The stability of the femoral component of a minimal invasive total hip replacement system

被引:4
|
作者
Willems, M. M. M.
Kooloos, J.
Gibbons, P.
Minderhoud, N.
Weernink, T.
Verdonschot, N.
机构
[1] Radboud Univ Nijmegen Med Ctr, Orthopaed Res Lab, NL-6500 HB Nijmegen, Netherlands
[2] Radboud Univ Nijmegen Med Ctr, Dept Anat & Embryol, Nijmegen, Netherlands
[3] DePuy Int, Dept Res & Dev, Leeds, W Yorkshire, England
关键词
total hip replacement; mechanical testing; stability; minimal invasive surgery;
D O I
10.1243/09544119H08104
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, the initial stability of the femoral component of a minimal invasive total hip replacement was biomechanically evaluated during simulated normal walking and chair rising. A 20 mm diameter canal was created in the femoral necks of five fresh frozen human cadaver bones and the femoral heads were resected at the smallest cross-sectional area of the neck. The relatively short, polished, taper-shaped prostheses were cemented centrally in this canal according to a standardized procedure. A servohydraulic testing machine was used to apply dynamic loads to the prosthetic head. Radiostereophotogrammetric analysis was used to measure rotations and translations between the prosthesis and bone. In addition, the reconstructions were loaded until failure in a static, displacement-controlled test. During the dynamic experiments, the femoral necks did not fail and no macroscopical damage was detected. Maximal values were found for normal walking with a mean rotation of about 0.2 degrees and a mean translation of about 120 mu m. These motions stabilized during testing. The mean static failure load was 4714 N. The results obtained in this study are promising and warrant further development of this type of minimal invasive hip prosthesis.
引用
收藏
页码:465 / 472
页数:8
相关论文
共 50 条
  • [21] Minimal invasive surgery for hip replacement: A new technique using the NILNAV Hip System
    Sherry, E
    Egan, M
    Warnke, PH
    Henderson, A
    Eslick, GD
    [J]. ANZ JOURNAL OF SURGERY, 2003, 73 (03) : 157 - 161
  • [22] REVISION TOTAL HIP-REPLACEMENT USING THE S-ROM FEMORAL COMPONENT
    CHANDLER, HP
    AYRES, DK
    TAN, RC
    ANDERSON, LC
    VARMA, AK
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1995, (319) : 130 - 140
  • [23] FRACTURE OF THE FEMORAL COMPONENT AFTER TOTAL HIP-REPLACEMENT - AN ANALYSIS OF 58 CASES
    CHAO, EYS
    COVENTRY, MB
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1981, 63 (07): : 1078 - 1094
  • [24] Femoral component failure in canine cemented total hip replacement: a report of two cases
    Yates, G. D.
    Wasik, S. M.
    Edwards, G. A.
    [J]. AUSTRALIAN VETERINARY JOURNAL, 2010, 88 (06) : 225 - 230
  • [26] Minimally invasive total hip replacement
    d'Imporzano, M.
    Pierannunzii, L.
    [J]. JOURNAL OF ORTHOPAEDICS AND TRAUMATOLOGY, 2006, 7 (01) : 42 - 50
  • [27] Femoral cementing techniques in total hip replacement
    Rice, J
    Prenderville, T
    Murray, P
    McCormack, B
    Quinlan, W
    [J]. INTERNATIONAL ORTHOPAEDICS, 1998, 22 (05) : 308 - 311
  • [28] Femoral osteolysis following total hip replacement
    Dattani, R.
    [J]. POSTGRADUATE MEDICAL JOURNAL, 2007, 83 (979) : 312 - 316
  • [29] Femoral cementing techniques in total hip replacement
    J. Rice
    T. Prenderville
    P. Murray
    B. McCormack
    W. Quinlan
    [J]. International Orthopaedics, 1998, 22 : 308 - 311
  • [30] Femoral Component Revision of Total Hip Arthroplasty
    Brown, Jonathon M.
    Mistry, Jaydev B.
    Cherian, Jeffrey J.
    Elmallah, Randa K.
    Chughtai, Morad
    Harwin, Steven F.
    Mont, Michael A.
    [J]. ORTHOPEDICS, 2016, 39 (06) : E1129 - E1139