A Dual-Layer TOF-DOI Detector Block for Whole-Body PET

被引:6
|
作者
Liu, Shitao [1 ]
An, Shaohui [1 ]
Li, Hongdi [1 ]
Wang, Chao [1 ]
Ramirez, Rocio A. [1 ]
Zhang, Yuxuan [1 ]
Baghaei, Hossain [1 ]
Wong, Wai-Hoi [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Expt Diagnost Imaging, Houston, TX 77030 USA
基金
美国国家卫生研究院;
关键词
Depth of interaction (DOI); lutetium gadolinium oxyorthosilicate (LGSO); lutetium yttrium orthosilicate (LYSO); photomultiplier quadrant sharing (PQS); positron emission tomography (PET); time of flight (TOF); OF-FLIGHT PET; POSITRON-EMISSION-TOMOGRAPHY; TIME RESOLUTION; DEPTH; SCINTILLATION; PERFORMANCE; INFORMATION; MODULES; ARRAY;
D O I
10.1109/TNS.2012.2210053
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We developed a dual-layer high-resolution detector block for whole-body positron-emission tomography systems. The top layer is a 13 x 13 array of lutetium yttrium orthosilicate (LYSO) crystals (2.85 x 2.85 x 10 mm(3)) and the bottom layer a 13 x 13 array of lutetium gadolinium oxyorthosilicate (LGSO) crystals (2.85 x 2.85 x 10 mm(3)). The LGSO layer was coupled to four Hamamatsu R11194HA photomultiplier tubes (PMTs) by a photomultiplier quadrant-sharing configuration. This block has both time-of-flight and depth-of-interaction functions. Using a Na-22 point source, the performance of this block was evaluated by coinciding with a reference detector which consisted of a 4 x 4 x 10 mm(3) LYSO crystal coupled to a Hamamatsu R9779 PMT. A total of 4.7% of the counts were mixed among the two layers. The LYSO layer had 11.4% energy resolution and 329 ps coincidence time resolution against the reference detector. The LGSO layer had 14.8% energy resolution and 356 ps coincidence time resolution against the reference detector.
引用
收藏
页码:1805 / 1808
页数:4
相关论文
共 50 条
  • [41] Geometry Optimization of a Dual-Layer Offset Detector for Use in Simultaneous PET/MR Neuroimaging
    Teimoorisichani, Mohammadreza
    Goertzen, Andrew L.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2019, 3 (03) : 275 - 284
  • [42] Diagnostic accuracy of whole-body PET/MRI and whole-body PET/CT for TNM staging in oncology
    Philipp Heusch
    Felix Nensa
    Benedikt Schaarschmidt
    Rupika Sivanesapillai
    Karsten Beiderwellen
    Benedikt Gomez
    Jens Köhler
    Henning Reis
    Verena Ruhlmann
    Christian Buchbender
    European Journal of Nuclear Medicine and Molecular Imaging, 2015, 42 : 42 - 48
  • [43] Diagnostic accuracy of whole-body PET/MRI and whole-body PET/CT for TNM staging in oncology
    Heusch, Philipp
    Nensa, Felix
    Schaarschmidt, Benedikt
    Sivanesapillai, Rupika
    Beiderwellen, Karsten
    Gomez, Benedikt
    Koehler, Jens
    Reis, Henning
    Ruhlmann, Verena
    Buchbender, Christian
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (01) : 42 - 48
  • [44] Simulation study of a brain PET scanner using TOF-DOI detectors equipped with first interaction position detection
    Li, Yingying
    Watanabe, Mitsuo
    Isobe, Takashi
    Ote, Kibo
    Tokui, Aoi
    Omura, Tomohide
    Liu, Huafeng
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (01):
  • [45] Wavelength discrimination (WLD) TOF-PET detector with DOI information
    Ullah, Muhammad Nasir
    Pratiwi, Eva
    Park, Jin Ho
    Lee, Kisung
    Choi, Hojong
    Yeom, Jung-Yeol
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (05):
  • [46] Study of a new PET detector design with combined DOI/TOF features
    Spanoudaki, Virginia
    Levin, Craig
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [47] A WHOLE-BODY IMAGING TECHNIQUE FOR PET
    GUERRERO, TM
    HOFFMAN, EJ
    DAHLBOM, M
    HAWKINS, R
    PHELPS, ME
    JOURNAL OF NUCLEAR MEDICINE, 1989, 30 (10) : 1750 - 1750
  • [48] Whole-body imaging with PET/MRI
    Gaa, J
    Rummeny, EJ
    Seemann, MD
    EUROPEAN JOURNAL OF MEDICAL RESEARCH, 2004, 9 (06) : 309 - 312
  • [49] Development of a TOF-PET Detector Capable of Four-Layer DOI Encoding with a Single-Layer Crystal Array
    Tsuda, Tomoaki
    Furumiya, Tetsuo
    Ohi, Lunichi
    Tonami, Hiromichi
    Kitamura, Keishi
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3012 - 3014
  • [50] Timing resolution improvement using DOI information in a four-layer scintillation detector for TOF-PET
    Shibuya, Kengo
    Nishikido, Fumihiko
    Tsuda, Tomoaki
    Kobayashi, Tetsuya
    Lam, Chihfung
    Yamaya, Taiga
    Yoshida, Eiji
    Inadama, Naoko
    Murayama, Hideo
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 593 (03): : 572 - 577