Histopathological comparison of biodegradable polymer and permanent polymer based sirolimus eluting stents in a porcine model of coronary stent implantation

被引:46
|
作者
Koppara, Tobias [1 ]
Joner, Michael [1 ]
Bayer, Gerd [2 ]
Steigerwald, Kristin [1 ]
Diener, Tobias [2 ]
Wittchow, Eric [2 ]
机构
[1] Deutsch Herzzentrum Munich, D-80636 Munich, Germany
[2] BIOTRONIK SE & Co KG, Erlangen, Germany
关键词
Porcine model; sirolimus-eluting stents; restenosis; POOLED ANALYSIS; DURABLE POLYMER; CLINICAL-OUTCOMES; RELEASE PROFILES; COATED STENTS; FOLLOW-UP; IN-VITRO; DRUG; RESTENOSIS; TRIAL;
D O I
10.1160/TH12-01-0043
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Biodegradable stent coatings were recently introduced as a potential solution to overcome sustained inflammatory responses observed with permanent polymer-based drug-eluting stents. In a preliminary study, selected biodegradable or permanent polymer-based sirolimus-eluting stent (SES) formulations were screened for effectiveness in comparison to bare metal stents (BMS) at 28 days. Subsequently, the most favourable SES formulation was compared to commercially available SES (Cypher (TM)) at 28,90 and 180 days to investigate the histopathologic response as well as tissue, blood and organ pharmacokinetics. Overlapping SES implantation was conducted to evaluate vascular healing at 28 days in this particular setting. SES with biodegradable poly (L-lactide) polymer (PLLA) or poly(lactide-co-glycolide) showed the most favourable outcome with regards to reductions in neointimal area in comparison to BMS at 28 days. The PLLA SES showed a similar reduction in neointimal area compared to Cypher (TM) at 28 days, with significant greater reductions at 90 and 180 days (1.7 +/- 0.7 mm(2) vs. 3.1 +/- 1.5 mm(2), p=0.03 and 1.8 +/- 1.2 mm(2) vs. 3.0 +/- 1.5 mm(2), p=0.01, respectively). Sirolimus vascular tissue concentrations were detectable up to 90 days following implantation. Overlapping stented segments showed favourable histopathologic results with respect to fibrin deposition and endothelialisation at 28 days. In conclusion, the use of PLLA as drug-eluting matrix resulted in mild inflammatory responses in the presence of effective sirolimus tissue concentrations. The greater efficacy observed at long-term follow-up in PLLA SES compared to Cypher (TM) may be a multifactorial result of stent design, polymer biocompatibility and improved release kinetics.
引用
收藏
页码:1161 / 1171
页数:11
相关论文
共 50 条
  • [1] HISTOPATHOLOGICAL COMPARISON OF BIODEGRADABLE POLYMER AND PERMANENT POLYMER BASED SIROLIMUS ELUTING STENTS IN A PORCINE MODEL
    Koppara, Tobias
    Wittchow, Eric
    Bayer, Gerd
    Steigerwald, Kristin
    Joner, Michael
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 59 (13) : E215 - E215
  • [2] SIROLIMUS: BIODEGRADABLE POLYMER DRUG ELUTING STENT VERSUS PERMANENT POLYMER DRUG ELUTING STENT
    Kumar, Karooa Aswinee
    Wang Hongyan
    Lou Dayuan
    Qu Peng
    HEART, 2012, 98 : E130 - E130
  • [3] Quantitative assessment of late lumen loss after biodegradable polymer and permanent polymer sirolimus-eluting stents implantation
    KAN Jing
    CHEN Feng
    LIU Li-ya
    XU Hai-mei
    LIN Ling
    LIU Yan
    ZHAO Ying-ying
    CHENG Jiu-pei
    CHEN Shao-liang
    中华医学杂志(英文版), 2013, 126 (06) : 1081 - 1085
  • [4] Quantitative assessment of late lumen loss after biodegradable polymer and permanent polymer sirolimus-eluting stents implantation
    Kan Jing
    Chen Feng
    Liu Li-ya
    Xu Hai-mei
    Lin Ling
    Liu Yan
    Zhao Ying-ying
    Cheng Jiu-pei
    Chen Shao-liang
    CHINESE MEDICAL JOURNAL, 2013, 126 (06) : 1081 - 1085
  • [5] Randomised comparison of vascular response to biodegradable polymer sirolimus eluting and permanent polymer everolimus eluting stents: An optical coherence tomography study
    Koppara, Tobias
    Tada, Tomohisa
    Xhepa, Erion
    Kutner, Sebastian
    Byrne, Robert A.
    Ibrahim, Tareq
    Laugwitz, Karl -Ludwig
    Kastrati, Adnan
    Joner, Michael
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2018, 258 : 42 - 49
  • [6] Comparison of Endothelial Barrier Functional Recovery After Implantation of a Novel Biodegradable-Polymer Sirolimus-Eluting Stent in Comparison to Durable- and Biodegradable-Polymer Everolimus-Eluting Stents
    Sakamoto, Atsushi
    Torii, Sho
    Jinnouchi, Hiroyuki
    Guo, Liang
    Cornelissen, Anne
    Kuntz, Salome
    Cheng, Qi
    Fernandez, Raquel
    Paek, Ka Hyun
    Harris, Kathryn
    Srivastava, Mukta C.
    Kolodgie, Frank D.
    Virmani, Renu
    Finn, Aloke, V
    CARDIOVASCULAR REVASCULARIZATION MEDICINE, 2021, 24 : 1 - 10
  • [7] Biodegradable polymer coated, paclitaxel eluting stents in the porcine coronary arteries
    Milewski, K
    Buszman, P
    Zurakowski, A
    Pajak, J
    Stojko, J
    Abu Samra, M
    AMERICAN JOURNAL OF CARDIOLOGY, 2005, 96 (7A): : 23H - 23H
  • [8] Efficacy and safety of domestic biodegradable polymer sirolimus-eluting stents and durable polymer sirolimus-eluting stents in the treatment of coronary artery disease
    Fang, Yi
    Bin, Wang
    Jing, Li
    Qiu Miaohan
    Han Yaling
    Shuang, Wang
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 70 (16) : C106 - C107
  • [9] Angioscopic Findings of Biodegradable Polymer-Coated Biolimus A9-Eluting Stent Implantation: Comparison With Sirolimus-Eluting Stent
    Awata, Masaki
    Uematsu, Masaaki
    Sera, Fusako
    Ishihara, Takayuki
    Watanabe, Tetsuya
    Fujita, Masashi
    Onishi, Toshinari
    Iida, Osamu
    Ishida, Yoshio
    Nanto, Shinsuke
    Nagata, Seiki
    AMERICAN JOURNAL OF CARDIOLOGY, 2011, 107 (8A): : 75A - 75A
  • [10] BIODEGRADABLE-POLYMER BIOLIMUS ELUTING STENTS SHOW BETTER RESULTS THAN PERMANENT-POLYMER PACLITAXEL ELUTING STENTS IN A PRECLINICAL CORONARY MODEL
    Perez de Prado, Armando
    Perez-Martinez, Claudia
    Cuellas-Ramon, Carlos
    Gonzalo-Orden, Jose M.
    Diego-Nieto, Alejandro
    Regueiro-Purrinos, Marta
    Lopez-Benito, Maria
    Ajenjo-Silverio, Jose M.
    Fernandez-Vazquez, Felipe
    BRITISH JOURNAL OF SURGERY, 2013, 100 : 19 - 19