Modes of Circular Magnetic Nano-Dots Using Coplanar Wave Guide Vector Network Analyzer Ferromagnetic Resonance
    Miyake, Kousaku
    Noh, Seung-Mo
    Kaneko, Tomoaki
    Imamura, Hiroshi
    Sahashi, Masashi
    IEEE TRANSACTIONS ON MAGNETICS, 2012, 48 (05) : 1782 - 1788
  • [35] Determination of the Initial Apical Canal Diameter by the First File to Bind or Cone-beam Computed Tomographic Measurements Using Micro-computed Tomography as the Gold Standard: An Ex Vivo Study in Human Cadavers
    Campello, Andrea F.
    Marceliano-Alves, Marilia F.
    Siqueira, Jose F., Jr.
    Marques, Fabio V.
    Guedes, Fabio R.
    Lopes, Ricardo T.
    Alvarenga, Fabricia O.
    Alves, Flavio R. F.
    JOURNAL OF ENDODONTICS, 2019, 45 (05) : 619 - 622
  • [36] Ex Vivo Fat Quantification Using Magnetic Resonance Imaging Hepatic Proton Density Fat Fraction in Human Cadaver Liver Predicts Early Graft Dysfunction after Implant: A Proof of Concept Study
    Satapathy, Sanjaya Kumar
    Dyer, Andrew
    Angel, Wesley
    Ranzinger, Rachel
    Jones, Anna V.
    Kedia, Satish
    Jiang, Yu
    Nezakatgoo, Nosratollah
    Helmick, Ryan
    Horton, Peter
    De La Borbolla, Luis Campos
    Vanatta, Jason M.
    Agbim, Uchenna
    Gonzalez, Humberto C.
    Nouer, Simonne S.
    Maliakkal, Benedict J.
    Nair, Satheesh
    Halford, Hollis
    Eason, James D.
    HEPATOLOGY, 2018, 68 : 657A - 657A
  • [37] Anatomic right ventricular measurements with live 3D echocardiography and magnetic resonance imaging: A comparison study using a new 4D analysis method
    Niemann, Petra S.
    Pinho, Luiz
    Balbach, Thomas
    Galuschky, Christian
    Brankenhagen, Michael
    Woodward, William J.
    Imus, Randal L.
    Silberbach, Michael
    Broberg, Craig
    Sheehan, Florence H.
    Jerosch-Herold, Michael
    Sahn, David J.
    CIRCULATION, 2006, 114 (18) : 799 - 799