Design and implementation of a film dosimetry audit tool for comparison of planned and delivered dose distributions in high dose rate (HDR) brachytherapy

被引:36
|
作者
Palmer, Antony L. [1 ,2 ]
Lee, Chris [3 ]
Ratcliffe, Ailsa J. [4 ]
Bradley, David [1 ]
Nisbet, Andrew [1 ,5 ]
机构
[1] Univ Surrey, Fac Engn & Phys Sci, Dept Phys, Surrey GU2 7JP, England
[2] Portsmouth Hosp NHS Trust, Queen Alexandra Hosp, Dept Med Phys, Portsmouth PO6 3LY, Hants, England
[3] Clatterbridge Canc Ctr NHS Fdn Trust, Wirral CH63 4JY, Merseyside, England
[4] Univ Hosp Coventry & Warwickshire NHS Trust, Dept Clin Phys & Bioengn, Coventry CV2 2DX, W Midlands, England
[5] Royal Surrey Cty Hosp NHS Fdn Trust, Dept Med Phys, Guildford GU2 7XX, Surrey, England
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2013年 / 58卷 / 19期
关键词
RADIOCHROMIC FILM; IR-192; BRACHYTHERAPY; QUALITY-CONTROL; GAFCHROMIC FILM; RECONSTRUCTION; AAPM; RECOMMENDATIONS; ACCURACY; VERIFICATION; CALIBRATION;
D O I
10.1088/0031-9155/58/19/6623
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A novel phantom is presented for 'full system' dosimetric audit comparing planned and delivered dose distributions in HDR gynaecological brachytherapy, using clinical treatment applicators. The brachytherapy applicator dosimetry test object consists of a near full-scatter water tank with applicator and film supports constructed of Solid Water, accommodating any typical cervix applicator. Film dosimeters are precisely held in four orthogonal planes bisecting the intrauterine tube, sampling dose distributions in the high risk clinical target volume, points A and B, bladder, rectum and sigmoid. The applicator position is fixed prior to CT scanning and through treatment planning and irradiation. The CT data is acquired with the applicator in a near clinical orientation to include applicator reconstruction in the system test. Gamma analysis is used to compare treatment planning system exported RTDose grid with measured multi-channel film dose maps. Results from two pilot audits are presented, using Ir-192 and Co-60 HDR sources, with a mean gamma passing rate of 98.6% using criteria of 3% local normalization and 3 mm distance to agreement (DTA). The mean DTA between prescribed dose and measured film dose at point A was 1.2 mm. The phantom was funded by IPEM and will be used for a UK national brachytherapy dosimetry audit.
引用
收藏
页码:6623 / 6640
页数:18
相关论文
共 50 条
  • [41] High dose rate (HDR) brachytherapy boost for prostate cancer gives high control rate
    Flynn, J. P.
    Kelly, D. A.
    Brookover, T.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : S323 - S324
  • [42] Interstitial High dose rate Brachytherapie (HDR-13T) of lip tumors, retrospective analysis in comparison with low dose rate brachytherapy
    Friedrich, E. E.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2007, 183 : 88 - 88
  • [43] Radiochromic film dosimetry for verification of dose distributions delivered with proton-beam radiosurgery
    Vatnitsky, SM
    Schulte, RWM
    Galindo, R
    Meinass, HJ
    Miller, DW
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (10): : 1887 - 1898
  • [44] Inversely planned catheter positions for high dose rate brachytherapy of the prostate
    Lessard, E
    Beaulieu, L
    Hsu, I
    Pouliot, J
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1957 - 1957
  • [45] Dose Delivered to the Lumbosacral Plexus From High-Dose-Rate Brachytherapy for Cervical Cancer
    Rash, Dominique
    Durbin-Johnson, Blythe
    Lim, Jihoon
    Dieterich, Sonja
    Huddleston, Adam
    Yi, Sun
    Mayadev, Jyoti
    [J]. INTERNATIONAL JOURNAL OF GYNECOLOGICAL CANCER, 2015, 25 (05) : 897 - 902
  • [46] Feasibility study of a dosimeter for in vivo measurements of the absolute dose delivered in high dose rate brachytherapy
    Carrara, M.
    Gambarini, G.
    Borroni, M.
    Tomatis, S.
    Cerrotta, A.
    Fallai, C.
    Olmi, P.
    Zonca, G.
    [J]. NUOVO CIMENTO DELLA SOCIETA ITALIANA DI FISICA B-BASIC TOPICS IN PHYSICS, 2010, 125 (5-6): : 667 - 673
  • [47] Temporal-Spatial Dose Pattern in Vaginal Cylinder Applicator for High Dose Rate (HDR) Brachytherapy
    Li, K.
    Saini, A.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : E369 - E369
  • [48] Monte Carlo dosimetry of high dose rate gynecologic interstitial brachytherapy
    Desbiens, Maxime
    D'Amours, Michel
    Afsharpour, Hossein
    Verhaegen, Frank
    Lavallee, Marie-Claude
    Thibault, Isabelle
    Vigneault, Eric
    Beaulieu, Luc
    [J]. RADIOTHERAPY AND ONCOLOGY, 2013, 109 (03) : 425 - 429
  • [49] Comparison of single and multiple dwell position methods in MammoSite high dose rate (HDR) brachytherapy planning
    Kim, Yongbok
    Trombetta, Mark G.
    Miften, Moyed
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2010, 11 (03): : 54 - 63
  • [50] In vivo dosimetry in high dose rate brachytherapy for cervical cancer treatments
    Gonzalez-Azcorra, S. A.
    Mota-Garcia, A.
    Poitevin-Chacon, M. A.
    Santamaria-Torruco, B. J.
    Rodriguez-Ponce, M.
    Herrera-Martinez, F. P.
    Gamboa, de Buen, I
    Ruiz-Trejo, C.
    Buenfil, A. E.