Hip resurfacing in a district general hospital: 6-year clinical results using the ReCap hip resurfacing system

被引:8
|
作者
van der Weegen, Walter [1 ]
Hoekstra, Henk J. [1 ]
Sijbesma, Thea [1 ]
Austen, Shennah [1 ]
Poolman, Rudolf W. [2 ]
机构
[1] St Anna Hosp, Dept Orthopaed Surg, NL-5664 EH Geldrop, Netherlands
[2] Onze Lieve Vrouw Hosp, Dept Orthopaed Surg, Amsterdam, Netherlands
来源
关键词
Hip resurfacing; Implant survival; Adverse reaction to metal debris; ReCap design; OF-THE-LITERATURE; RISK-FACTORS; PSEUDOTUMOR FORMATION; CHROMIUM IONS; ARTHROPLASTY; REPLACEMENT; FAILURE; COBALT; SURVIVORSHIP; COMPONENT;
D O I
10.1186/1471-2474-13-247
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The purpose of our study was to prospectively report the clinical results of 280 consecutive hips (240 patients) who received a ReCap Hip Resurfacing System implant (Biomet Inc., Warsaw, USA) in a single district general hospital. Literature reports a large variation in clinical results between different resurfacing designs and published results using this particular design are scarce. Methods: Mean follow up was 3.3 years (1.0 to 6.3) and four patients were lost to follow-up. All patients were diagnosed with end-stage hip osteoarthritis, their mean age was 54 years and 76.4% of all patients were male. Results: There were 16 revisions and four patients reported a Harris Hip Score < 70 points at their latest follow up. There were no pending revisions. Kaplan-Meier implant survival probability, with revision for any reason as endpoint, was 93.5% at six years follow-up (95%-CI: 88.8-95.3). There were no revisions for Adverse Reactions to Metal Debris (ARMD) and no indications of ARMD in symptomatic non-revised patients, although diagnostics were limited to ultrasound scans. Conclusions: This independent series confirms that hip resurfacing is a demanding procedure, and that implant survival of the ReCap hip resurfacing system is on a critical level in our series. In non-revised patients, reported outcomes are generally excellent.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Five Year Results of the First US FDA-Approved Hip Resurfacing Device
    Su, Edwin P.
    Housman, Lawrence R.
    Masonis, John L.
    Noble, John W., Jr.
    Engh, C. Anderson
    [J]. JOURNAL OF ARTHROPLASTY, 2014, 29 (08): : 1571 - 1575
  • [22] Operationstechnik und klinische Ergebnisse des DUROM™-Hip-Resurfacing-Systems mit dem lateralen ZugangSurgical Principles and Clinical Experiences with the DUROM™ Hip Resurfacing System Using a Lateral Approach
    Sascha Gravius
    Torsten Mumme
    Oliver Weber
    Philipp Berdel
    Dieter Christian Wirtz
    [J]. Operative Orthopädie und Traumatologie, 2009, 21 : 586 - 601
  • [23] The mean ten-year results of metal-on-metal hybrid hip resurfacing arthroplasty
    Amstutz, H. C.
    Le Duff, M. J.
    [J]. BONE & JOINT JOURNAL, 2018, 100B (11): : 1424 - 1433
  • [24] Durom® hip resurfacing arthroplasty: Five-year results of the first 50 consecutive patients
    Naal, F. D.
    Pilz, R.
    Guggi, T.
    Munzinger, U.
    Hersche, O.
    Leunig, M.
    [J]. SWISS MEDICAL WEEKLY, 2010, 140 (23-24) : 2S - 2S
  • [25] 10-year results of the Birmingham Hip Resurfacing: a non-designer case series
    Hunter, Thomas J. A.
    Moores, Thomas S.
    Morley, David
    Manoharan, Gopikanthan
    Collier, Simon G.
    Shaylor, Phillip J.
    [J]. HIP INTERNATIONAL, 2018, 28 (01) : 50 - 52
  • [26] Computer-assisted hip resurfacing surgery using the Acrobot® Navigation System
    Barrett, A. R. W.
    Davies, B. L.
    Gomes, M. P. S. F.
    Harris, S. I.
    Henckel, J.
    Jakopec, M.
    Kannan, V.
    Rodriguez y Baena, F. M.
    Cobb, J. P.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H7) : 773 - 785
  • [27] Clinical and Radiographic Results of Metal-on-Metal Hip Resurfacing with a Minimum Ten-Year Follow-up
    Amstutz, Harlan C.
    Le Duff, Michel J.
    Campbell, Patricia A.
    Gruen, Thomas A.
    Wisk, Lauren E.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (16): : 2663 - 2671
  • [28] Cup Anteversion in Hip Resurfacing: Validation of EBRA and the Presentation of a Simple Clinical Grading System
    Langton, David J.
    Sprowson, Andrew P.
    Mahadeva, Dhirendra
    Bhatnagar, Sharad
    Holland, James P.
    Nargol, Antoni V. F.
    [J]. JOURNAL OF ARTHROPLASTY, 2010, 25 (04): : 607 - 613
  • [29] RESURFACING ARTHROPLASTY OF THE HIP - BIOMECHANICAL, MORPHOLOGICAL, AND CLINICAL ASPECTS BASED ON THE RESULTS OF A PRELIMINARY CLINICAL-STUDY
    GOLDIE, IF
    BUNKETORP, O
    GUNTERBERG, B
    HANSSON, T
    MYRHAGE, R
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1979, 95 (03) : 149 - 157
  • [30] Lessons learned from early clinical experience and results of 300 ASR® hip resurfacing implantations
    Siebel, T
    Maubach, S
    Morlock, MM
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) : 345 - 353