Comparison of two detector systems for cone beam CT small animal imaging - a preliminary study

被引:0
|
作者
Meng, Yang [1 ]
Shaw, Chris C. [1 ]
Liu, Xinming [1 ]
Altunbas, Mustafa C. [1 ]
Wang, Tianpeng [1 ]
Chen, Lingyun [1 ]
Tu, Shu-Ju [1 ]
Kappadath, S. Cheenu [1 ]
Lai, Chao-Jen [1 ]
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Imaging Phys, Digital Imaging Res Lab, Houston, TX 77030 USA
关键词
cone beam CT; CCD; flat panel; small animal imaging;
D O I
10.1117/12.656690
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose: To compare two detector systems - one based on the charge-coupled device (CCD) and image amplifier, the other based on a-Si/CsI flat panel, for cone beam computed-tomography (CT) imaging of small animals. A high resolution, high framing rate detector system for the cone beam CT imaging of small animals was developed. The system consists of a 2048x3072x12 bit CCD optically coupled to an image amplifier and an x-ray phosphor screen. The CCD has an intrinsic pixel size of 12 gin but the effective pixel size can be adjusted through the magnification adjustment of the optical coupling systems. The system is used in conjunction with an x-ray source and a rotating stage for holding and rotating the scanned object in the cone beam CT imaging experiments. The advantages of the system include but are not limited to the ability to adjust the effective pixel size and to achieve extremely high spatial resolution and temporal resolution. However, the need to use optical coupling compromises the detective quanta efficiency (DQE) of the system. In this paper, the imaging characteristics of the system were presented and compared with those of an aSi/CsI flat-panel detector system.
引用
收藏
页数:6
相关论文
共 50 条
  • [11] Comparison of effective dose for imaging of mandible between multi-detector CT and cone-beam CT
    Jeong, Dae-Kyo
    Lee, Sang-Chul
    Huh, Kyung-Hoe
    Yi, Won-Jin
    Heo, Min-Suk
    Lee, Sam-Sun
    Choi, Soon-Chul
    IMAGING SCIENCE IN DENTISTRY, 2012, 42 (02) : 65 - 70
  • [12] Circular Cone-beam Micro-CT for Small Animal Imaging with Truncated Data
    Cho, Seungryong
    Yu, Lifeng
    Pelizzari, Charles A.
    Pan, Xiaochuan
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2889 - 2891
  • [13] A preliminary study of intensity-weighted ROI imaging in cone-beam CT
    Cho, Seungryong
    Pearson, Erik
    Xia, Dan
    Han, Xiao
    Pelizzari, Charles A.
    Pan, Xiaochuan
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [14] Comparison measurements of DQE for two flat panel detectors:: Fluoroscopic detector vs. cone beam CT detector
    Benitez, Ricardo Betancourt
    Ning, Ruola
    Conover, David
    MEDICAL IMAGING 2006: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2006, 6142
  • [15] Integration of Bioluminescence Imaging and Cone-Beam CT for Image- Guided Small Animal Irradiation
    Weersink, R.
    Lindsay, P.
    Clarkson, R.
    Ansell, S.
    Wilson, G.
    Jaffray, D.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [16] Post acquisition small animal respiratory gated imaging using micro cone-beam CT
    Hu, JC
    Haworth, ST
    Molthen, RC
    Dawson, CA
    MEDICAL IMAGING 2004: PHYSIOLOGY, FUNCTION, AND STRUCTURE FROM MEDICAL IMAGES, 2004, 5 (23): : 350 - 360
  • [17] Feasibility of Applying SMEIR Method On Small Animal 4D Cone Beam CT Imaging
    Zhong, Y.
    Zhang, Y.
    Shao, Y.
    Wang, J.
    MEDICAL PHYSICS, 2016, 43 (06) : 3901 - 3901
  • [18] Estimating the magnitude of scatter in small animal cone-beam CT
    Chow, PL
    Vu, NT
    Chatziioannou, AF
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2752 - 2754
  • [19] Imaging properties of cone beam breast CT - effects of detector properties and imaging conditions
    Shaw, C
    Chen, L
    Altunbas, M
    Wang, T
    Lai, C
    Tu, S
    Liu, X
    Kappadath, S
    MEDICAL PHYSICS, 2005, 32 (06) : 2064 - 2064
  • [20] PIXSCAN: Pixel detector CT scanner for small animal imaging
    Delpierre, P.
    Basolo, S.
    Berar, J. F.
    Bonissent, A.
    Breugnon, P.
    Boudet, N.
    Caillot, B.
    Clemens, J. C.
    Debarbieux, F.
    Dinkespiler, B.
    Khouri, R.
    Koudobine, I.
    Matarazzo, V.
    Meessen, C.
    Menouni, M.
    Morel, C.
    Mouget, C.
    Pangaud, P.
    Peyrin, F.
    Rougon, G.
    Sappey-Marinier, D.
    Valton, S.
    Vigeolas, E.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2381 - 2385