Characteristics of a plastic scintillation detector in photon beam dosimetry

被引:3
|
作者
Das, Indra [1 ,3 ]
Sohn, Jooyoung [1 ]
Lim, Sara [1 ]
Sengupta, Bishwambhar [1 ]
Feijoo, Marcos [2 ]
Yadav, Poonam [1 ]
机构
[1] Northwestern Univ, Feinberg Sch Med, Northwestern Mem Hosp, Dept Radiat Oncol, Chicago, IL USA
[2] Blue Phys LLC, Lutz, FL USA
[3] Northwestern Univ, Northwestern Mem Hosp, Feinberg Sch Med, Dept Radiat Oncol, Chicago, IL 60611 USA
来源
关键词
characterization; photon beam; plastic scintillating detector; small field dosimetry; CRYSTAL DIAMOND DETECTOR; EXRADIN W1 SCINTILLATOR; RELATIVE DOSIMETRY; PROTOCOL; MV;
D O I
10.1002/acm2.14209
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Plastic scintillating detectors (PSD) have gained popularity due to small size and are ideally suited in small-field dosimetry due to no correction needed and hence detector reading can be compared to dose. Likewise, these detectors are active and water equivalent. A new PSD from Blue Physics is characterized in photon beam.Purpose Innovation in small-field dosimetry detector has led us to examine Blue Physics PSD (BP-PSD) for use in photon beams from linear accelerator.Method sBP-PSD was acquired and its characteristics were evaluated in photon beams from a Varian TrueBeam. Data were collected in a 3D water tank. Standard parameters; dose, dose rate, energy, angular dependence and temperature dependence were studied. Depth dose, profiles and output in a reference condition as well as small fields were measured.Results BP-PSD is versatile and provides data very similar to an ion chamber when Cerenkov radiation is properly accounted. This device measures data pulse by pulse which very few detectors can perform. The differences between ion chamber data and PSD are < 2% in most cases. The angular dependence is a bit pronounces to 1.5% which is due to PSD housing. Depth dose and profiles are comparable within < 1% to an ion chamber. For small fields this detector provides suitable field output factor compared to other detectors and Monte Carlo (MC) simulated data without any added correction factor.Conclusions The characteristics of Blue Physics PSD is uniquely suitable in photon beam and more so in small fields. The data are reproducible compared to ion chamber for most parameters and ideally suitable for small-field dosimetry without any correction factor.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Characterization of a microSilicon diode detector for small-field photon beam dosimetry
    Akino, Yuichi
    Fujiwara, Masateru
    Okamura, Keita
    Shiomi, Hiroya
    Mizuno, Hirokazu
    Isohashi, Fumiaki
    Suzuki, Osamu
    Seo, Yuji
    Tamari, Keisuke
    Ogawa, Kazuhiko
    JOURNAL OF RADIATION RESEARCH, 2020, 61 (03) : 410 - 418
  • [22] Plastic scintillator preparation and coupling in scintillation dosimetry
    Archambault, L.
    Ayotte, G.
    Gingras, L.
    Beddar, A.
    Beaulieu, L.
    MEDICAL PHYSICS, 2006, 33 (06) : 2129 - 2130
  • [23] Plastic scintillation dosimetry and its application to radiotherapy
    Beddar, A. S.
    RADIATION MEASUREMENTS, 2006, 41 : S124 - S133
  • [24] Parameterization of the Photon Beam Dosimetry
    Lebron, S.
    Li, J.
    Yan, G.
    Kahler, D.
    Liu, C.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [25] Patient Specific QA for External Beam Radiotherapy Using the HYPERSCINT Plastic Scintillation Detector
    Goulet, M.
    MEDICAL PHYSICS, 2020, 47 (06) : E825 - E826
  • [26] Dosimetric Evaluation of Plastic Scintillation Detector for Skin Collimation in Electron Beam Radiation Therapy
    Yang, K.
    Sharma, S.
    Ju, A.
    MEDICAL PHYSICS, 2022, 49 (06) : E830 - E830
  • [27] Characteristics and Effective Point of Measurement of a MicroSilicon Detector in Electron Beam Dosimetry
    Das, I.
    Srivastava, S.
    MEDICAL PHYSICS, 2020, 47 (06) : E714 - E715
  • [28] WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .1. PHYSICAL CHARACTERISTICS AND THEORETICAL CONSIDERATIONS
    BEDDAR, AS
    MACKIE, TR
    ATTIX, FH
    PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10): : 1883 - 1900
  • [29] Characterization of a Low-Cost Plastic Fiber Array Detector for Proton Beam Dosimetry
    Ozkan Loch, Cigdem
    Eichenberger, Michael Alexander
    Togno, Michele
    Zinsli, Simon Pascal
    Egloff, Martina
    Papa, Angela
    Ischebeck, Rasmus
    Lomax, Antony John
    Peier, Peter
    Safai, Sairos
    SENSORS, 2020, 20 (20) : 1 - 13
  • [30] CyberKnife multi-site small beam dosimetry with a new plastic scintillator detector
    Russo, S.
    Masi, L.
    Bergantin, A.
    De Martin, E.
    Doro, R.
    Frassanito, C.
    Fumagalli, M. L.
    Martinotti, A. S.
    Rondi, E.
    Vigorito, S.
    Mancosu, P.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S733 - S733