Characteristics of a plastic scintillation detector in photon beam dosimetry

被引:3
|
作者
Das, Indra [1 ,3 ]
Sohn, Jooyoung [1 ]
Lim, Sara [1 ]
Sengupta, Bishwambhar [1 ]
Feijoo, Marcos [2 ]
Yadav, Poonam [1 ]
机构
[1] Northwestern Univ, Feinberg Sch Med, Northwestern Mem Hosp, Dept Radiat Oncol, Chicago, IL USA
[2] Blue Phys LLC, Lutz, FL USA
[3] Northwestern Univ, Northwestern Mem Hosp, Feinberg Sch Med, Dept Radiat Oncol, Chicago, IL 60611 USA
来源
关键词
characterization; photon beam; plastic scintillating detector; small field dosimetry; CRYSTAL DIAMOND DETECTOR; EXRADIN W1 SCINTILLATOR; RELATIVE DOSIMETRY; PROTOCOL; MV;
D O I
10.1002/acm2.14209
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Plastic scintillating detectors (PSD) have gained popularity due to small size and are ideally suited in small-field dosimetry due to no correction needed and hence detector reading can be compared to dose. Likewise, these detectors are active and water equivalent. A new PSD from Blue Physics is characterized in photon beam.Purpose Innovation in small-field dosimetry detector has led us to examine Blue Physics PSD (BP-PSD) for use in photon beams from linear accelerator.Method sBP-PSD was acquired and its characteristics were evaluated in photon beams from a Varian TrueBeam. Data were collected in a 3D water tank. Standard parameters; dose, dose rate, energy, angular dependence and temperature dependence were studied. Depth dose, profiles and output in a reference condition as well as small fields were measured.Results BP-PSD is versatile and provides data very similar to an ion chamber when Cerenkov radiation is properly accounted. This device measures data pulse by pulse which very few detectors can perform. The differences between ion chamber data and PSD are < 2% in most cases. The angular dependence is a bit pronounces to 1.5% which is due to PSD housing. Depth dose and profiles are comparable within < 1% to an ion chamber. For small fields this detector provides suitable field output factor compared to other detectors and Monte Carlo (MC) simulated data without any added correction factor.Conclusions The characteristics of Blue Physics PSD is uniquely suitable in photon beam and more so in small fields. The data are reproducible compared to ion chamber for most parameters and ideally suitable for small-field dosimetry without any correction factor.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Passively scattered proton beam entrance dosimetry with a plastic scintillation detector
    Wootton, Landon
    Holmes, Charles
    Sahoo, Narayan
    Beddar, Sam
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (03): : 1185 - 1198
  • [2] Plastic scintillation detector validation for kV dosimetry
    Lessard, F.
    Archambault, L.
    Plamondon, M.
    Despres, P.
    Therriault-Proulx, F.
    Beddar, S.
    Beaulieu, L.
    [J]. MEDICAL PHYSICS, 2012, 39 (07) : 4646 - 4646
  • [3] Proton Beam Entrance Dosimetry with the Exradin W1 Plastic Scintillation Detector
    Wootton, L.
    Holmes, C.
    Sahoo, N.
    Beddar, S.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3727 - 3727
  • [4] Development of a Liquid Scintillation Detector for External Beam Dosimetry
    Sinn, D.
    Mackenzie, M.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3741 - 3742
  • [5] Plastic Scintillation Detector Calibration Procedure for Brachytherapy in Vivo Dosimetry
    Herreros, A.
    Perez-Calatayud, J.
    Ballester, F.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E473 - E473
  • [6] A high sensitivity plastic scintillation detector for in vivo dosimetry of LDR brachytherapy
    Therriault-Proulx, F.
    Beaulieu, L.
    Beddar, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S116 - S116
  • [7] Photon and electron beam dosimetry with a CVD diamond detector
    Fidanzio, A
    Azario, L
    Viola, P
    Ascarelli, P
    Cappelli, E
    Conte, G
    Piermattei, A
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 524 (1-3): : 115 - 123
  • [8] Technical Note: Characteristics of a microSilicon X shielded diode detector for photon beam dosimetry
    Akino, Yuichi
    Okamura, Keita
    Das, Indra J.
    Isohashi, Fumiaki
    Seo, Yuji
    Tamari, Keisuke
    Hirata, Takero
    Hayashi, Kazuhiko
    Inoue, Shinichi
    Ogawa, Kazuhiko
    [J]. MEDICAL PHYSICS, 2021, 48 (04) : 2004 - 2009
  • [9] Validating plastic scintillation detectors for photon dosimetry in the radiologic energy range
    Lessard, Francois
    Archambault, Louis
    Plamondon, Mathieu
    Despres, Philippe
    Therriault-Proulx, Francois
    Beddar, Sam
    Beaulieu, Luc
    [J]. MEDICAL PHYSICS, 2012, 39 (09) : 5308 - 5316
  • [10] Review of plastic and liquid scintillation dosimetry for photon, electron, and proton therapy
    Beaulieu, Luc
    Beddar, Sam
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (20): : R305 - R343