Factors influencing size-specific dose estimates of selected computed tomography protocols at two clinical practices in South Africa

被引:2
|
作者
Sebelego, Ida-Keshia [1 ]
Acho, Sussan [2 ]
van der Merwe, Belinda [1 ]
Rae, William I. D. [2 ,3 ]
机构
[1] Cent Univ Technol, Fac Hlth & Environm Sci, Dept Clin Sci, ZA-9301 Bloemfontein, South Africa
[2] Univ Free State, Fac Hlth Sci, Dept Med Phys, ZA-9300 Bloemfontein, South Africa
[3] Prince Wales Hosp, Med Imaging Dept, Randwick 2133, Australia
基金
新加坡国家研究基金会;
关键词
DIAGNOSTIC REFERENCE LEVELS; TUBE CURRENT MODULATION; IMAGE QUALITY; CT; IMPACT;
D O I
10.1093/rpd/ncad059
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The study aimed to determine the factors that impact the size-specific dose estimate (SSDE) for computed tomography (CT) examinations of the chest-abdomen-pelvis and abdomen-pelvis protocols in two clinical radiology practices and evaluate the image quality of these protocols. Imaging parameters, protocols, dose metrics from the CT units and size-related parameters to calculate the SSDE were documented. The image quality of the CT images was assessed using an image subtraction algorithm. The SSDE increased as the volumetric CT dose index (CTDIvol), and the patient's body mass index increased, respectively. Significant differences (p < 0.001) occurred between the two hospitals regarding image quality. However, these differences were not indicative of differences in the diagnostic performances for task-based imaging protocols. Different clinical protocols should be reviewed to optimise dose. The inclusion of the pre-monitoring sequence, age of the machine and the scan requisition parameters impacted the SSDEs. Image quality should be assessed to evaluate the consistency of image quality between protocols applied by different CT units when assessing SSDEs.
引用
收藏
页码:588 / 602
页数:15
相关论文
共 50 条
  • [21] Determination of radiation dose during computed tomography examinations in southern Saudi Arabian hospitals using size-specific dose estimates
    Saeed, Mohammed K.
    Asiri, Abdullah Ali M.
    MAEJO INTERNATIONAL JOURNAL OF SCIENCE AND TECHNOLOGY, 2021, 15 (02) : 147 - 158
  • [22] Size-Specific Dose Estimates Based on Water-Equivalent Diameter and Effective Diameter in Computed Tomography Coronary Angiography
    Xu, Jian
    Wang, Xiangquan
    Xiao, Huawei
    Xu, Jianguo
    MEDICAL SCIENCE MONITOR, 2019, 25 : 9299 - 9305
  • [23] Institutional computed tomography diagnostic reference levels based on water-equivalent diameter and size-specific dose estimates
    Boos, Johannes
    Thomas, Christoph
    Appel, Elisabeth
    Klosterkemper, Yan
    Schleich, Christoph
    Aissa, Joel
    Bethge, Oliver Th
    Antoch, Gerald
    Kroepil, Patric
    JOURNAL OF RADIOLOGICAL PROTECTION, 2018, 38 (02) : 536 - 548
  • [24] ACCURACY OF SIZE-SPECIFIC DOSE ESTIMATE CALCULATION FROM CENTER SLICE IN COMPUTED TOMOGRAPHY
    Boos, Johannes
    Kroepil, Patric
    Bethge, Oliver Th.
    Aissa, Joel
    Schleich, Christoph
    Sawicki, Lino Morris
    Heinzler, Niklas
    Antoch, Gerald
    Thomas, Christoph
    RADIATION PROTECTION DOSIMETRY, 2018, 178 (01) : 8 - 19
  • [25] Development of size-specific institutional diagnostic reference levels for computed tomography protocols in neck imaging
    Steuwe A.
    Thomas C.
    Kraus B.
    Bethge O.T.
    Aissa J.
    Klosterkemper Y.
    Antoch G.
    Boos J.
    Journal of Radiological Protection, 2020, 40 (01) : 68 - 82
  • [26] USEFULNESS OF SIZE-SPECIFIC DOSE ESTIMATES IN PEDIATRIC COMPUTED TOMOGRAPHY: REVALIDATION OF LARGE-SCALE PEDIATRIC CT DOSE SURVEY DATA IN JAPAN
    Yamazaki, Daisuke
    Miyazaki, Osamu
    Takei, Yasutaka
    Matsubara, Kosuke
    Shinozaki, Masafumi
    Shimada, Yoshiya
    Suzuki, Shoichi
    Muramatsu, Yoshihisa
    RADIATION PROTECTION DOSIMETRY, 2018, 179 (03) : 254 - 262
  • [27] Assessment of patient size for size-specific dose estimates using two methods: a comparative study
    Abuhaimed, A.
    Martin, C.
    Demirkaya, O.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S906 - S906
  • [28] Size-Specific Dose Estimates (SSDE) Factors for Head Scans of Different Imaging Systems
    Abuhaimed, A.
    Martin, C. J.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S759 - S759
  • [29] Local Diagnostic Reference Levels and Size-Specific Patient Dose in Computed Tomography at a Greek Hospital
    Stasinou, D.
    Zoiopoulou, V.
    Antonakos, J.
    Argentos, S.
    Efstathopoulos, E.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [30] Effect of table height displacement and patient center deviation on size-specific dose estimates calculated from computed tomography localizer radiographs
    Kuriyama, Kazuki
    Matsubara, Kosuke
    Hisahara, Shu
    Nagata, Yukie
    Nosaka, Rumiko
    Goto, Rika
    Yanano, Nobutaka
    Shimizu, Kanichiro
    Shoji, Tomokazu
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (02) : 665 - 672