Accuracy of the Minicardio system for heart rate variability analysis compared to ECG

被引:0
|
作者
Cassirame, J. [1 ]
Stuckey, M. I. [2 ,3 ]
Sheppard, F.
Tordi, N. [4 ,5 ]
机构
[1] Univ Franche Comte, Univ Hosp Besancon, EA Physiopathol Cardiovasc & Prevent 3920, F-25030 Besancon, France
[2] Univ Western Ontario, Dept Kinesiol, London, ON, Canada
[3] Lawson Hlth Res Inst, London, ON, Canada
[4] Univ Hosp Besancon, INSERM CIT808, Ctr Invest Clin Innovat Technol, Besancon, France
[5] Univ Franche Comte, EA Fonct & Dysfonct Epitheliale 4267, F-25030 Besancon, France
来源
关键词
Heart rate; Heart function tests; Resistance training; R-R INTERVALS; POLAR S810; VENTILATORY THRESHOLDS; RATE DYNAMICS; RATE MONITOR; VALIDITY; TIME;
D O I
暂无
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
Aim. The purpose of this investigation was twofold: firstly, to assess the accuracy of heart rate (HR) measurements from Minicardio compared to a standard electrocardiogram (ECG) and secondly, to compare the heart rate variability analysis carried out for each set of data. Methods. Fifteen young participants volunteered for this study and were instrumented for the simultaneous recording of HR via ECG and Minicardio. HR was recorded continuously during 10 minutes of supine rest and 2 consecutive exercise bouts (target HR at 130 bpm and 150 bpm). R-R intervals were determined in order to compare the number and type of artifacts in each data set. The device were then compared by determining the type and number of artifacts, correlations between HRV analysis parameters and Bland-Altman limits of agreement. Results. No artifact was produced from ECG. With Minicardio, no artifact was produced during supine rest, during exercise there were 2 type 1 (long R-R interval) and 7 type 2 artifacts (short R-R interval). For time domain analysis, Pearson coefficients were 1 for SDNN and mean R-R interval and 0.99 for RMSSD and pNN50. For frequency domain analysis, Pearson coefficients were 0.99 for high frequency and low frequency power, and 0.99 for Poincare plots SD1 and SD2. Limits of agreement from Bland-Altman analysis showed an excellent agreement between all devices Conclusion. The similarity between ECG and Minicardio encourages us to use Minicardio as a portable HR recorder for subjects at rest and during moderate cycling exercise.
引用
下载
收藏
页码:248 / 254
页数:7
相关论文
共 50 条
  • [1] ECG monitoring system for heart rate variability analysis
    University Politehnica Bucharest, Faculty of Electrical Engineering, 313 Splaiul Independentei, Romania
    IMEKO TC Symp. Novelties Electr. Meas. Instrum.,
  • [2] System to ECG Signals Variability Analysis: Heart Rate Variability and QT Interval Variability
    Petry, D.
    Palodeto, V.
    Suzuki, D. O. H.
    Marques, J. L. B.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 1160 - +
  • [3] ECG system with heart rate variability application
    Naiyanetr, P
    Charoentong, P
    Charoensuk, W
    TENCON 2004 - 2004 IEEE REGION 10 CONFERENCE, VOLS A-D, PROCEEDINGS: ANALOG AND DIGITAL TECHNIQUES IN ELECTRICAL ENGINEERING, 2004, : B435 - B437
  • [4] ECG MEASUREMENT AND ACQUISITION SYSTEM FOR DYNAMIC ANALYSIS OF HEART RATE VARIABILITY PARAMETERS
    Boev, Anatolie
    Stanciu, Marcel
    Pantelimon, Brandusa
    UNIVERSITY POLITEHNICA OF BUCHAREST SCIENTIFIC BULLETIN SERIES C-ELECTRICAL ENGINEERING AND COMPUTER SCIENCE, 2011, 73 (01): : 191 - 198
  • [5] Influence of ECG sampling rate in fetal heart rate variability analysis
    De Jonckheere, J.
    Garabedian, C.
    Charlier, P.
    Champion, C.
    Servan-Schreiber, E.
    Storme, L.
    Debarge, V.
    Jeanne, M.
    Logier, R.
    2017 39TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2017, : 2027 - 2030
  • [6] Reliability and accuracy of heart rate variability metrics versus ECG segment duration
    James McNames
    Mateo Aboy
    Medical and Biological Engineering and Computing, 2006, 44 : 747 - 756
  • [7] Extraction of Fetal ECG from Abdominal ECG and Heart Rate Variability Analysis
    Lamesgin, Gizeaddis
    Kassaw, Yonas
    Assefa, Dawit
    AFRO-EUROPEAN CONFERENCE FOR INDUSTRIAL ADVANCEMENT, AECIA 2014, 2015, 334 : 65 - 76
  • [8] Reliability and accuracy of heart rate variability metrics versus ECG segment duration
    McNames, James
    Aboy, Mateo
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2006, 44 (09) : 747 - 756
  • [9] Reliability and accuracy of heart rate variability metrics versus ECG segment duration
    McNames, J
    Thong, T
    Goldstein, B
    PROCEEDINGS OF THE 25TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: A NEW BEGINNING FOR HUMAN HEALTH, 2003, 25 : 212 - 215
  • [10] An ECG Arrhythmia Classification and Heart Rate Variability Analysis System based on Android Platform
    Xue, Shouhai
    Chen, Xianxiang
    Fang, Zhen
    Xia, Shanhong
    2015 2nd International Symposium on Future Information and Communication Technologies for Ubiquitous HealthCare (Ubi-Health Tech), 2015,