EPID Dosimetry - Configuration and Pre-Treatment IMRT Verification

被引:0
|
作者
Grzadziel, A. [1 ]
Smolinska, B. [1 ]
Rutkowski, R. [1 ]
Slosarek, K. [1 ]
机构
[1] Ctr Oncol, MSC Mem Inst, Branch Gliwice, Radiotherapy & Brachytherapy Planning Dept, Gliwice, Poland
关键词
IMRT; aSI EPID; treatment plan verification; QA;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Electronic Portal Imaging Device (EPID) is used for patient setup during radiotherapy sessions. Dosimetric verification is done using the ion chambers, diodes and thermoluminescence detectors. In Intensity Modulated Radiation Therapy (IMRT) the dosimetry is sophisticated and time consuming task. The simultaneous use of an amorphous silicon (aSi) detectors and Transit Dosimetry (TD) option in Treatment Planning System (TPS) Eclipse, Varian enables dosimetric pre-treatment verification of the IMRT plans. The purpose of this study was to calibrate the EPID and TPS and to evaluate the usefulness of that method for dose verification in IMRT technique. First step was calibration of the aSi EPID mounted on three linear accelerators Clianc23EXS, Varian. Afterwards, configuration of the calculation algorithm in TPS was carried out. Then dosimetric characteristics of the EPID were investigated. The EPID response depending on the beam mode, treatment time and static square field size was measured. The same measurements were repeated twice for three accelerators and analyzed. Furthermore, three IMRT plans were treated for the pretreatment dose evaluation. The calculated dose matrix was compared with delivered one. The similarity of the calculated and measured fluency maps were evaluated by means of gamma index in Eclipse. The linearity of the EPID signal was proven. For both beam modes EPIDs response is proportional to treatment time and field size, within the considered field size range. However, the obtained results can slightly differ for the different accelerators that are equipped with different imaging devices. The gamma evaluation indicates good correlation between predicted and acquired EPID image. Although some differences in a high gradient area were found. We found EPID based pre-treatment IMRT verification method as a good Quality Assurance (QA) procedure. Quite frequent control of the method and the periodic recalibration of the used device are required.
引用
收藏
页码:1932 / 1936
页数:5
相关论文
共 50 条
  • [31] Sensitivity of an Elekta iView GT a-Si EPID model to delivery errors for pre-treatment verification of IMRT fields
    Sri Herwiningsih
    Peta Hanlon
    Andrew Fielding
    [J]. Australasian Physical & Engineering Sciences in Medicine, 2014, 37 : 763 - 770
  • [32] Sensitivity of an Elekta iView GT a-Si EPID model to delivery errors for pre-treatment verification of IMRT fields
    Herwiningsih, Sri
    Hanlon, Peta
    Fielding, Andrew
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (04) : 763 - 770
  • [33] Patient Specific Pre-treatment QA Verification using an EPID Approach
    Quino, Luis A. Vazquez
    Chen, Xiaoming
    Fitzpatrick, Mathew
    Shi, Chengyu
    Stathakis, Sotirios
    Gutierrez, Alonso
    Esquivel, Carlos
    Mavroidis, Panayiotis
    Alkhatib, Hassan
    Papanikolaou, Niko
    [J]. TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2014, 13 (01) : 1 - 10
  • [34] New imaging devices for pre-treatment plan verification in IMRT
    Slosarek, K.
    Smolinska, B.
    Rutkowski, R.
    Grzadziel, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S466 - S466
  • [35] Transit dosimetry in IMRT with an a-Si EPID in direct detection configuration
    Sabet, Mahsheed
    Rowshanfarzad, Pejman
    Vial, Philip
    Menk, Frederick W.
    Greer, Peter B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (15): : N295 - N306
  • [36] Large scale implementation of clinical EPID dosimetry for IMRT dose verification
    Mans, A.
    Stroom, J.
    Tielenburg, R.
    Vijlbrief, R.
    Buijs, M.
    Wendling, M.
    McDermott, L.
    Van Herk, M.
    Mijnheer, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S99 - S99
  • [37] Portal Dosimetry for Pre-Treatment Verification of IMRT Plan: A Comparison with 2D Ion Chamber Array
    Sharma, D.
    Mhatre, V.
    Heigrujam, M.
    Talapatra, K.
    Mallik, S.
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 3234 - +
  • [38] Independent verification of gantry angle for pre-treatment VMAT QA using EPID
    Department of Radiation Oncology, Duke University Medical Center, Durham, NC 27707, United States
    [J]. Phys. Med. Biol., 20 (6587-6600):
  • [39] Independent verification of gantry angle for pre-treatment VMAT QA using EPID
    Adamson, Justus
    Wu, Qiuwen
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (20): : 6587 - 6600
  • [40] EPID-based 3D dosimetry for pre-treatment FFF VMAT stereotactic body radiotherapy plan verification using dosimetry Check™
    Chendi, Agnese
    Botti, Andrea
    Orlandi, Matteo
    Sghedoni, Roberto
    Iori, Mauro
    Cagni, Elisabetta
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 81 : 227 - 236