Dose Reduction in Automatic Optimization Parameter of Full Field Digital Mammography: Breast Phantom Study

被引:8
|
作者
Ko, Myung-Su [1 ]
Kim, Hak Hee [2 ,3 ]
Cha, Joo Hee [2 ,3 ]
Shin, Hee Jung [2 ,3 ]
Kim, Jeoung Hyun [4 ]
Kim, Min Jeong [1 ]
机构
[1] Asan Med Ctr, Hlth Screening & Promot Ctr, Seoul 138736, South Korea
[2] Univ Ulsan, Coll Med, Asan Med Ctr, Dept Radiol, Seoul, South Korea
[3] Univ Ulsan, Coll Med, Asan Med Ctr, Res Inst Radiol, Seoul, South Korea
[4] Kyung Hee Univ, Coll Med, Kyung Hee Univ Hosp Gangdong, Dept Radiol, Seoul, South Korea
关键词
Breast; Imaging phantom; Mammography; Radiation dosage; SCREEN-FILM MAMMOGRAPHY;
D O I
10.4048/jbc.2013.16.1.90
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: We evaluated the impact of three automatic optimization of parameters (AOP) modes of digital mammography on the dose and image quality. Methods: Computerized Imaging Reference Systems phantoms were used. A total of 12 phantoms with different thickness and glandularity were imaged. We analyzed the average glandular dose (AGD) and entrance surface exposure (ESE) of 12 phantoms imaged by digital mammography in three modes of AOP; namely standard mode (STD), contrast mode (CNT), and dose mode (DOSE). Moreover, exposure factors including kVp, mAs, and target/filter combination were evaluated. To evaluate the quality of the obtained digital image, two radiologists independently counted the objects of the phantoms. Results: According to the AOP modes, the score of masses and specks was sorted as CNT>STD = DOSE. There was no difference in the score of fiber among the three modes. The score of image preference was sorted as CNT>STD>DOSE. The AGD, ESE, and mAs were sorted as CNT>STD>DOSE. The kVp was sorted as CNT=STD>DOSE. The score of all test objects in the phantom image was on a downtrend with increasing breast thickness. The score of masses was different among the three groups; 20-21%>30%>50% glandularity. The score of specks was sorted as 20-21%=30%>50% glandularity. The score of fibers was sorted as 30% >20-21% = 50% glandularity. The score of image preference was not different among the three glandularity groups. The AGD, ESE, kVp, and mAs were correlated with breast thickness, but not correlated with glandularity. Conclusion: The DOSE mode offers significant improvement (19.1-50%) in dose over the other two modes over a range of breast thickness and breast glandularity with acceptable image quality. Owning knowledge of the three AOP modes may reduce unnecessary radiation exposure by utilizing the proper mode according to its purpose.
引用
收藏
页码:90 / 96
页数:7
相关论文
共 50 条
  • [1] Dose reduction in full-field digital mammography: an anthropomorphic breast phantom study
    Obenauer, S
    Hermann, KP
    Grabbe, E
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (907): : 478 - 482
  • [2] Dose reduction in full-field digital mammography - An anthropomorphic breast phantom study.
    Obenauer, S
    Hermann, KP
    Grabbe, E
    [J]. DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2003, : 145 - 148
  • [3] Dose reduction in Full-Field Digital Electromagnetic Radiation Mammography using Noise reduction Method: Phantom Study
    Choi, Seokyoon
    Seo, Jeongmin
    [J]. JOURNAL OF MAGNETICS, 2019, 24 (04) : 744 - 751
  • [4] Evaluation of breast phantom masses with full field digital mammography
    Yang, WT
    Lai, CJ
    Whitman, GJ
    Murphy, WA
    Dryden, MJ
    Kushwaha, AC
    Sahin, AA
    Shaw, C
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 182 (04) : 107 - 108
  • [5] Eye radiation dose from breast cancer screening using full field digital mammography and digital breast tomosynthesis: A phantom study
    Ali, R. M. K. M.
    Hogg, P.
    [J]. RADIOGRAPHY, 2024, 30 (01) : 141 - 144
  • [6] Automatic evaluation of breast density for full-field digital mammography
    Lou, SL
    Fan, Y
    [J]. MEDICAL IMAGING 2000: IMAGE PROCESSING, PTS 1 AND 2, 2000, 3979 : 1362 - 1369
  • [7] Digital luminescence mammography (CR) versus full-field digital mammography (DR): A phantom study
    Schulz-Wendtland, RW
    Aichinger, UG
    Lell, MM
    Kuchar, I
    Tartsch, M
    Bautz, WA
    [J]. RADIOLOGY, 2002, 225 : 416 - 417
  • [8] X-ray spectrum optimization of full-field digital mammography: Simulation and phantom study
    Bernhardt, Philipp
    Mertelmeier, Thomas
    Hoheisel, Martin
    [J]. MEDICAL PHYSICS, 2006, 33 (11) : 4337 - 4349
  • [9] Threshold in breast compression reduction for full-field digital mammography and digital breast tomosynthesis
    Afandy, A. N.
    Tori, M. B.
    Bintalib, S. O.
    Soh, B. L. P.
    [J]. RADIOGRAPHY, 2024, 30 (01) : 217 - 225
  • [10] Full-field digital mammography in comparison to screen-film mammography - A phantom study
    Obenauer, S
    Schorn, C
    Funke, M
    Fischer, U
    Grabbe, E
    Hermann, KP
    [J]. IWDM 2000: 5TH INTERNATIONAL WORKSHOP ON DIGITAL MAMMOGRAPHY, 2001, : 499 - 503