Overlapping ablation using a coaxial radiofrequency electrode and multiple cannulae system:: Experimental study in ex-vivo bovine liver

被引:14
|
作者
Choi, D
Lim, HK
Kim, MJ
Lee, J
Kim, SK
Kim, EY
Kim, S
Kim, SH
机构
[1] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Radiol, Seoul 135710, South Korea
[2] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Ctr Imaging Sci, Seoul 135710, South Korea
[3] Kangbuk Samsung Hosp, Dept Radiol, Seoul, South Korea
关键词
D O I
10.3348/kjr.2003.4.2.117
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To assess the sizes and configurations of thermal zones after over lapping ablations using a coaxial radiofrequency (RF) electrode and multiple cannulae in ex-vivo bovine liver. Materials and Methods: For ablation procedures, a coaxial RF electrode and 2 introducer set was used. Employing real-time ultrasound guidance and overlapping techniques in explanted, fresh bovine liver, we created five kinds of thermal, zones with one (n=10), two (n=8), four (n=3), and six ablation spheres (n=3). 2 Following ablation, MR images were obtained and the dimensions of all thermal zones were measured on the longitudinal or transverse section of specimens. The shape of the composite ablation zones was evaluated using three-dimensional MR image reconstruction. Results: At gross pathologic examination of ten single-ablation zones (spheres), the long-axis (transverse) and short-axis lengths of zones ranged from 3.7 to 4.4 (mean, 4.1) cm and from 3.5 to 4.0 (mean, 3.7) cm, respectively. The long-axis (transverse) and short-axis lengths of double-ablation zones (cylinders) with 23% overlap ranged from 7.0 to 7.7 (mean, 7.3) cm and from 3.0 to 3.9 (mean, 3.5) cm, respectively; those with 58% overlap ranged from 6.0 to 6.4 (mean, 6.2) cm and from 3.8 to 4.6 (mean, 4.3) cm, respectively. The long-axis (diagonal) and short-axis lengths on a transverse section of four-ablation zones (cakes) ranged from 8.5 to 9.7 (mean, 9.1) cm and from 3.0 to 4.1 (mean, 3.7) cm, respectively. Gross pathologic examination of three composite six-ablation zones (spheres) showed that the long-axis (diagonal) and short-axis lengths of zones ranged from 9.0 to 9.9 (mean, 9.4) cm and from 6.8 to 7.5 (mean, 7.2) cm, respectively. T2-weighted MR images depicted low-signal thermal zones containing multiple curvilinear and spotty regions, of hyperintensity. Conclusion: Using a coaxial RF electrode and multiple cannulae, together with ultrasound guidance and precise overlapping ablation techniques, We successfully created predictable thermal zones in ex-vivo bovine liver.
引用
收藏
页码:117 / 123
页数:7
相关论文
共 50 条
  • [1] Ex Vivo Experimental Study of the Ablation Area of Bovine Liver Using STARmed Radiofrequency Ablation
    Ishikawa, Toru
    Hasegawa, Iori
    Hirosawa, Hiroshi
    Honmou, Tsubasa
    Sakai, Nobuyuki
    Igarashi, Takanori
    Yamazaki, Shun
    Kobayashi, Takamasa
    Sato, Toshifumi
    Iwanaga, Akito
    Sano, Tomoe
    Yokoyama, Junji
    Honma, Terasu
    IN VIVO, 2023, 37 (03): : 1047 - 1051
  • [2] Radiofrequency ablation with a new perfused-cooled electrode using a single pump:: An experimental study in Ex vivo bovine liver
    Kim, SK
    Se, JW
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2005, 28 (06) : 779 - 788
  • [3] Radiofrequency Ablation with a New Perfused-Cooled Electrode Using a Single Pump: An Experimental Study in Ex Vivo Bovine Liver
    Seung Kwon Kim
    Jung Wook Seo
    CardioVascular and Interventional Radiology, 2005, 28 : 779 - 788
  • [4] Unidirectional Ablation Electrode to Minimize Thermal Injury During Radiofrequency Ablation: An Experimental Study in an Ex Vivo Bovine Liver Model
    Na, Dong Gyu
    Lee, Jeong Hyun
    Kim, Sun Mi
    Lim, Hyun Kyung
    Baek, Jung Hwan
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 22 (07) : 935 - 940
  • [5] An ex-vivo experimental study on optimization of bipolar radiofrequency liver ablation using perfusion-cooled electrodes
    Lee, JM
    Han, JK
    Kim, SH
    Lee, JY
    Shin, KS
    Choi, BI
    ACTA RADIOLOGICA, 2005, 46 (05) : 443 - 451
  • [6] Experimental evaluation of the effective ablation zone on ex-vivo bovine liver samples
    Bottiglieri, Anna
    Elahi, Adnan
    Shahzad, Atif
    Donlon, Padraig T.
    Dennedy, Michael C.
    O'Halloran, Martin
    2018 EMF-MED 1ST WORLD CONFERENCE ON BIOMEDICAL APPLICATIONS OF ELECTROMAGNETIC FIELDS (EMF-MED 2018), 2018,
  • [7] A newly developed perfused umbrella electrode for radiofrequency ablation:: An ex vivo evaluation study in bovine liver
    Bruners, Philipp
    Pfeffer, Jochen
    Kazim, Rana M.
    Guenther, Rolf W.
    Schmitz-Rode, Thomas
    Mahnken, Andreas H.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2007, 30 (05) : 992 - 998
  • [8] A Newly Developed Perfused Umbrella Electrode for Radiofrequency Ablation: An Ex Vivo Evaluation Study in Bovine Liver
    Philipp Bruners
    Jochen Pfeffer
    Rana M. Kazim
    Rolf W. Günther
    Thomas Schmitz-Rode
    Andreas H. Mahnken
    CardioVascular and Interventional Radiology, 2007, 30 : 992 - 998
  • [9] Ex Vivo Experiment of Bovine Liver using Cool-Tip™ Radiofrequency Ablation System
    Mah, Y. H.
    Ng, K. H.
    Abdullah, B. J. J.
    Kwek, K. H.
    Wong, J. H. D.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 6, 2009, 25 : 209 - +
  • [10] Moving-Shot versus Fixed Electrode Techniques for Radiofrequency Ablation: Comparison in an Ex-Vivo Bovine Liver Tissue Model
    Ha, Eun Ju
    Baek, Jung Hwan
    Lee, Jeong Hyun
    KOREAN JOURNAL OF RADIOLOGY, 2014, 15 (06) : 836 - 843