Friction Measurement in the Biaxial Rocking Motion Hip Joint Simulator

被引:13
|
作者
Saikko, Vesa [1 ]
机构
[1] Aalto Univ, Dept Engn Design & Prod, FIN-02015 Helsinki, Finland
来源
基金
芬兰科学院;
关键词
biomechanics; bone; friction; prosthetics; wear; SLIDE TRACK ANALYSIS; WEAR; POLYETHYLENE; CUPS; COCR;
D O I
10.1115/1.2991121
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The established biaxial rocking motion (BRM) hip joint simulator was complemented by a novel friction measurement accessory. This simple and practical system, which was easily added to an existing BRM design, is described in detail and shown to perform well in long-term wear tests and in comparative tests of 24 h duration involving several different bearing couples. The system was based on the measurement of frictional torque about the leaning axis of the lower component, the femoral head. In the 28 mm CoCr-on-ultrahigh molecular weight polyethylene articulation with diluted calf serum lubricant at body temperature and 1 kN static load, the maximum value of frictional torque during a cycle was 1.2 N m on the average. The alternative system based on the measurement of torque about the vertical loading axis was shown to be less sensitive and highly insensitive when contact area was small, as was the case with alumina-on-alumina.
引用
收藏
页码:1 / 8
页数:8
相关论文
共 50 条
  • [41] On a hip joint simulator electric actuators integrated design and optimization
    Pellicciari, M
    Barbanti, G
    Andrisano, AO
    DESIGN 2002: Proceedings of the 7th International Design Conference, Vols 1 and 2, 2002, : 873 - 878
  • [42] A simulated synovial fluid for wear characterization of artificial hip joints by a hip joint simulator
    Takadama, H
    Mizuno, M
    BIOCERAMICS 18, PTS 1 AND 2, 2006, 309-311 : 1273 - 1276
  • [43] Motion study of the hip joint in extreme postures
    Caecilia Charbonnier
    Lazhari Assassi
    Pascal Volino
    Nadia Magnenat-Thalmann
    The Visual Computer, 2009, 25 : 873 - 882
  • [44] Studies of the Adhesive Component of Friction in Hip Joint Endoprostheses
    Yakupov R.R.
    Minasov B.S.
    Schuster L.S.
    Bilyalov A.R.
    Chertovskikh S.V.
    Nikitin V.V.
    Valeev M.M.
    Mavlyutov T.R.
    Minasov T.B.
    Emaev I.I.
    Biomedical Engineering, 2017, 51 (04) : 271 - 274
  • [45] The range of active motion at the hip joint of men
    Moore, JB
    Vaughn, JO
    JOURNAL OF BONE AND JOINT SURGERY, 1928, 10 : 248 - 257
  • [46] Accurate simulation of hip joint range of motion
    Kang, MJ
    Sadri, H
    Moccozet, L
    Magnenat-Thalmann, N
    Hoffmeyer, P
    CA 2002: PROCEEDINGS OF THE COMPUTER ANIMATION 2002, 2002, : 215 - 219
  • [47] Motion study of the hip joint in extreme postures
    Charbonnier, Caecilia
    Assassi, Lazhari
    Volino, Pascal
    Magnenat-Thalmann, Nadia
    VISUAL COMPUTER, 2009, 25 (09): : 873 - 882
  • [48] Comparison of distal radioulnar joint reconstructions using an active joint motion simulator
    Gofton, WT
    Gordon, KD
    Dunning, CE
    Johnson, JA
    King, GJW
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2005, 30A (04): : 733 - 742
  • [49] MULTILEVEL NUMERICAL MODEL OF HIP JOINT ACCOUNTING FOR FRICTION IN THE HIP RESURFACING ENDOPROSTHESIS
    Eremina, Galina M.
    Smolin, Alexey Yu
    FACTA UNIVERSITATIS-SERIES MECHANICAL ENGINEERING, 2019, 17 (01) : 29 - 38
  • [50] Joint Space Measurement as a Predictor for Joint Replacement for Hip Arthroscopy
    Zacher, J.
    ORTHOPADE, 2015, 44 (04): : 315 - 315