Accuracy of the Minicardio system for heart rate variability analysis compared to ECG

被引:0
|
作者
Cassirame, J. [1 ]
Stuckey, M. I. [2 ,3 ]
Sheppard, F.
Tordi, N. [4 ,5 ]
机构
[1] Univ Franche Comte, Univ Hosp Besancon, EA Physiopathol Cardiovasc & Prevent 3920, F-25030 Besancon, France
[2] Univ Western Ontario, Dept Kinesiol, London, ON, Canada
[3] Lawson Hlth Res Inst, London, ON, Canada
[4] Univ Hosp Besancon, INSERM CIT808, Ctr Invest Clin Innovat Technol, Besancon, France
[5] Univ Franche Comte, EA Fonct & Dysfonct Epitheliale 4267, F-25030 Besancon, France
来源
关键词
Heart rate; Heart function tests; Resistance training; R-R INTERVALS; POLAR S810; VENTILATORY THRESHOLDS; RATE DYNAMICS; RATE MONITOR; VALIDITY; TIME;
D O I
暂无
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
Aim. The purpose of this investigation was twofold: firstly, to assess the accuracy of heart rate (HR) measurements from Minicardio compared to a standard electrocardiogram (ECG) and secondly, to compare the heart rate variability analysis carried out for each set of data. Methods. Fifteen young participants volunteered for this study and were instrumented for the simultaneous recording of HR via ECG and Minicardio. HR was recorded continuously during 10 minutes of supine rest and 2 consecutive exercise bouts (target HR at 130 bpm and 150 bpm). R-R intervals were determined in order to compare the number and type of artifacts in each data set. The device were then compared by determining the type and number of artifacts, correlations between HRV analysis parameters and Bland-Altman limits of agreement. Results. No artifact was produced from ECG. With Minicardio, no artifact was produced during supine rest, during exercise there were 2 type 1 (long R-R interval) and 7 type 2 artifacts (short R-R interval). For time domain analysis, Pearson coefficients were 1 for SDNN and mean R-R interval and 0.99 for RMSSD and pNN50. For frequency domain analysis, Pearson coefficients were 0.99 for high frequency and low frequency power, and 0.99 for Poincare plots SD1 and SD2. Limits of agreement from Bland-Altman analysis showed an excellent agreement between all devices Conclusion. The similarity between ECG and Minicardio encourages us to use Minicardio as a portable HR recorder for subjects at rest and during moderate cycling exercise.
引用
下载
收藏
页码:248 / 254
页数:7
相关论文
共 50 条
  • [11] The corner frequencies of the ECG amplifier for heart rate variability analysis
    Hejjel, L
    Kellenyi, L
    PHYSIOLOGICAL MEASUREMENT, 2005, 26 (01) : 39 - 47
  • [12] Comparative Analysis of the Equivital EQ02 Lifemonitor with Holter Ambulatory ECG Device for Continuous Measurement of ECG, Heart Rate, and Heart Rate Variability: A Validation Study for Precision and Accuracy
    Akintola, Abimbola A.
    van de Pol, Vera
    Bimmel, Daniel
    Maan, Arie C.
    van Heemst, Diana
    FRONTIERS IN PHYSIOLOGY, 2016, 7
  • [13] Ambulatory ECG and analysis of heart rate variability in Parkinson's disease
    Haapaniemi, TH
    Pursiainen, V
    Korpelainen, JT
    Huikuri, HV
    Sotaniemi, KA
    Myllylä, VV
    JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 2001, 70 (03): : 305 - 310
  • [14] Influence of Undetected ECG R waves on the Heart Rate Variability Analysis
    Naka, H.
    Tohara, T.
    Inoue, K.
    2008 PROCEEDINGS OF SICE ANNUAL CONFERENCE, VOLS 1-7, 2008, : 3068 - 3072
  • [15] Comparative Heart Rate Variability Analysis of ECG, Holter and PPG Signals
    Georgieva-Tsaneva, Galya N.
    Gospodinova, Evgeniya
    INTERNATIONAL JOURNAL OF ADVANCED COMPUTER SCIENCE AND APPLICATIONS, 2021, 12 (12) : 464 - 470
  • [16] High resolution ECG and heart rate variability
    Voss, A
    Wessel, N
    Dietz, R
    PROCEEDINGS OF THE 18TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 18, PTS 1-5, 1997, 18 : 1622 - 1623
  • [17] Evaluation of Effect of Music on Human Nervous System by Heart Rate Variability Analysis Using ECG Sensor
    Wu, Min-Hao
    Chang, Ting-Cheng
    SENSORS AND MATERIALS, 2021, 33 (02) : 739 - 753
  • [18] System for Heart Rate Variability Analysis in Athletes
    Petry, D.
    Marques, J. L. B.
    5TH LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2011): SUSTAINABLE TECHNOLOGIES FOR THE HEALTH OF ALL, PTS 1 AND 2, 2013, 33 (1-2): : 311 - 314
  • [19] Accuracy of pulse rate variability parameters obtained from finger plethysmogram: A comparison with heart rate variability parameters obtained from ECG
    Kageyama, T
    Kabuto, M
    Kaneko, T
    Nishikido, N
    JOURNAL OF OCCUPATIONAL HEALTH, 1997, 39 (02) : 154 - 155
  • [20] Motion Artifact Removal in Ambulatory ECG Signal for Heart Rate Variability Analysis
    Berwal, Deepak
    Vandana, C. R.
    Dewan, Sourya
    Jiji, C., V
    Baghini, Maryam Shojaei
    IEEE SENSORS JOURNAL, 2019, 19 (24) : 12432 - 12442