Comparing Autorefractors for Measurement of Accommodation

被引:26
|
作者
Aldaba, Mikel [1 ]
Gomez-Lopez, Selena [1 ]
Vilaseca, Meritxell [1 ]
Pujol, Jaume [1 ]
Arjona, Montserrat [1 ]
机构
[1] Univ Politecn Cataluna, Ctr Sensors Instruments & Syst Dev CD6, Barcelona, Spain
关键词
accommodation; accommodative response; static; dynamic; WAM-5500; PowerRef-II; ECCENTRIC INFRARED PHOTOREFRACTOR; AGE-RELATED-CHANGES; CLINICAL-EVALUATION; REFRACTIVE ERROR; STATISTICAL-METHODS; GRAND SEIKO; POWERREFRACTOR; DYNAMICS; AGREEMENT; WAM-5500;
D O I
10.1097/OPX.0000000000000685
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To compare the static and dynamic accommodative responses measured with the WAM-5500 and the PowerRef-II autorefractors. Methods The dynamic and static monocular accommodative responses were measured with the WAM-5500 and the PowerRef-II instruments in 30 pre-presbyopic patients (23.66 [3.19] years). The spherical equivalent was measured at 0.00, 2.50, and 5.00 diopters (D) of accommodative stimulation for the static measurements. The subjective refraction was also determined. Dynamic accommodation was measured for abrupt changes of stimulus vergence of 2.00 D. Mean and peak velocities of accommodation and disaccommodation were evaluated. For the PowerRef-II, dynamic measurements were calculated for sampling frequencies of 5 and 25 Hz. Results For far distance static results, the differences between subjective and WAM-5500 measurements were 0.07 (+/- 0.21) D (p = 0.093) and those between subjective and PowerRef-II measurements were 0.70 (+/- 0.47) D (p = 0.001). The difference in the response measured with both instruments was 0.08 (+/- 0.32) D (p = 0.194) for 2.50 D and -0.32 (+/- 0.48) D (p = 0.001) for 5.00 D of stimulation. For the dynamic mode, the PowerRef-II at 25 Hz measured faster mean and peak velocities of accommodation and disaccommodation than the WAM-5500, with statistically significant (p < 0.05) differences of 0.68 (+/- 1.01), 0.67 (+/- 0.98), 1.26 (+/- 1.19), and 1.42 (+/- 1.53) D/s, respectively. With a sampling frequency of 5 Hz for the PowerRef-II, these differences, which were statistically significant (p < 0.05), were reduced to 0.52 (+/- 0.90), 0.49 (+/- 0.91), 0.83 (+/- 1.07), and 0.83 (+/- 1.31) D/s, respectively. Conclusions There is good agreement between subjective refraction and WAM-5500 measurements. In contrast, the PowerRef-II produced more hyperopic results. There were no differences among instruments at 2.50 D of static stimulation; however, differences were found at 5.00 D. In the dynamic measurements, the PowerRef-II measured faster velocities, partly attributed to the difference in the sampling frequency.
引用
收藏
页码:1003 / 1011
页数:9
相关论文
共 50 条
  • [21] Measurement of Gas-Surface Accommodation
    Trott, W. M.
    Rader, D. J.
    Castaneda, J. N.
    Torczynski, J. R.
    Gallis, M. A.
    RAREFIED GAS DYNAMICS, 2009, 1084 : 621 - 628
  • [22] OBJECTIVE, CONTINUOUS MEASUREMENT OF ACCOMMODATION IN THE EYE
    KRUEGER, H
    HESSEN, J
    BIOMEDIZINISCHE TECHNIK, 1982, 27 (06): : 142 - 147
  • [23] Measurement of accommodation using optical biometry
    Nurispahic, A.
    Kotliar, K.
    Lanzl, I.
    OPHTHALMOLOGE, 2008, 105 (04): : 369 - 375
  • [24] CRITICAL SUBJECTIVE MEASUREMENT OF AMPLITUDE OF ACCOMMODATION
    ATCHISON, DA
    CAPPER, EJ
    MCCABE, KL
    OPTOMETRY AND VISION SCIENCE, 1994, 71 (11) : 699 - 706
  • [25] OBJECTIVE AND SUBJECTIVE MEASUREMENT OF TONIC ACCOMMODATION
    BULLIMORE, MA
    GILMARTIN, B
    HOGAN, RE
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 1986, 6 (01) : 57 - 62
  • [26] MEASUREMENT OF TONIC ACCOMMODATION AS A FUNCTION OF WAVELENGTH
    DECHRISTE, S
    SIMONET, P
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 1995, 18 (6-7): : 439 - 446
  • [27] The measurement of accommodation using wavefront technology
    Raymond, WR
    Otto, CS
    Minarcik, JR
    Schallhorn, SC
    Parmley, VC
    George, RK
    Mazzoli, RA
    Apodaca, AJ
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2002, 43 : U36 - U36
  • [28] Measurement of accommodation related MEG responses
    Takeda, T
    Hashimoto, K
    Kumagai, T
    Endo, H
    BIOMAG 96: PROCEEDINGS OF THE TENTH INTERNATIONAL CONFERENCE ON BIOMAGNETISM, VOLS I & II, 2000, : 881 - 884
  • [29] Measurement of amplitude of accommodation in young persons
    Sergienko, Nikolai M.
    Nikonenko, Dmitry P.
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2015, 98 (04) : 359 - 361
  • [30] SUBJECTIVE MEASUREMENT OF ACCOMMODATION WITH LASER LIGHT
    HENNESSY, RT
    LEIBOWITZ, H
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1970, 60 (12) : 1700 - +