Characteristics of a plastic scintillation detector in photon beam dosimetry

被引:3
|
作者
Das, Indra [1 ,3 ]
Sohn, Jooyoung [1 ]
Lim, Sara [1 ]
Sengupta, Bishwambhar [1 ]
Feijoo, Marcos [2 ]
Yadav, Poonam [1 ]
机构
[1] Northwestern Univ, Feinberg Sch Med, Northwestern Mem Hosp, Dept Radiat Oncol, Chicago, IL USA
[2] Blue Phys LLC, Lutz, FL USA
[3] Northwestern Univ, Northwestern Mem Hosp, Feinberg Sch Med, Dept Radiat Oncol, Chicago, IL 60611 USA
来源
关键词
characterization; photon beam; plastic scintillating detector; small field dosimetry; CRYSTAL DIAMOND DETECTOR; EXRADIN W1 SCINTILLATOR; RELATIVE DOSIMETRY; PROTOCOL; MV;
D O I
10.1002/acm2.14209
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Plastic scintillating detectors (PSD) have gained popularity due to small size and are ideally suited in small-field dosimetry due to no correction needed and hence detector reading can be compared to dose. Likewise, these detectors are active and water equivalent. A new PSD from Blue Physics is characterized in photon beam.Purpose Innovation in small-field dosimetry detector has led us to examine Blue Physics PSD (BP-PSD) for use in photon beams from linear accelerator.Method sBP-PSD was acquired and its characteristics were evaluated in photon beams from a Varian TrueBeam. Data were collected in a 3D water tank. Standard parameters; dose, dose rate, energy, angular dependence and temperature dependence were studied. Depth dose, profiles and output in a reference condition as well as small fields were measured.Results BP-PSD is versatile and provides data very similar to an ion chamber when Cerenkov radiation is properly accounted. This device measures data pulse by pulse which very few detectors can perform. The differences between ion chamber data and PSD are < 2% in most cases. The angular dependence is a bit pronounces to 1.5% which is due to PSD housing. Depth dose and profiles are comparable within < 1% to an ion chamber. For small fields this detector provides suitable field output factor compared to other detectors and Monte Carlo (MC) simulated data without any added correction factor.Conclusions The characteristics of Blue Physics PSD is uniquely suitable in photon beam and more so in small fields. The data are reproducible compared to ion chamber for most parameters and ideally suitable for small-field dosimetry without any correction factor.
引用
下载
收藏
页数:11
相关论文
共 50 条
  • [41] BETA DETECTOR UTILIZING PLASTIC SCINTILLATION FIBERS
    HARDIE, G
    EZOP, JJ
    PRESTON, CC
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1964, 7 (02): : 340 - &
  • [42] NEW RADIATION DETECTOR OF PLASTIC SCINTILLATION FIBER
    IMAI, SI
    SORAMOTO, S
    MOCHIKI, KI
    IGUCHI, T
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (04): : 1093 - 1097
  • [43] Technical note: Visual, rapid, scintillation point dosimetry for in vivo MV photon beam radiotherapy treatments
    Decker, Savannah M.
    Bruza, Petr
    Zhang, Rongxiao
    Williams, Benjamin B.
    Jarvis, Lesley A.
    Pogue, Brian W.
    Gladstone, David J.
    MEDICAL PHYSICS, 2024, 51 (08) : 5754 - 5763
  • [44] Exradin W1 plastic scintillation detector for in vivo skin dosimetry in passive scattering proton therapy
    Alsanea, Fahed
    Wootton, Landon
    Sahoo, Narayan
    Kudchadker, Rajat
    Mahmood, Usama
    Beddar, Sam
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 47 : 58 - 63
  • [45] Evaluation of effective field size characteristics for small megavoltage photon beam dosimetry
    Keivan, H.
    Maskani, R.
    Shahbazi-Gahrouei, D.
    Shanei, A.
    Pandesh, S.
    Sereshke, E. Tarighati
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2022, 20 (01): : 163 - 168
  • [46] BEAM QUALITY SPECIFICATION FOR PHOTON-BEAM DOSIMETRY
    KOSUNEN, A
    ROGERS, DWO
    MEDICAL PHYSICS, 1993, 20 (04) : 1181 - 1188
  • [47] Monte Carlo calculations of correction factors for plastic phantoms in clinical photon and electron beam dosimetry
    Araki, Fujio
    Hanyu, Yuji
    Fukuoka, Miyoko
    Matsumoto, Kenji
    Okumura, Masahiko
    Oguchi, Hiroshi
    MEDICAL PHYSICS, 2009, 36 (07) : 2992 - 3001
  • [48] EVALUATION OF THE DOSIMETRIC CHARACTERISTICS OF A DIAMOND DETECTOR FOR PHOTON-BEAM MEASUREMENTS
    RUSTGI, SN
    MEDICAL PHYSICS, 1995, 22 (05) : 567 - 570
  • [49] Development of an inorganic scintillation detector system for in vivo dosimetry for brachytherapy
    Kertzscher, G.
    Beddar, S.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S89 - S90
  • [50] Photon dosimetry using selective data sampling with Particle Swarm optimization algorithm based on NaI(Tl) scintillation detector
    Taheri Balanoji, Seyed Mortaza
    Zaki Dizaji, Hossein
    Abdi Saray, Akbar
    KERNTECHNIK, 2022, 87 (03) : 364 - 372