Assessment of the primary stability of root analog zirconia implants designed using cone beam computed tomography software by means of the Periotest® device: An ex vivo study. A preliminary report

被引:20
|
作者
Matys, Jacek [1 ]
Swider, Katarzyna [1 ]
Flieger, Rafal [1 ]
Dominiak, Marzena [1 ]
机构
[1] Wroclaw Med Univ, Dent Surg Dept, Wroclaw, Poland
来源
关键词
RAZI; primary stability; root analog zirconia implant; Periotest; RESONANCE FREQUENCY-ANALYSIS; DENTAL IMPLANTS; CLINICAL-TRIAL; FOLLOW-UP; IMMEDIATE; BONE; TECHNOLOGY; PLACEMENT; ACCURACY; TORQUE;
D O I
10.17219/acem/65069
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background. The implant primary stability is a fundamental prerequisite for a success of osseointegration process which determines the prosthetic reconstruction time. Objectives. The aim of the present study was to assess the quality and precision of modern conical bone computer tomography (CBCT) software in preparing root analog zirconia implants (RAZIs) by measuring its primary stability by means of the Periotest device. Material and methods. Thirteen pig jaws with proper erupted first premolar (P1) teeth were used in the study. The CBCT examination was conducted in the area of the P1 tooth in each mandible. The 3-dimensional (3D) view of each tooth was designed from CBCT scan. The created 3D images were used to prepare root analog zirconia implants milled from a medical-grade zirconia block by means of laboratory milling. The RAZIs and titanium implants were placed into an alveolar socket after the tooth had been removed. The primary stability of the teeth before their extraction (G1), RAZIs (G2) and titanium implants (G3) were checked by Periotest devices. Results. The mean results in PTV were: 15.9, 3.35, 12.7 for G1, G2 and G3 group, respectively. RAZIs during immediate loading achieved a significantly higher primary stability (lower Periotest value) as compared to the teeth and implants. Conclusions. The modern CBCT device allows us to design a precise image of an extracted tooth for the purpose of manufacturing a root analog implant. The additional feature of the surgical protocol using RAZI is the possibility of avoiding the augmentation procedure, which reduces the whole cost of the treatment.
引用
收藏
页码:803 / 809
页数:7
相关论文
共 5 条
  • [1] Mandibular Bone Structure Analysis Using Cone Beam Computed Tomography vs Primary Implant Stability: An Ex Vivo Study
    Pauwels, Ruben
    Sessirisombat, Somchai
    Panmekiate, Soontra
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2017, 32 (06) : 1257 - +
  • [2] Mapping artifacts generated in a tooth adjacent to titanium and zirconia implants located in the endomass and exomass in cone beam computed tomography: an ex vivo study
    Andrade-Bortoletto, Maria Fernanda Silva
    Fontenele, Rocharles Cavalcante
    Farias-Gomes, Amanda
    Freitas, Deborah Queiroz
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2024, 137 (01): : 73 - 82
  • [3] To Analyze the Mesiobuccal Root Canal of Maxillary First Molar using Cone Beam Computed Tomography: An Ex-vivo Study
    Vasisth, Diwakar
    Surana, Pratik
    Gupta, Ekta
    Bhargava, Triveni
    Sahoo, Nivedita
    Sonkesriya, Subhash
    JOURNAL OF PHARMACY AND BIOALLIED SCIENCES, 2023, 15 : S867 - S870
  • [4] Assessment of the Accuracy of Obturator Bulbs that are Produced by Using Cone Beam Computed Tomography and Stereolithography in Maxilloctemy Defects: An Ex Vivo Study
    Murat, Sema
    JOURNAL OF ACADEMIC RESEARCH IN MEDICINE-JAREM, 2018, 8 (03): : 186 - 190
  • [5] Assessment of furcal perforations in the vicinity of different root canal sealers using a cone beam computed tomography system with and without the application of artifact reduction mode: an ex vivo investigation on extracted human teeth
    Kamburoglu, Kivanc
    Yilmaz, Funda
    Yeta, Elif Naz
    Ozen, Dogukan
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2016, 121 (06): : 657 - 665