Design and implementation of a film dosimetry audit tool for comparison of planned and delivered dose distributions in high dose rate (HDR) brachytherapy

被引:36
|
作者
Palmer, Antony L. [1 ,2 ]
Lee, Chris [3 ]
Ratcliffe, Ailsa J. [4 ]
Bradley, David [1 ]
Nisbet, Andrew [1 ,5 ]
机构
[1] Univ Surrey, Fac Engn & Phys Sci, Dept Phys, Surrey GU2 7JP, England
[2] Portsmouth Hosp NHS Trust, Queen Alexandra Hosp, Dept Med Phys, Portsmouth PO6 3LY, Hants, England
[3] Clatterbridge Canc Ctr NHS Fdn Trust, Wirral CH63 4JY, Merseyside, England
[4] Univ Hosp Coventry & Warwickshire NHS Trust, Dept Clin Phys & Bioengn, Coventry CV2 2DX, W Midlands, England
[5] Royal Surrey Cty Hosp NHS Fdn Trust, Dept Med Phys, Guildford GU2 7XX, Surrey, England
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2013年 / 58卷 / 19期
关键词
RADIOCHROMIC FILM; IR-192; BRACHYTHERAPY; QUALITY-CONTROL; GAFCHROMIC FILM; RECONSTRUCTION; AAPM; RECOMMENDATIONS; ACCURACY; VERIFICATION; CALIBRATION;
D O I
10.1088/0031-9155/58/19/6623
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A novel phantom is presented for 'full system' dosimetric audit comparing planned and delivered dose distributions in HDR gynaecological brachytherapy, using clinical treatment applicators. The brachytherapy applicator dosimetry test object consists of a near full-scatter water tank with applicator and film supports constructed of Solid Water, accommodating any typical cervix applicator. Film dosimeters are precisely held in four orthogonal planes bisecting the intrauterine tube, sampling dose distributions in the high risk clinical target volume, points A and B, bladder, rectum and sigmoid. The applicator position is fixed prior to CT scanning and through treatment planning and irradiation. The CT data is acquired with the applicator in a near clinical orientation to include applicator reconstruction in the system test. Gamma analysis is used to compare treatment planning system exported RTDose grid with measured multi-channel film dose maps. Results from two pilot audits are presented, using Ir-192 and Co-60 HDR sources, with a mean gamma passing rate of 98.6% using criteria of 3% local normalization and 3 mm distance to agreement (DTA). The mean DTA between prescribed dose and measured film dose at point A was 1.2 mm. The phantom was funded by IPEM and will be used for a UK national brachytherapy dosimetry audit.
引用
收藏
页码:6623 / 6640
页数:18
相关论文
共 50 条
  • [31] Dose determination in rectum and urethra with TLD dosimeters in prostate brachytherapy with High Dose Rate (HDR)
    Gonzalez, J
    Rodriguez, A
    Crelgo, D
    Villacé, A
    Fernandez, P
    Sanchez, A
    Vivanco, J
    Sanz, D
    De la Ru, A
    Prada, P
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S135 - S135
  • [32] Results of high dose rate (HDR) brachytherapy boost for prostate cancer
    Kelly, D
    Flynn, J
    Brookover, W
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 71 : S18 - S18
  • [33] Commissioning Integrated System in High-Dose-Rate (HDR) Brachytherapy
    Saw, C.
    Juliano, J.
    Wagner, H.
    [J]. MEDICAL PHYSICS, 2011, 38 (06) : 3516 - U64
  • [34] EFFECT OF ANISOTROPY CORRECTIONS ON THE DYNAMIC DOSE CALCULATIONS IN HIGH-DOSE-RATE (HDR) BRACHYTHERAPY
    CHO, SH
    MULLERRUNKEL, R
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (07): : 1181 - 1188
  • [35] Single dose high dose-rate (HDR) brachytherapy as monotherapy for localised prostate cancer
    Tharmalingam, H.
    Tsang, Y. M.
    Hoskin, P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S337 - S337
  • [36] An automated optimization tool for high-dose-rate (HDR) prostate brachytherapy with divergent needle pattern
    de Battisti, M. Borot
    Maenhout, M.
    de Senneville, B. Denis
    Hautvast, G.
    Binnekamp, D.
    Lagendijk, J. J. W.
    van Vulpen, M.
    Moerland, M. A.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (19): : 7567 - 7583
  • [37] Comparison of Two VMAT Planning Techniques Using Planned and Delivered Dose Distributions
    Mathew, D.
    Edlund, S.
    Watanabe, Y.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2913 - 2913
  • [38] Impact of new software tool in treatment planning for prostate high-dose rate (HDR) brachytherapy
    Sankreacha, R
    Morton, G
    Keller, B
    [J]. MEDICAL PHYSICS, 2003, 30 (06) : 1388 - 1388
  • [39] High dose rate (HDR) prostate boost brachytherapy: A comparison of acute and chronic toxicities between three dose escalation groups
    Boyea, G
    Martinez, A
    Wallace, M
    Chen, P
    Gustafson, G
    Lewis, L
    Flynn, C
    Axiak, A
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S188 - S188
  • [40] Verification of high dose rate brachytherapy dose distributions with EBT3 Gafchromic film quality control techniques
    Palmer, Antony L.
    Nisbet, Andrew
    Bradley, David
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (03): : 497 - 511