Calibration method and photon flux influences tiled flat-panel photon counting detector image uniformity in computed tomography

被引:3
|
作者
Inkinen, S. I. [1 ]
Juntunen, M. A. K. [1 ,2 ,3 ,4 ]
Kotiaho, A. [1 ,2 ,3 ,4 ]
Nieminen, M. T. [1 ,2 ,3 ,4 ]
机构
[1] Univ Oulu, Res Unit Med Imaging Phys & Technol, Aapistie 5A, FI-90220 Oulu, Finland
[2] Univ Oulu, Med Res Ctr, Aapistie 5A, FI-90220 Oulu, Finland
[3] Oulu Univ Hosp, Aapistie 5A, FI-90220 Oulu, Finland
[4] Oulu Univ Hosp, Dept Diagnost Radiol, POB 50, FI-90029 Oulu, Finland
关键词
Computerized Tomography (CT) and Computed Radiography (CR); Data processing methods; Image reconstruction in medical imaging; X-ray detectors; CT; REDUCTION; ARTIFACTS;
D O I
10.1088/1748-0221/15/08/T08005
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The aim of this study is to compare how different calibration methods influence the image quality of photon-counting detector computed tomography (PCD-CT) at high and low photon fluxes. We investigate the performance of flat-field correction, signal-to-equivalent thickness calibration (STC), and polynomial correction (PC) methods using polymethyl methacrylate (PMMA) and iron as calibration materials. Two different cylindrical imaging phantoms containing contrast targets were scanned: an agar phantom and a phantom consisting of titanium hip implant embedded in agar. The scans were acquired using 120 kVp, and the energy thresholds of the PCD were set at 10 keV and 60 keV to obtain low energy (10-60 keV), high energy (60-120 keV) and total energy images (10-120 keV). Additionally, virtual monochromatic images (VMIs) with energies between 60-180 keV with 20 keV increments were generated from PC data. The reconstructions were made using filtered back projection, and image quality was assessed by evaluating image noise, contrast-to-noise ratio (CNR), and image uniformity. Overall, STC with PMMA as calibration material yielded the best image quality in terms of CNR and uniformity. Flat-field correction produced uniform reconstruction at low photon flux, but the performance degraded substantially at high flux. STC with iron as calibration material did not improve the reconstructions of the titanium hip implant. The beam hardening effects arising from metal were reduced when the VMI energy was increased while the CNR evaluated from agar phantom decreased with increasing energy of the VMI. Over the methods investigated, STC with PMMA was the most optimal calibration method for PCD-CT, yielding excellent image uniformity with both photon flux conditions.
引用
收藏
页数:19
相关论文
共 50 条
  • [1] Photon Counting Detector Computed Tomography
    Taguchi, Katsuyuki
    Ballabriga, Rafael
    Campbell, Michael
    Darambara, Dimitra G.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2022, 6 (01) : 1 - 4
  • [2] Dose and image quality of flat-panel detector volume computed tomography for sinus imaging
    Alspaugh, J.
    Christodoulou, E.
    Goodsitt, M.
    Stayman, J.
    MEDICAL PHYSICS, 2007, 34 (06) : 2634 - 2634
  • [3] Proposal for calibration of a single-photon counting detector without the need of input photon flux calibration
    Gerrits, Thomas
    JOURNAL OF OPTICS, 2016, 18 (05)
  • [4] Planar cone-beam computed tomography with a flat-panel detector
    Kim, S. H.
    Kim, D. W.
    Youn, H.
    Kim, D.
    Kam, S.
    Jeon, H.
    Kim, H. K.
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [5] Photon-counting detector computed tomography in cardiac imaging
    Sharma, Simran P.
    Lemmens, Marie-Julie D. K.
    Smulders, Martijn W.
    Budde, Ricardo P. J.
    Hirsch, Alexander
    Mihl, Casper
    NETHERLANDS HEART JOURNAL, 2024, 32 (11) : 405 - 416
  • [6] CT technology: photon-counting detector computed tomography
    Layer, Yannik Christian
    Kravchenko, Dmitrij
    Dell, Tatjana
    Kuetting, Daniel
    RADIOLOGIE, 2023, 63 (07): : 497 - 506
  • [8] Using the Medipix2 detector for photon counting computed tomography
    Niederloehner, Daniel
    Nachtrab, Frank
    Michel, Thilo
    Anton, Gisela
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2327 - 2331
  • [9] Potential benefits of photon counting detector computed tomography in pediatric imaging
    Horst, Kelly K.
    Yu, Lifeng
    Mccollough, Cynthia H.
    Esquivel, Andrea
    Thorne, Jamison E.
    Rajiah, Prabhakar Shanta
    Baffour, Francis
    Hull, Nathan C.
    Weber, Nikkole M.
    Thacker, Paul G.
    Thomas, Kristen B.
    Binkovitz, Larry A.
    Guerin, Julie B.
    Fletcher, Joel G.
    BRITISH JOURNAL OF RADIOLOGY, 2023, 96 (1152):
  • [10] Microstructural Imaging of Bone by Photon-Counting Detector Computed Tomography
    Burghardt, Andrew
    Korajac, Aida
    Strassl, Andreas
    Sadoughi, Saghi
    Kazakia, Galateia
    Patsch, Janina
    JOURNAL OF BONE AND MINERAL RESEARCH, 2023, 38 : 342 - 343