Comparison of action levels for patient-specific quality assurance of intensity modulated radiation therapy and volumetric modulated arc therapy treatments

被引:33
|
作者
Mancuso, Gordon M. [2 ,3 ]
Fontenot, Jonas D. [1 ,2 ,3 ]
Gibbons, John P. [1 ,2 ,3 ]
Parker, Brent C. [4 ]
机构
[1] Mary Bird Perkins Canc Ctr, Dept Phys, Baton Rouge, LA 70809 USA
[2] Louisiana State Univ, Dept Phys & Astron, Baton Rouge, LA 70803 USA
[3] Agr & Mech Coll, Baton Rouge, LA 70803 USA
[4] Univ Texas Med Branch, Dept Radiat Oncol, Galveston, TX 77555 USA
关键词
volumetric modulated arc therapy; intensity modulated radiation therapy; quality assurance; action levels; PASSING RATES; IMRT; VMAT; DOSIMETRY; DELIVERY; QA; VERIFICATION; TOMOTHERAPY; PHANTOM;
D O I
10.1118/1.4729738
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To perform a comprehensive and systematic comparison of fixed-beam IMRT and volumetric modulated arc therapy (VMAT) patient-specific QA measurements for a common set of geometries using typical measurement methods. Methods: Fixed-beam IMRT and VMAT plans were constructed for structure set geometries provided by AAPM Task Group 119. The plans were repeatedly delivered across multiple measurement sessions, and the resulting dose distributions were measured with (1) radiochromic film and ionization chamber and (2) a commercial two-dimensional diode array. The resulting QA measurements from each delivery technique were then analyzed, compared, and tested for statistically significant differences. Results: Although differences were noted between QA results for some plans, neither modality showed consistently better agreement of measured and planned doses: of the 22 comparisons, IMRT showed better QA results in 11 cases, and VMAT showed better QA results in 11 cases. No statistically significant differences (p < 0.05) between IMRT and VMAT QA results were found for point doses measured with an ionization chamber, planar doses measured with radiochromic film, or planar doses measured with a two-dimensional diode array. Conclusions: These results suggest that it is appropriate to apply patient-specific QA action levels derived from fixed-beam IMRT to VMAT. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4729738]
引用
收藏
页码:4378 / 4385
页数:8
相关论文
共 50 条
  • [1] Effect of dose grid resolution on the results of patient-specific quality assurance for intensity-modulated radiation therapy and volumetric modulated arc therapy
    Chun, M.
    Kim, J., I
    Oh, G. H.
    Wu, H. G.
    Park, J. M.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2020, 18 (03): : 521 - 530
  • [2] A Systematic Approach to Patient-Specific Quality Assurance for Volumetric Modulated Arc Therapy
    SONG Yu-lin
    Ceferino Obcemea
    Boris Mueller
    Borys Mychalczak
    ChineseJournalofBiomedicalEngineering, 2015, 24 (02) : 58 - 65
  • [3] Prospective Clinical Validation of Virtual Patient-Specific Quality Assurance of Volumetric Modulated Arc Therapy Radiation Therapy Plans
    Wall, Phillip D. H.
    Hirata, Emily
    Morin, Olivier
    Valdes, Gilmer
    Witztum, Alon
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2022, 113 (05): : 1091 - 1102
  • [4] A retrospective analysis for patient-specific quality assurance of volumetric-modulated arc therapy plans
    Li, Guangjun
    Wu, Kui
    Peng, Guang
    Zhang, Yingjie
    Bai, Sen
    MEDICAL DOSIMETRY, 2014, 39 (04) : 309 - 313
  • [5] Comparison of patient-specific intensity modulated radiation therapy quality assurance for the prostate across multiple institutions
    Kubo, Kazuki
    Monzen, Hajime
    Shimomura, Kohei
    Matsumoto, Kenji
    Sato, Tomoharu
    Tamura, Mikoto
    Nakamatsu, Kiyoshi
    Ishii, Kentaro
    Kawamorita, Ryu
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2019, 24 (06) : 600 - 605
  • [6] A study to establish reasonable action limits for patient-specific quality assurance in intensity-modulated radiation therapy
    Both, Stefan
    Alecu, Ionut M.
    Stan, Andrada R.
    Alecu, Marius
    Ciura, Andrei
    Hansen, Jeremy M.
    Alecu, Rodica
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2007, 8 (02): : 1 - 8
  • [7] Dosimetric Comparison of Volumetric Modulated Arc Therapy and Intensity Modulated Radiation Therapy for Whole Brain Hippocampal Sparing Radiation Therapy Treatments
    Kendall, E.
    Higby, C.
    Algan, O.
    Ahmad, S.
    Hossain, S.
    MEDICAL PHYSICS, 2016, 43 (06) : 3587 - 3587
  • [8] Virtual pretreatment patient-specific quality assurance of volumetric modulated arc therapy using deep learning
    Yoganathan, S. A.
    Ahmed, Sharib
    Paloor, Satheesh
    Torfeh, Tarraf
    Aouadi, Souha
    Al-Hammadi, Noora
    Hammoud, Rabih
    MEDICAL PHYSICS, 2023, 50 (12) : 7891 - 7903
  • [9] Using an Electronic Portal Imaging Device for Patient-Specific Volumetric Modulated Arc Therapy Quality Assurance
    Bakhtiari, M.
    Kumaraswamy, L.
    de Boer, S.
    Malhotra, H.
    Podgorsak, M.
    MEDICAL PHYSICS, 2010, 37 (06) : 3087 - +
  • [10] Patient-Specific Quality Assurance Protocol for Volumetric Modulated Arc Therapy using Dose Volume Histogram
    Low, Christopher
    Toye, Warren
    Phung, Peter
    Huston, Christopher
    JOURNAL OF MEDICAL PHYSICS, 2018, 43 (02) : 112 - 118