Precision and accuracy of commercial assays for vancomycin therapeutic drug monitoring: evaluation based on external quality assessment scheme

被引:19
|
作者
Chen, Chao-Yang [1 ]
Li, Meng-Ya [1 ,2 ]
Ma, Ling-Yun [1 ]
Zhai, Xing-Yu [1 ,2 ]
Luo, Dao-Huang [1 ,2 ]
Zhou, Ying [1 ,2 ]
Liu, Zhen-Ming [3 ]
Cui, Yi-Min [1 ,2 ]
机构
[1] Peking Univ, Dept Pharm, Hosp 1, Beijing, Peoples R China
[2] Peking Univ, Sch Pharmaceut Sci, Dept Pharm Adm & Clin Pharm, Hlth Sci Ctr, Beijing, Peoples R China
[3] Peking Univ, Sch Pharmaceut Sci, State Key Lab Nat & Biomimet Drugs, Hlth Sci Ctr, Beijing, Peoples R China
关键词
FLUORESCENCE POLARIZATION IMMUNOASSAY; LIQUID-CHROMATOGRAPHY; PERFORMANCE; PHARMACOKINETICS; STANDARD; SERUM;
D O I
10.1093/jac/dkaa150
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Background: Vancomycin remains a mainstay of the treatment of Gram-positive bacterial infections. It is crucial to accurately determine vancomycin serum concentration for adequate dose adjustment. Objectives: To evaluate the precision and accuracy of commercial assay techniques for vancomycin concentration and to assess the comparability of vancomycin detection methods in Chinese laboratories. Methods: Human serum samples spiked with known concentrations of vancomycin were provided to laboratories participating in the external quality assessment scheme (EQAS). Assay methods included chemiluminescence, enzyme immunoassay (EIA) and so on. The dispersion of the measurements was analysed and the robust coefficient of variation (rCV), relative percentage difference (RPD) and satisfactory rate for method groups were calculated. Moreover, performance of the Chinese laboratories was assessed. Results: A total of 657 results from 75 laboratories were collected, including 84 samples from 10 Chinese laboratories. The median rCV, median RPD and satisfactory rates classified by methods ranged from 1.85% to 15.87%, #14.75% to 13.34% and 94.59% to 100.00%, respectively. Significant differences were seen in precision, between kinetic interaction of microparticles in solution (KIMS) and other methods, and in accuracy, between enzyme-multiplied immunoassay technique (EMIT), fluorescence polarization immunoassay (FPIA) and other techniques. Vancomycin detection in China mainly depended on the chemiluminescence and EMIT methods, which tended to result in lowermeasurements. Conclusions: Although almost all assays in this study achieved an acceptable performance for vancomycin serumconcentration monitoring, obvious inconsistencies between methods were still observed. Chinese laboratories were more likely to underestimate vancomycin concentrations. Thus, recognizing inconsistencies between methods and regular participation in vancomycin EQAS are essential.
引用
收藏
页码:2110 / 2119
页数:10
相关论文
共 50 条
  • [31] Evaluation of the Analytical Performance of Two Therapeutic Drug Monitoring Assays on the Atellica CI Analyzer
    Huff, J.
    Valdivia, B.
    Wilkins, G.
    Arrode-Bruses, G.
    Leipold, H.
    Snyder, J.
    CLINICAL CHEMISTRY, 2024, 70 : I242 - I242
  • [32] THERAPEUTIC DRUG MONITORING OF VANCOMYCIN (TDM-V) IN NEONATES AND PAEDIATRICS: EVALUATION OF GUIDELINES AND RECOMMENDATIONS
    Diaz, L.
    Garcia, I
    Hernandez, R.
    Munoz, M.
    Frias, J.
    Ramirez, E.
    Borobia, A. M.
    Carcas, A. J.
    BASIC & CLINICAL PHARMACOLOGY & TOXICOLOGY, 2016, 119 : 33 - 33
  • [33] QUALITY-CONTROL IN THERAPEUTIC DRUG-MONITORING - INTRA-LABORATORY PRECISION
    GERSON, B
    ANNALS OF CLINICAL AND LABORATORY SCIENCE, 1980, 10 (05): : 445 - 445
  • [34] Clinical Practice Guidelines for Therapeutic Drug Monitoring of Vancomycin in the Framework of Model-Informed Precision Dosing: A Consensus Review by the Japanese Society of Chemotherapy and the Japanese Society of Therapeutic Drug Monitoring
    Matsumoto, Kazuaki
    Oda, Kazutaka
    Shoji, Kensuke
    Hanai, Yuki
    Takahashi, Yoshiko
    Fujii, Satoshi
    Hamada, Yukihiro
    Kimura, Toshimi
    Mayumi, Toshihiko
    Ueda, Takashi
    Nakajima, Kazuhiko
    Takesue, Yoshio
    PHARMACEUTICS, 2022, 14 (03)
  • [35] Evaluation of clinical cases in External Quality Assessment Scheme (EQAS) for the urinary sediment
    Fogazzi, Giovanni B.
    Secchiero, Sandra
    Garigali, Giuseppe
    Plebani, Mario
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2014, 52 (06) : 845 - 852
  • [36] Results and feasibility of an external quality assessment scheme for self-monitoring of blood glucose
    Kristensen, Gunn B. B.
    Nerhus, Kari
    Thue, Geir
    Sandberg, Sverre
    CLINICAL CHEMISTRY, 2006, 52 (07) : 1311 - 1317
  • [37] Evaluation of published vancomycin pharmacokinetic infant models with therapeutic drug monitoring data from a pediatric population
    Comisar, Craig M.
    Jayaraman, Bhuvana
    Barrett, Jeffrey S.
    JOURNAL OF CLINICAL PHARMACOLOGY, 2008, 48 (09): : 1123 - 1123
  • [38] Vancomycin therapeutic drug monitoring service quality indices and clinical effectiveness outcomes: A retrospective cohort and clinical audit
    Al-Sulaiti, Fatima K. K.
    Nader, Ahmed
    El-Mekaty, Eman
    Elewa, Hazem
    Al-Badriyeh, Daoud
    El-Zubair, Ahmed
    Saad, Mohamed O. O.
    Awaisu, Ahmed
    JOURNAL OF THE AMERICAN COLLEGE OF CLINICAL PHARMACY, 2020, 3 (04): : 778 - 785
  • [39] External quality assessment scheme for HbA1c assays in Thailand: A 5-year experience
    Suparak, Supaporn
    Pratumvinit, Busadee
    Ngueanchanthong, Kanokwan
    Unpol, Petai
    Thanomsakyuth, Ariya
    Setthaudom, Chavachol
    Kunakorn, Mongkol
    Rojanawiwat, Archawin
    Uppapong, Ballang
    PRACTICAL LABORATORY MEDICINE, 2022, 31
  • [40] Towards precision medicine: Therapeutic drug monitoring-guided dosing of vancomycin and β-lactam antibiotics to maximize effectiveness and minimize toxicity
    Cusumano, Jaclyn A.
    Klinker, Kenneth P.
    Huttner, Angela
    Luther, Megan K.
    Roberts, Jason A.
    LaPlante, Kerry L.
    AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2020, 77 (14) : 1104 - 1112