The stability of the femoral component of a minimal invasive total hip replacement system

被引:4
|
作者
Willems, M. M. M.
Kooloos, J.
Gibbons, P.
Minderhoud, N.
Weernink, T.
Verdonschot, N.
机构
[1] Radboud Univ Nijmegen Med Ctr, Orthopaed Res Lab, NL-6500 HB Nijmegen, Netherlands
[2] Radboud Univ Nijmegen Med Ctr, Dept Anat & Embryol, Nijmegen, Netherlands
[3] DePuy Int, Dept Res & Dev, Leeds, W Yorkshire, England
关键词
total hip replacement; mechanical testing; stability; minimal invasive surgery;
D O I
10.1243/09544119H08104
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, the initial stability of the femoral component of a minimal invasive total hip replacement was biomechanically evaluated during simulated normal walking and chair rising. A 20 mm diameter canal was created in the femoral necks of five fresh frozen human cadaver bones and the femoral heads were resected at the smallest cross-sectional area of the neck. The relatively short, polished, taper-shaped prostheses were cemented centrally in this canal according to a standardized procedure. A servohydraulic testing machine was used to apply dynamic loads to the prosthetic head. Radiostereophotogrammetric analysis was used to measure rotations and translations between the prosthesis and bone. In addition, the reconstructions were loaded until failure in a static, displacement-controlled test. During the dynamic experiments, the femoral necks did not fail and no macroscopical damage was detected. Maximal values were found for normal walking with a mean rotation of about 0.2 degrees and a mean translation of about 120 mu m. These motions stabilized during testing. The mean static failure load was 4714 N. The results obtained in this study are promising and warrant further development of this type of minimal invasive hip prosthesis.
引用
收藏
页码:465 / 472
页数:8
相关论文
共 50 条
  • [1] FEMORAL COMPONENT LOOSENING IN TOTAL HIP-REPLACEMENT
    PATERSON, JMH
    FULFORD, PC
    DENHAM, RA
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1987, 69 (01): : 155 - 155
  • [2] THE SURVIVAL OF THE CEMENTED FEMORAL COMPONENT OF A TOTAL HIP-REPLACEMENT
    RITTER, MA
    CAMPBELL, ED
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1989, (243) : 143 - 147
  • [3] Is the length of the femoral component important in primary total hip replacement?
    Feyen, H.
    Shimmin, A. J.
    [J]. BONE & JOINT JOURNAL, 2014, 96B (04): : 442 - 448
  • [4] MECHANICAL FAILURE IN THE FEMORAL COMPONENT IN TOTAL HIP-REPLACEMENT
    BANNISTER, G
    ORTH, MC
    [J]. ORTHOPEDIC CLINICS OF NORTH AMERICA, 1988, 19 (03) : 567 - 573
  • [5] CAUSES OF FRACTURES OF THE FEMORAL COMPONENT IN TOTAL HIP-REPLACEMENT
    GALANTE, JO
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1980, 62 (04): : 670 - 673
  • [6] OPTIMIZATION OF THE FEMORAL COMPONENT FIXATION IN TOTAL HIP-REPLACEMENT
    BOURNE, RB
    OH, I
    HARRIS, WH
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1980, 62 (01): : 133 - 134
  • [7] ASEPTIC LOOSENING OF THE FEMORAL COMPONENT IN CEMENTED TOTAL HIP-REPLACEMENT
    JONES, PR
    HUKINS, DWL
    PORTER, ML
    DAVIES, KE
    TIMPERLEY, AJ
    HARDINGE, K
    TAYLOR, CJ
    [J]. JOURNAL OF BIOMEDICAL ENGINEERING, 1992, 14 (05): : 379 - 384
  • [8] Bilateral femoral component breakage in total hip replacement: A case report
    Murray, P
    Rorabeck, CH
    [J]. CANADIAN JOURNAL OF SURGERY, 1998, 41 (03) : 236 - 238
  • [9] BENDING AND FRACTURE OF THE FEMORAL COMPONENT IN CEMENTED TOTAL HIP-REPLACEMENT
    JONES, PR
    HUKINS, DWL
    PORTER, ML
    DAVIES, KE
    HARDINGE, K
    TAYLOR, CJ
    [J]. JOURNAL OF BIOMEDICAL ENGINEERING, 1992, 14 (01): : 9 - 15
  • [10] Material selection in the design of the femoral component of cemented total hip replacement
    El-Sheikh, HF
    MacDonald, BJ
    Hashmi, MSJ
    [J]. JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2002, 122 (2-3) : 309 - 317