Precision and accuracy of commercial assays for vancomycin therapeutic drug monitoring: evaluation based on external quality assessment scheme

被引:19
|
作者
Chen, Chao-Yang [1 ]
Li, Meng-Ya [1 ,2 ]
Ma, Ling-Yun [1 ]
Zhai, Xing-Yu [1 ,2 ]
Luo, Dao-Huang [1 ,2 ]
Zhou, Ying [1 ,2 ]
Liu, Zhen-Ming [3 ]
Cui, Yi-Min [1 ,2 ]
机构
[1] Peking Univ, Dept Pharm, Hosp 1, Beijing, Peoples R China
[2] Peking Univ, Sch Pharmaceut Sci, Dept Pharm Adm & Clin Pharm, Hlth Sci Ctr, Beijing, Peoples R China
[3] Peking Univ, Sch Pharmaceut Sci, State Key Lab Nat & Biomimet Drugs, Hlth Sci Ctr, Beijing, Peoples R China
关键词
FLUORESCENCE POLARIZATION IMMUNOASSAY; LIQUID-CHROMATOGRAPHY; PERFORMANCE; PHARMACOKINETICS; STANDARD; SERUM;
D O I
10.1093/jac/dkaa150
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Background: Vancomycin remains a mainstay of the treatment of Gram-positive bacterial infections. It is crucial to accurately determine vancomycin serum concentration for adequate dose adjustment. Objectives: To evaluate the precision and accuracy of commercial assay techniques for vancomycin concentration and to assess the comparability of vancomycin detection methods in Chinese laboratories. Methods: Human serum samples spiked with known concentrations of vancomycin were provided to laboratories participating in the external quality assessment scheme (EQAS). Assay methods included chemiluminescence, enzyme immunoassay (EIA) and so on. The dispersion of the measurements was analysed and the robust coefficient of variation (rCV), relative percentage difference (RPD) and satisfactory rate for method groups were calculated. Moreover, performance of the Chinese laboratories was assessed. Results: A total of 657 results from 75 laboratories were collected, including 84 samples from 10 Chinese laboratories. The median rCV, median RPD and satisfactory rates classified by methods ranged from 1.85% to 15.87%, #14.75% to 13.34% and 94.59% to 100.00%, respectively. Significant differences were seen in precision, between kinetic interaction of microparticles in solution (KIMS) and other methods, and in accuracy, between enzyme-multiplied immunoassay technique (EMIT), fluorescence polarization immunoassay (FPIA) and other techniques. Vancomycin detection in China mainly depended on the chemiluminescence and EMIT methods, which tended to result in lowermeasurements. Conclusions: Although almost all assays in this study achieved an acceptable performance for vancomycin serumconcentration monitoring, obvious inconsistencies between methods were still observed. Chinese laboratories were more likely to underestimate vancomycin concentrations. Thus, recognizing inconsistencies between methods and regular participation in vancomycin EQAS are essential.
引用
收藏
页码:2110 / 2119
页数:10
相关论文
共 50 条
  • [41] Influence of apixaban on commonly used coagulation assays: results from the Belgian national External Quality Assessment Scheme
    Van Blerk, M.
    Bailleul, E.
    Chatelain, B.
    Demulder, A.
    Devreese, K.
    Douxfils, J.
    Jacquemin, M.
    Jochmans, K.
    Mullier, F.
    Wijns, W.
    China, B.
    Vernelen, K.
    Soumali, M. R.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2017, 39 (04) : 402 - 408
  • [42] Performance evaluation of 14 specific proteins measurement checked by an External Quality Assessment Scheme
    Secchiero, Sandra
    Sciacovelli, Laura
    Plebani, Mario
    CLINICA CHIMICA ACTA, 2020, 502 : 73 - 83
  • [43] A RNA based external quality assessment pilot scheme for the molecular testing of sarcomas
    Deans, S.
    Fairley, J.
    VIRCHOWS ARCHIV, 2014, 465 : S323 - S323
  • [44] External Quality Assessment Scheme for HIV-1 Drug-Resistance Genotyping in Thailand
    Saeng-aroon, Siriphan
    Saipradit, Nonglak
    Loket, Ruangchai
    Klamkhai, Nattapong
    Boonmuang, Ratrawee
    Kaewprommal, Pavita
    Prommajan, Korrakot
    Takeda, Naokazu
    Sungkanuparph, Somnuek
    Shioda, Tatsuo
    Sangkitporn, Somchai
    Motomura, Kazushi
    AIDS RESEARCH AND HUMAN RETROVIRUSES, 2018, 34 (12) : 1028 - 1035
  • [45] Characterisation and evaluation of external quality assessment scheme serum - Discussion paper from the European external quality assessment (EQA) organisers working group C
    Middle, JG
    Libeer, JC
    Malakhov, V
    Penttila, I
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 1998, 36 (02) : 119 - 130
  • [46] An evaluation of analytic coals for assays of drugs - A College of American Pathologists Therapeutic Drug Monitoring Survey study
    Steele, BW
    Wang, E
    Palomaki, GE
    Klee, GG
    Elin, RJ
    Soldin, SJ
    Witte, DL
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2001, 125 (06) : 729 - 735
  • [47] IMPLEMENTATION AND EVALUATION OF A PILOT PHARMACIST-LED VANCOMYCIN THERAPEUTIC DRUG MONITORING SERVICE IN AN ACUTE GENERAL HOSPITAL
    Cheung, E. M.
    Mo, K. Y.
    Ng, T. M.
    Leung, W. Y.
    Lam, T. T.
    VALUE IN HEALTH, 2018, 21 : S62 - S62
  • [48] Intravenous Vancomycin Therapeutic Drug Monitoring in Children: Evaluation of a Pharmacy-Driven Protocol and Collaborative Practice Agreement
    Olson, Jared
    Hersh, Adam L.
    Sorensen, Jeffrey
    Zobell, Jeffrey
    Anderson, Collin
    Thorell, Emily A.
    JOURNAL OF THE PEDIATRIC INFECTIOUS DISEASES SOCIETY, 2020, 9 (03) : 334 - 341
  • [49] ADAPTATION AND EVALUATION OF MONOCLONAL-ANTIBODY BASED THERAPEUTIC DRUG ASSAYS ON COULTER CPA
    LUCAS, F
    LANDRUM, E
    HO, S
    RICARDO, R
    BEDEVIA, J
    CLINICAL CHEMISTRY, 1987, 33 (06) : 943 - 943
  • [50] Population pharmacokinetic model of Vancomycin based on therapeutic drug monitoring data in critically ill septic patients
    Kovacevic, Tijana
    Miljkovic, Branislava
    Kovacevic, Pedja
    Dragic, Sasa
    Momcicevic, Danica
    Avram, Sanja
    Jovanovic, Marija
    Vucicevic, Katarina
    JOURNAL OF CRITICAL CARE, 2020, 55 : 116 - 121