The stability of the femoral component of a minimal invasive total hip replacement system

被引:4
|
作者
Willems, M. M. M.
Kooloos, J.
Gibbons, P.
Minderhoud, N.
Weernink, T.
Verdonschot, N.
机构
[1] Radboud Univ Nijmegen Med Ctr, Orthopaed Res Lab, NL-6500 HB Nijmegen, Netherlands
[2] Radboud Univ Nijmegen Med Ctr, Dept Anat & Embryol, Nijmegen, Netherlands
[3] DePuy Int, Dept Res & Dev, Leeds, W Yorkshire, England
关键词
total hip replacement; mechanical testing; stability; minimal invasive surgery;
D O I
10.1243/09544119H08104
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, the initial stability of the femoral component of a minimal invasive total hip replacement was biomechanically evaluated during simulated normal walking and chair rising. A 20 mm diameter canal was created in the femoral necks of five fresh frozen human cadaver bones and the femoral heads were resected at the smallest cross-sectional area of the neck. The relatively short, polished, taper-shaped prostheses were cemented centrally in this canal according to a standardized procedure. A servohydraulic testing machine was used to apply dynamic loads to the prosthetic head. Radiostereophotogrammetric analysis was used to measure rotations and translations between the prosthesis and bone. In addition, the reconstructions were loaded until failure in a static, displacement-controlled test. During the dynamic experiments, the femoral necks did not fail and no macroscopical damage was detected. Maximal values were found for normal walking with a mean rotation of about 0.2 degrees and a mean translation of about 120 mu m. These motions stabilized during testing. The mean static failure load was 4714 N. The results obtained in this study are promising and warrant further development of this type of minimal invasive hip prosthesis.
引用
收藏
页码:465 / 472
页数:8
相关论文
共 50 条
  • [41] The effect of femoral component offset on prosthetic joint stability in total hip arthroplasty.
    Royhal, R
    [J]. JOURNAL OF INVESTIGATIVE MEDICINE, 2000, 48 (01) : 59A - 59A
  • [42] Fracture of a ceramic component in total hip replacement
    Whittingham-Jones, P.
    Mann, B.
    Coward, P.
    Hart, A. J.
    Skinner, J. A.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2012, 94B (04): : 570 - 573
  • [43] Effect of proximal femoral bone support on the fixation of a press-fit noncemented total hip replacement femoral component
    Sangiorgio, Sophia N.
    Ebramzadeh, Edward
    Borkowski, Sean L.
    Oakes, Daniel A.
    Reid, Jeremy J.
    Bengs, Benjamin C.
    [J]. JOURNAL OF APPLIED BIOMATERIALS & FUNCTIONAL MATERIALS, 2013, 11 (01): : 26 - 34
  • [44] Influence of Femoral Component Design on Proximal Femoral Bone Mass After Total Hip Replacement A Randomized Controlled Trial
    Slullitel, Pablo A.
    Mahatma, Mohit M.
    Farzi, Mohsen
    Grammatopoulos, George
    Wilkinson, J. Mark
    Beaule, Paul E.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2021, 103 (01): : 74 - 83
  • [45] Total hip replacement in avascular femoral head necrosis
    Betsch, M.
    Tingart, M.
    Driessen, A.
    Quack, V.
    Rath, B.
    [J]. ORTHOPADE, 2018, 47 (09): : 751 - 756
  • [46] FEMORAL FRACTURES IN CONJUNCTION WITH TOTAL HIP-REPLACEMENT
    TURNER, RH
    SCOTT, RD
    LEITZES, SM
    AUFRANC, OE
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1973, A 55 (08): : 1766 - 1766
  • [47] Osteonecrosis of the femoral head THE TOTAL HIP REPLACEMENT SOLUTION
    Issa, K.
    Pivec, R.
    Kapadia, B. H.
    Banerjee, S.
    Mont, M. A.
    [J]. BONE & JOINT JOURNAL, 2013, 95B (11): : 46 - 50
  • [48] FEMORAL FRACTURES IN CONJUNCTION WITH TOTAL HIP-REPLACEMENT
    SCOTT, RD
    TURNER, RH
    LEITZES, SM
    AUFRANC, OE
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1975, A 57 (04): : 494 - 501
  • [49] Femoral stem wear in cemented total hip replacement
    Zhang, H-Y
    Blunt, L.
    Jiang, X-Q
    Brown, L.
    Barrans, S.
    Zhao, Y.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2008, 222 (H5) : 583 - 592
  • [50] PROXIMAL FEMORAL REPLACEMENT TOTAL HIP-ARTHROPLASTY