Multipolar radiofrequency ablation with internally cooled electrodes:: Experimental study in ex vivo bovine liver with mathernatic modeling

被引:64
|
作者
Clasen, S
Schmidt, D
Boss, A
Dietz, K
Kröber, SM
Claussen, CD
Pereira, PL
机构
[1] Univ Tubingen, Dept Diagnost Radiol, D-72076 Tubingen, Germany
[2] Univ Tubingen, Dept Med Biometry, Tubingen, Germany
[3] Univ Tubingen, Inst Pathol, D-7400 Tubingen, Germany
关键词
D O I
10.1148/radiol.2382050571
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the size and geometry of thermally induced coagulation by Using multipolar radiofrequency (RF) ablation and to determine a mathematic model to predict coagulation volume. Materials and Methods: Multipolar RF ablations (n = 80) were pet-formed in ex vivo bovine livers by using three internally cooled bipolar same shaft. Applicators were placed in a trianglilar array (spacing, 2-5 cm) and were activated in multipolar mode (power output, 75-225 W). The size and geometry of the coagulation zone, together with ablation time, were assessed. Mathematic functions were fitted. and the goodness or fit was assessed by using r(2). Results: Coagulation volome, short-axis diameter, and ablation time were dependent on power output an([ applicator distance. The maximum zone of coagulation (volume, 324 cm(3); short-axis diameter. 8.4 cm: ablation time, 193 min) was induced with a power output or 75 W at an applicator distance of 5 cm. Coagulation volume and ablation time decreased as power output increased. Power outputs Of 100-125 W at applicator distances of 2-4 cm led to a reasonable compromise between coagulation volume and ablation time. At 2 cm (100 W), coagulation volume, 3 short-axis diameter, and ablation time were 66 cm, 4.5 cm, and 19 min, respectively; at 3 cm (100 W), go cm(3),13 5.2 3 cm, and 22 min, respectively; at 4 cm (100 W) 132 cm 6.1 cm, and 27 min, respectively; at 2 cm (125 W), 56 cm 3 4.2 cm. and 9 min, respectively: at 3 cot (125 W), 73 cm 4.9 cm, and 12 min. respectively; and at. 4 cm (125 W) 103 cm(3), 5.5 cm, and 16 min, respectively. At applicator distances of 4 ern (>125 W) and 5 cm (>100 W), the zones of coagulation were not confluent. Coagulation on volume (r(2) = 0.80) and RF ablation time (r(2) = 0.93) were determined by using the mathematic model. Conclusions: Multipolar RF ablation with three bipolar applicators may produce large volumes of confluent coagulation ex vivo. A compromise is necessary between prolonged RF ablations at lower power outputs, which produce larger volumes of coagulation. and faster RF ablations at higher power outputs, which produce smaller volumes of coagulation. (c) RSNA, 2006.
引用
收藏
页码:881 / 890
页数:10
相关论文
共 50 条
  • [41] Radiofrequency Ablation of Spine An Experimental Study in an Ex Vivo Bovine and In Vivo Swine Model for Feasibility in Spine Tumor
    You, Nam Kyu
    Lee, Hye Young
    Shin, Dong Ah
    Choi, Gwi Hyun
    Yi, Seong
    Kim, Keung Nyun
    Yoon, Do Heum
    Park, Jon
    SPINE, 2013, 38 (18) : E1121 - E1127
  • [42] Overlapping ablation using a coaxial radiofrequency electrode and multiple cannulae system:: Experimental study in ex-vivo bovine liver
    Choi, D
    Lim, HK
    Kim, MJ
    Lee, J
    Kim, SK
    Kim, EY
    Kim, S
    Kim, SH
    KOREAN JOURNAL OF RADIOLOGY, 2003, 4 (02) : 117 - 123
  • [43] Bipolar radiofrequency ablation in ex vivo bovine liver with the open-perfused system versus the cooled-wet system
    Lee, JM
    Han, JK
    Kim, SH
    Sohn, KL
    Choi, SH
    Choi, BI
    EUROPEAN RADIOLOGY, 2005, 15 (04) : 759 - 764
  • [44] Bipolar radiofrequency ablation in ex vivo bovine liver with the open-perfused system versus the cooled-wet system
    Jeong Min Lee
    Joon Koo Han
    Se Hyung Kim
    Kyu Li Sohn
    Seung Hong Choi
    Byung Ihn Choi
    European Radiology, 2005, 15 : 759 - 764
  • [45] Comparison of Four Microwave Ablation Devices: An Experimental Study in ex Vivo Bovine Liver
    Hoffmann, Ruediger
    Rempp, Hansjoerg
    Erhard, Ludwig
    Blumenstock, Gunnar
    Pereira, Philippe L.
    Claussen, Claus D.
    Clasen, Stephan
    RADIOLOGY, 2013, 268 (01) : 89 - 97
  • [46] A newly developed perfused umbrella electrode for radiofrequency ablation:: An ex vivo evaluation study in bovine liver
    Bruners, Philipp
    Pfeffer, Jochen
    Kazim, Rana M.
    Guenther, Rolf W.
    Schmitz-Rode, Thomas
    Mahnken, Andreas H.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2007, 30 (05) : 992 - 998
  • [47] A Newly Developed Perfused Umbrella Electrode for Radiofrequency Ablation: An Ex Vivo Evaluation Study in Bovine Liver
    Philipp Bruners
    Jochen Pfeffer
    Rana M. Kazim
    Rolf W. Günther
    Thomas Schmitz-Rode
    Andreas H. Mahnken
    CardioVascular and Interventional Radiology, 2007, 30 : 992 - 998
  • [48] Optimization of wet radiofrequency ablation using a perfused-cooled electrode:: A comparative study in Ex vivo bovine livers
    Lee, JM
    Han, JK
    Kim, SH
    Lee, Y
    Shin, KS
    Han, CJ
    Lee, MW
    Choi, JI
    Choi, BI
    KOREAN JOURNAL OF RADIOLOGY, 2004, 5 (04) : 250 - 257
  • [49] Internally cooled antenna for microwave ablation: Results in ex vivo and in vivo porcine livers
    Wang, Yang
    Sun, Yuanyuan
    Feng, Lei
    Gao, Yongyan
    Ni, Xiaoxia
    Liang, Ping
    EUROPEAN JOURNAL OF RADIOLOGY, 2008, 67 (02) : 357 - 361
  • [50] The vascular cooling effect in hepatic multipolar radiofrequency ablation leads to incomplete ablation ex vivo
    Poch, Franz G. M.
    Rieder, Christian
    Ballhausen, Hanne
    Knappe, Verena
    Ritz, Joerg-Peter
    Gemeinhardt, Ole
    Kreis, Martin E.
    Lehmann, Kai S.
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2016, 32 (07) : 749 - 756