Radiation dose vs. image quality for low-dose CT protocols of the head for maxillofacial surgery and oral implant planning

被引:46
|
作者
Loubele, M
Jacobs, R
Maes, F
Schutyser, F
Debaveye, D
Bogaerts, R
Coudyzer, W
Vandermeulen, D
van Cleynenbreugel, J
Marchal, G
Suetens, P
机构
[1] Katholieke Univ Leuven, Univ Hosp Gasthuisberg, Med Image Compt ESAT, PSI, B-3000 Louvain, Belgium
[2] Katholieke Univ Leuven, Oral Imaging Ctr, Sch Dent Oral Pathol & Maxillofacial Surg, B-3000 Louvain, Belgium
[3] Medicim NV, B-9100 St Niklaas, Belgium
[4] Katholieke Univ Leuven, Univ Hosp Gasthuisberg, Unit Person Dosemetry Radiat Protect, B-3000 Louvain, Belgium
[5] Katholieke Univ Leuven, Univ Hosp Gasthuisberg, Dept Radiol, B-3000 Louvain, Belgium
关键词
D O I
10.1093/rpd/nci749
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The goal of this study was to determine the acquisition parameters for a low-dose multi-slice CT protocol and to compare the effective dose and the image quality of this low-dose protocol with the image quality of a clinical multi-slice CT protocol, routinely used for visualisation of the head. The low-dose protocol was derived from a clinical multi-slice CT protocol by lowering mA s and kV and increasing the pitch. The low-dose protocol yielded a dose reduction from 1.5 to 0.18 mSv for a multi-slice CT scan of the whole head, whereas noise in the low-dose CT images was increased. For bone segmentation, noise could be reduced by use of a non-linear edge preserving smoothing filter. Tests on ESP and skull phantom indicated that the accuracy of the measurements on low-dose CT is acceptable for image-based planning of maxillofacial and oral implant surgery, reducing the dose by a factor of 8.
引用
收藏
页码:211 / 216
页数:6
相关论文
共 50 条
  • [41] Low-dose CBCT protocols in implant dentistry: a systematic review
    Kaaber, Laurits
    Matzen, Louise Hauge
    Schropp, Lars
    Spin-Neto, Rubens
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2024, 138 (03): : 427 - 439
  • [42] Split vs. Single Bolus CT Urography: Comparison of Scan Time, Image Quality and Radiation Dose
    Morrison, Nicole
    Bryden, Sherrie
    Costa, Andreu F.
    TOMOGRAPHY, 2021, 7 (02) : 210 - 218
  • [43] Low-Dose Temporal Bone CT in Infants and Young Children: Effective Dose and Image Quality
    Nauer, C. B.
    Rieke, A.
    Zubler, C.
    Candreia, C.
    Arnold, A.
    Senn, P.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2011, 32 (08) : 1375 - 1380
  • [44] Low-Dose Chest CT Protocols for Imaging COVID-19 Pneumonia: Technique Parameters and Radiation Dose
    Suliman, Ibrahim I.
    Khouqeer, Ghada A.
    Ahmed, Nada A.
    Abuzaid, Mohamed M.
    Sulieman, Abdelmoneim
    LIFE-BASEL, 2023, 13 (04):
  • [45] Low-Dose High-Resolution Photon-Counting CT of the Lung: Radiation Dose and Image Quality in the Clinical Routine
    Woeltjen, Matthias Michael
    Niehoff, Julius Henning
    Michael, Arwed Elias
    Horstmeier, Sebastian
    Moenninghoff, Christoph
    Borggrefe, Jan
    Kroeger, Jan Robert
    DIAGNOSTICS, 2022, 12 (06)
  • [46] Image Quality Measurement for Low-Dose Human Lung CT Scans
    Bognar, Gergo
    2015 38TH INTERNATIONAL CONFERENCE ON TELECOMMUNICATIONS AND SIGNAL PROCESSING (TSP), 2015, : 587 - 591
  • [47] Effects of sparse sampling schemes on image quality in low-dose CT
    Abbas, Sajid
    Lee, Taewon
    Shin, Sukyoung
    Lee, Rena
    Cho, Seungryong
    MEDICAL PHYSICS, 2013, 40 (11)
  • [48] Phase II Trial of High-Dose Radiotherapy vs. Low-Dose Radiation, Demonstrating Low-Dose Mediated Immune-Cell Infiltration
    He, K.
    Patel, R. R.
    Barsoumian, H. B.
    Chang, J. Y.
    Tang, C.
    Comeaux, N. I.
    Ning, M. S.
    Gandhi, S.
    Davies, M.
    Glitza, I.
    Hong, D. S.
    Heymach, J.
    Nguyen, Q. N.
    Diab, A.
    Chun, S. G.
    Lee, P.
    Blumenschein, G. R.
    Fossella, F. V.
    Chen, D.
    Welsh, J. W.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : S118 - S118
  • [49] The effect of varying low-dose protocols on perceived image quality in multidetector CT in a rabbit model of acute appendicitis
    Marcus Hörmann
    Marcel O. Philipp
    Heidrun Eberl
    Ursula Windberger
    Wolfgang Matzek
    Cornelia Schäfer-Prokop
    Viktor M. Metz
    European Radiology, 2004, 14 : 1465 - 1471
  • [50] DOSE AND IMAGE QUALITY IN SOME LOW-DOSE MAMMOGRAPHIC SYSTEMS
    THOMASON, C
    WEBSTER, EW
    PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (04): : 787 - 787