Dosimetric Accuracy of an HDR Brachytherapy Treatment Planning System for Different Irradiation Lengths with Monte Carlo Simulation

被引:0
|
作者
Acun Bucht, Hediye [1 ]
Bozkurt, Ahmet [2 ]
Kemikler, Gonul [3 ]
机构
[1] Harran Univ, Dept Biophys, Fac Med, Sanliurfa, Turkey
[2] Akdeniz Univ, Dept Biomed Engn, Fac Engn, Antalya, Turkey
[3] Istanbul Univ, Dept Med Phys, Inst Oncol, Istanbul, Turkey
来源
关键词
Dose rate profile; HDR brachytherapy; Monte Carlo; TPS; DOSE-RATE BRACHYTHERAPY; IR-192; PLUS; 12I;
D O I
10.5505/tjo.2019.1876
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
OBJECTIVE The purpose of the present study was to verify calculated dose rate profiles of Ir-192 for four different irradiation lengths and various numbers of dwell positions by using treatment planning system (TPS) calculations and Monte Carlo (MC) simulations. METHODS Dose rate profiles per air-kerma strength in mu Gy m(2)/h (abbreviated as U) were calculated on transverse axis for irradiation lengths of 2, 3, 4, and 5 cm by Nucletron Oncentra TPS. The same irradiation scenarios were simulated with Monte Carlo N-Particle (MCNP) Transport Code. The MC and TPS calculated dose rate profiles were compared for all settings. RESULTS The difference between the calculated dose rate profiles by MC and TPS was found within 2% at an interval of 2 mm-8 cm away from the source for all irradiation lengths. The largest discrepancy of 5.2% was computed at a distance of 10 cm from the source for 5 cm irradiation length. CONCLUSION There is a good consistency between the calculated dose rate profiles by MCNP and Oncentra TPS. The deviation between the calculated dose rate values slightly increases as the distance from the source center increases >5 cm from the source, and its quantity depends on the number of dwell positions.
引用
收藏
页码:36 / 44
页数:9
相关论文
共 50 条
  • [1] Monte Carlo simulation of HDR Brachytherapy dosimetric parameters in different mediums
    Shukor, Nor Shazleen Ab
    Musarudin, Marianie
    Abdullah, Reduan
    Aziz, Mohd Zahri Abd
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2022, 200
  • [2] Monte Carlo simulation and dosimetry measurements of an experimental approach for in vitro HDR brachytherapy irradiation
    Geraldo, Jony M.
    Andrade, Lidia M.
    Nogueira, Luciana B.
    Grundmann, Cassia T.
    Fonseca, Telma C. F.
    Gomes, Bruno G. S.
    Leite, Miguel T. T.
    Mafra, Arnoldo
    Goncalves, Kennedy B.
    Furtado, Clascidia A.
    Ladeira, Luiz O.
    [J]. APPLIED RADIATION AND ISOTOPES, 2021, 172
  • [3] PATIENT-SPECIFIC MONTE CARLO SIMULATION FOR THE DOSIMETRIC VALIDATION OF AN ADVANCED 1921R BRACHYTHERAPY TREATMENT PLANNING SYSTEM
    Zourari, K.
    Pantelis, E.
    Sakelliou, L.
    Georgiou, E.
    Karaiskos, P.
    Papagiannis, P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S491 - S491
  • [4] COMPARISON OF HDR BRACHYTHERAPY SOURCES IN MONTE CARLO INVERSE PLANNING
    D'Amours, M.
    Liu, D.
    Lessard, E.
    Pouliot, J.
    Verhaegen, F.
    Beaulieu, L.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 91 : S15 - S15
  • [5] Monte Carlo Dosimetric Study of SAVI HDR Applicator for Partial Breast Irradiation
    Zhang, Zhengdong
    Dou, Kai
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3756 - 3756
  • [6] Dosimetric evaluation of a monte carlo IMRT treatment planning system with the MIMiC
    Rassiah-Szegedi, P.
    Fuss, M.
    Sheikh-Bagheri, D.
    Szegedi, M.
    Papanikolaoul, N.
    Stathakis, S.
    Salter, B.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2441 - 2442
  • [7] An integrated Monte Carlo dosimetric verification system for radiotherapy treatment planning
    Yamamoto, T.
    Mizowaki, T.
    Miyabe, Y.
    Takegawa, H.
    Narita, Y.
    Yano, S.
    Nagata, Y.
    Teshima, T.
    Hiraoka, M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (07): : 1991 - 2008
  • [8] Dosimetric Evaluation of Treatment Planing System on High-Dose-Rate (HDR) Brachytherapy Using Monte Carlo Method and Radiochomic Film
    Pavan, G.
    Martins, M.
    Salmon, H.
    Anderson, E.
    Nardi, S.
    Fairbanks, L.
    Silva, D.
    Cursino, F.
    Junior, J.
    Silva, A.
    Santos, D.
    Werner, F.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S743 - S743
  • [9] Dosimetric comparison between three dimensional treatment planning system, Monte Carlo simulation and gel dosimetry in nasopharynx phantom for high dose rate brachytherapy
    Fazli, Zeynab
    Sadeghi, Mahdi
    Zahmatkesh, M. H.
    Mahdavi, Seied Rabei
    Tenreiro, Claudio
    [J]. JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2013, 9 (03) : 402 - 409
  • [10] A comparison of HDR near source dosimetry using a treatment planning system, Monte Carlo simulation; and radiochromic film
    Amoush, Ahmad
    Luckstead, Marcus
    Lamba, Michael
    Elson, Howard
    Kassing, William