Radiofrequency Ablation of Spine An Experimental Study in an Ex Vivo Bovine and In Vivo Swine Model for Feasibility in Spine Tumor

被引:5
|
作者
You, Nam Kyu [1 ]
Lee, Hye Young [2 ]
Shin, Dong Ah [2 ]
Choi, Gwi Hyun [2 ]
Yi, Seong [2 ]
Kim, Keung Nyun [2 ]
Yoon, Do Heum [2 ]
Park, Jon [3 ]
机构
[1] Ajou Univ, Sch Med, Dept Neurosurg, Suwon 441749, South Korea
[2] Yonsei Univ, Coll Med, Dept Neurosurg, Spine & Spinal Cord Inst, Seoul 120752, South Korea
[3] Stanford Univ, Sch Med, Dept Neurosurg, Stanford, CA 94305 USA
关键词
ablation volume; bovine; experiment; ex vivo; internally cooled electrode; in vivo; metastasis; radiofrequency ablation; spine; spine tumor; swine; temperature; RENAL-CELL CARCINOMA; THERMAL ABLATION; VERTEBROPLASTY; CORD; METASTASES; EXPERIENCE; MANAGEMENT; ELECTRODE;
D O I
10.1097/BRS.0b013e31829c2e12
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. An experimental study of radiofrequency ablation (RFA) of spine in an ex vivo bovine and in vivo swine animal model. Objective. To study the feasibility of RFA for spine tumors close to the spinal cord, to examine the safety and efficacy of RFA, and to suggest quantitative guidelines for clinical application. Summary of Background Data. RFA has received increased attention as an effective and minimally invasive method for treating soft tissue tumors. However, there is currently only anecdotal evidence to support RFA of spinal tumors and only a few experimental studies have been conducted. Methods. We performed ex vivo experiments by producing 10 RFA zones in extracted bovine spines and an in vivo study by producing 8 RFA zones in a swine spine using internally cooled electrodes. The volume and diameter of ablation zones were evaluated and analyzed by the corresponding energy and ablation times. Results. In the ex vivo study, the average diameters of the ablation zones were 3.05 cm, 1.85 cm, and 1.26 cm, for the D1, D2, and D3 zones, respectively, and the average ablation volume was 4.19 cm(3). In the in vivo study, the average diameters were 2.51 cm, 2.05 cm, and 1.28 cm, respectively, and the ablation volume was 6.80 cm(3). The ablation zones demonstrated a positive correlation with ablation time, but the coefficients were 0.942 ex vivo and 0.257 in vivo. The temperature in the ex vivo study was inversely proportional to distance, with a maximal temperature of 63.7 degrees C at 10 mm; however, the maximum temperature was 38.2 degrees C in the in vivo study. Conclusion. This study demonstrated that sufficient RFA zone volume could be induced, which suggests that RFA is feasible and safe for application to human spinal tumors with predictability.
引用
收藏
页码:E1121 / E1127
页数:7
相关论文
共 50 条
  • [1] Ex Vivo Experimental Study of the Ablation Area of Bovine Liver Using STARmed Radiofrequency Ablation
    Ishikawa, Toru
    Hasegawa, Iori
    Hirosawa, Hiroshi
    Honmou, Tsubasa
    Sakai, Nobuyuki
    Igarashi, Takanori
    Yamazaki, Shun
    Kobayashi, Takamasa
    Sato, Toshifumi
    Iwanaga, Akito
    Sano, Tomoe
    Yokoyama, Junji
    Honma, Terasu
    IN VIVO, 2023, 37 (03): : 1047 - 1051
  • [2] Unidirectional Ablation Electrode to Minimize Thermal Injury During Radiofrequency Ablation: An Experimental Study in an Ex Vivo Bovine Liver Model
    Na, Dong Gyu
    Lee, Jeong Hyun
    Kim, Sun Mi
    Lim, Hyun Kyung
    Baek, Jung Hwan
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 22 (07) : 935 - 940
  • [3] Algorithm Optimization for Multitined Radiofrequency Ablation: Comparative Study in ex Vivo and in Vivo Bovine Liver
    Appelbaum, Liat
    Sosna, Jacob
    Pearson, Robert
    Perez, Sarah
    Nissenbaum, Yizhak
    Mertyna, Pawel
    Libson, Eugene
    Goldberg, S. Nahum
    RADIOLOGY, 2010, 254 (02) : 430 - 440
  • [4] Direct Current Combined With Bipolar Radiofrequency Ablation: An Ex Vivo Feasibility Study
    Tanaka, Toshihiro
    Penzkofer, Tobias
    Isfort, Peter
    Bruners, Philipp
    Disselhorst-Klug, Catherine
    Junker, Elmar
    Kichikawa, Kimihiko
    Schmitz-Rode, Thomas
    Mahnken, Andreas H.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 34 (03) : 631 - 636
  • [5] Direct Current Combined With Bipolar Radiofrequency Ablation: An Ex Vivo Feasibility Study
    Toshihiro Tanaka
    Tobias Penzkofer
    Peter Isfort
    Philipp Bruners
    Catherine Disselhorst-Klug
    Elmar Junker
    Kimihiko Kichikawa
    Thomas Schmitz-Rode
    Andreas H. Mahnken
    CardioVascular and Interventional Radiology, 2011, 34 : 631 - 636
  • [6] Microwave Ablation Compared with Radiofrequency Ablation for Breast Tissue in an Ex Vivo Bovine Udder Model
    Toshihiro Tanaka
    Saskia Westphal
    Peter Isfort
    Till Braunschweig
    Tobias Penzkofer
    Philipp Bruners
    Kimihiko Kichikawa
    Thomas Schmitz-Rode
    Andreas H. Mahnken
    CardioVascular and Interventional Radiology, 2012, 35 : 914 - 920
  • [7] Microwave Ablation Compared with Radiofrequency Ablation for Breast Tissue in an Ex Vivo Bovine Udder Model
    Tanaka, Toshihiro
    Westphal, Saskia
    Isfort, Peter
    Braunschweig, Till
    Penzkofer, Tobias
    Bruners, Philipp
    Kichikawa, Kimihiko
    Schmitz-Rode, Thomas
    Mahnken, Andreas H.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2012, 35 (04) : 914 - 920
  • [8] Multipolar radiofrequency ablation with internally cooled electrodes:: Experimental study in ex vivo bovine liver with mathernatic modeling
    Clasen, S
    Schmidt, D
    Boss, A
    Dietz, K
    Kröber, SM
    Claussen, CD
    Pereira, PL
    RADIOLOGY, 2006, 238 (03) : 881 - 890
  • [9] The distribution of heat in bone during radiofrequency ablation of an ex vivo bovine model of osteoid osteoma
    Greenberg, A.
    Weyel, T. Berenstein
    Sosna, J.
    Applbaum, J.
    Peyser, A.
    BONE & JOINT JOURNAL, 2014, 96B (05): : 677 - 683
  • [10] Geometry and temperature distribution during radiofrequency tissue ablation:: An experimental ex vivo model
    Lobik, L
    Leveillee, RJ
    Hoey, MF
    JOURNAL OF ENDOUROLOGY, 2005, 19 (02) : 242 - 247