BENDING AND FRACTURE OF THE FEMORAL COMPONENT IN CEMENTED TOTAL HIP-REPLACEMENT

被引:10
|
作者
JONES, PR
HUKINS, DWL
PORTER, ML
DAVIES, KE
HARDINGE, K
TAYLOR, CJ
机构
[1] ROYAL PRESTON HOSP,PRESTON PR2 4QF,LANCS,ENGLAND
[2] WRIGHTINGTON HOSP,CTR HIP SURG,WIGAN WN6 9EP,LANCS,ENGLAND
来源
JOURNAL OF BIOMEDICAL ENGINEERING | 1992年 / 14卷 / 01期
关键词
COMPUTER IMAGE ANALYSIS; STEM FRACTURE; TOTAL HIP REPLACEMENT;
D O I
10.1016/0141-5425(92)90029-K
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A computer method was used to make 41 measurements on the geometry of insertion of the femoral component in 200 Charnley total hip replacements. Surgery had been performed at least 12 years before, giving results which were classified as: success (90); fracture (56); or loose (54), according to rigid selection criteria. Fracture was associated with heavier patients in which there was poor proximal fixation of the femoral component but adequate distal fixation. Stems with a medial disposition proximally were more common in the fracture group than in the successful or loose groups. Sequential measurements of bending and subsequent fracture were made on the follow-up radiographs of 24 of the 200 cases (6 fracture and 18 successful). These measurements allowed bending to be detected at an earlier stage than by simple inspection of the radiographs.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 50 条