Purpose. To compare anterior chamber depth (ACD) measurements using two non-contact optical devices, Pentacam and IOLMaster, and a contact device, ultrasonic A-scan in phakic and pseudophakic eyes. Methods. Ninety phakic and 94 pseudophakic eyes were enrolled in this prospective Study. The difference between ACD measurements by the three devices was analyzed using the repeated-measures analysis of variance, and agreement among the three measurements was investigated. The accuracy of detecting the anterior lens surface of the intraocular lens (IOL) with or without blue-blocker was also assessed. Results. In phakic eyes, the Pentacam measured the deepest ACD with the smallest standard deviation (3.26 +/- 0.41, 3.20 +/- 0.45, and 3.12 +/- 0.44 mm measured by Pentacam, IOLMaster, and A-scan, respectively, p < 0.001). In contrast, the Pentacam measurement of ACD had the largest standard deviation in pseudophakic eyes (4.05 +/- 0.58, 4.06 +/- 0.46, and 3.81 +/- 0.41 mm by Pentacam, IOL-Master, and A-scan, respectively, p < 0.001). The Pentacam instrument failed to correctly identify the anterior lens surface in 26.4% of IOLs without blue-blocker and in 58.5% of blue light-filtering IOLs (p = 0.016). Manual correction of the Pentacam image and subsequent ACD measurement improved the agreement between Pentacam and the other two devices. After manual adjustment, the average ACD value increased, and the standard deviation decreased significantly (4.34 +/- 0.28 mm). Conclusions. ACD measurements by Pentacam, IOLMaster, and A-scan were in better agreement in phakic eyes when compared with pseudophakic eyes. In pseudophakic eyes, the 95% limits of agreements between Pentacam and IOLMaster as well as Pentacam and A-scan were unsatisfactory. IOLs with blue-blocker might interfere with ACD measurements. However, this error can be corrected manually in the Pentacam examination.
引用
收藏
页码:1193 / 1200
页数:8
相关论文
共 50 条
  • [31] Depth of focus in pseudophakic eyes
    Nikolai M. Sergienko
    Yury N. Kondratenko
    Nikolai N. Tutchenko
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2008, 246
  • [32] Characteristics of rhegmatogenous retinal detachment in pseudophakic and phakic eyes
    Mahroo, O. A. R.
    Dybowski, R.
    Wong, R.
    Williamson, T. H.
    EYE, 2012, 26 (08) : 1114 - 1121
  • [33] Depth of focus in pseudophakic eyes
    Sergienko, Nikolai M.
    Kondratenko, Yury N.
    Tutchenko, Nikolai N.
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2008, 246 (11) : 1623 - 1627
  • [34] Anterior chamber depth measurement variation
    Seah, SKL
    Foster, PJ
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1997, 38 (04) : 812 - 812
  • [35] Fluorescence lifetime distribution in phakic and pseudophakic healthy eyes
    Dysli, Chantal
    Dysli, Muriel
    Wolf, Sebastian
    Zinkernagel, Martin
    PLOS ONE, 2023, 18 (01):
  • [36] Characteristics of rhegmatogenous retinal detachment in pseudophakic and phakic eyes
    O AR Mahroo
    R Dybowski
    R Wong
    T H Williamson
    Eye, 2012, 26 : 1114 - 1121
  • [37] Safety of DSAEK in pseudophakic eyes with anterior chamber lenses and Fuchs endothelial dystrophy
    Esquenazi, S.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2009, 93 (04) : 558 - 559
  • [38] Descemet membrane endothelial keratoplasty (DMEK) in phakic eyes with shallow anterior chamber
    Siggel, Robert
    Heindl, Ludwig M.
    Cursiefen, Claus
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2015, 253 (05) : 817 - 819
  • [39] Anterior chamber maintenance during drainage of a suprachoroidal hemorrhage in two phakic eyes
    Birt, CM
    Berger, AR
    OPHTHALMIC SURGERY AND LASERS, 1996, 27 (09): : 739 - 745
  • [40] Descemet membrane endothelial keratoplasty (DMEK) in phakic eyes with shallow anterior chamber
    Robert Siggel
    Ludwig M. Heindl
    Claus Cursiefen
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2015, 253 : 817 - 819