Comparison of two detector systems for cone beam CT small animal imaging - a preliminary study

被引:0
|
作者
Meng, Yang [1 ]
Shaw, Chris C. [1 ]
Liu, Xinming [1 ]
Altunbas, Mustafa C. [1 ]
Wang, Tianpeng [1 ]
Chen, Lingyun [1 ]
Tu, Shu-Ju [1 ]
Kappadath, S. Cheenu [1 ]
Lai, Chao-Jen [1 ]
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Imaging Phys, Digital Imaging Res Lab, Houston, TX 77030 USA
关键词
cone beam CT; CCD; flat panel; small animal imaging;
D O I
10.1117/12.656690
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose: To compare two detector systems - one based on the charge-coupled device (CCD) and image amplifier, the other based on a-Si/CsI flat panel, for cone beam computed-tomography (CT) imaging of small animals. A high resolution, high framing rate detector system for the cone beam CT imaging of small animals was developed. The system consists of a 2048x3072x12 bit CCD optically coupled to an image amplifier and an x-ray phosphor screen. The CCD has an intrinsic pixel size of 12 gin but the effective pixel size can be adjusted through the magnification adjustment of the optical coupling systems. The system is used in conjunction with an x-ray source and a rotating stage for holding and rotating the scanned object in the cone beam CT imaging experiments. The advantages of the system include but are not limited to the ability to adjust the effective pixel size and to achieve extremely high spatial resolution and temporal resolution. However, the need to use optical coupling compromises the detective quanta efficiency (DQE) of the system. In this paper, the imaging characteristics of the system were presented and compared with those of an aSi/CsI flat-panel detector system.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Comparison of Two Imaging Setups for Small Animal Cone Beam CT
    Yang, Y.
    Armour, M.
    Siewerdsen, J.
    Iordachita, I.
    Wong, J.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [2] Comparison of two cone-beam CT scanners dedicated to small animal research
    Cody, D
    Ross, W
    MEDICAL PHYSICS, 2002, 29 (06) : 1322 - 1322
  • [3] Dual-energy cone-beam Micro-CT for animal imaging: Preliminary study
    Cho, Seungryong
    Sidky, Emil
    Bian, Junguo
    Pan, Xiaochuan
    MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [4] Flat panel detector-based cone beam volume CT breast imaging: preliminary phantom study
    Ning, RL
    Chen, BA
    Conover, D
    McHugh, L
    Cullinan, J
    Yu, RF
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 601 - 610
  • [5] Flat panel detector-based cone beam CT imaging: Preliminary system evaluation
    Ning, RL
    Conover, D
    Yu, Y
    Lu, XH
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 616 - 620
  • [6] Performance Comparison of Two Different Cone-Beam CT Systems
    Chan, P.
    Li, S.
    Chan, M.
    MEDICAL PHYSICS, 2008, 35 (06) : 2649 - +
  • [7] Novel Detector Configurations in Cone-Beam CT Systems: A Simulation Study
    Karali, Evangelia
    Michail, Christos
    Fountos, George
    Kalyvas, Nektarios
    Valais, Ioannis
    CRYSTALS, 2024, 14 (05)
  • [8] Flat panel detector-based cone beam CT lung imaging: preliminary system evaluation
    Ning, R
    Conover, D
    Yu, Y
    Cai, WX
    Lu, XH
    MEDICAL IMAGING 2005: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2005, 5745 : 147 - 158
  • [9] Preliminary system characterization of flat panel detector-based cone beam CT for breast imaging
    Ning, RL
    Yu, Y
    Conover, D
    Lu, XH
    He, HG
    Chen, ZK
    Schiffhauer, L
    Cullinan, J
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 292 - 303
  • [10] Cone-beam CT breast imaging with a flat panel detector - a simulation study
    Chen, LY
    Shaw, CC
    Tu, SJ
    Altunbas, M
    Wang, T
    Lai, CJ
    Liu, XM
    Kappadath, SC
    MEDICAL IMAGING 2005: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2005, 5745 : 943 - 951