Effect of X-Ray Tube Voltage and Radiation Dose on Modulation Transfer Function in Digital Mammography

被引:0
|
作者
Karaaslan, Melike Kaya [1 ]
Avci, Onur Muhammed [2 ]
Muzoglu, Nedim [2 ]
Gundogdu, Ozcan [1 ]
机构
[1] Kocaeli Univ, Inst Sci, Dept Biomed Engn, Kocaeli, Turkey
[2] Istanbul Univ, Inst Sci, Dept Biomed Engn, Istanbul, Turkey
关键词
digital mammography; modulation transfer function (MTF); radiation dose; X-ray tube voltage; MTF;
D O I
10.1109/TIPTEKNO56568.2022.9960162
中图分类号
Q813 [细胞工程];
学科分类号
摘要
The Modulation Transfer Function (MTF) is a physical metric that explains the relationship between spatial resolution and contrast in the frequency domain for image quality evaluation. This study investigates the effect of the X-ray tube voltage (kVp) and radiation dose on MTF measurement in digital mammography. The MTF was determined for four radiation dose levels at two different kVp values according to the IEC 62220-1-2 standard. In conclusion, there was no significant difference between the results, and the MTF was determined to be independent of kVp and radiation dose.