System characterization and performance studies with MACACO III Compton camera

被引:8
|
作者
Barrientos, L. [1 ]
Borja-Lloret, M. [1 ]
Casana, J. V. [1 ]
Hueso-Gonzalez, F. [1 ]
Ros, A. [1 ]
Roser, J. [1 ]
Senra, C. [1 ]
Solaz, C. [1 ]
Viegas, R. [1 ]
Llosa, G. [1 ]
机构
[1] CSIC UV, Inst Fis Corpuscular IF, Valencia, Spain
关键词
Compton camera; Scintillator crystals; Silicon photomultipliers; PROTON THERAPY; RECONSTRUCTION; FEASIBILITY; TELESCOPE; BEAMS;
D O I
10.1016/j.radphyschem.2023.110922
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The IRIS group of IFIC-Valencia has developed a Compton camera prototype. The system detectors are made of Lanthanum (III) bromide scintillator crystals coupled to silicon photomultipliers. Two models of silicon photomultipliers arrays with different micro pixel pitch (25 and 50 mu m) have been chosen as possible candidates to improve the response of the new system. Characterization studies with a 22Na point-like source have indicated that the 25 mu m photodetector provided better performance in terms of energy resolution (5.2% FWHM at 511 keV) and angular resolution (6.9 degrees FWHM at 1275 keV), and more stability with temperature variations. In addition, MACACO III imaging capabilities have been assessed using a structure composed of thirty-seven 22Na point-like sources. Furthermore, in order to evaluate possible ways of improving the system performance, several studies have been carried out by means of simulations both in realistic and performance improved conditions. In this work, the system performance is evaluated for its future application in different areas.
引用
收藏
页数:13
相关论文
共 50 条
  • [31] A Compton Camera prototype simulation study: Camera performance and first tests of range monitoring capabilities
    Lozano, Valencia I. I.
    Dedes, G.
    Aldawood, S.
    Liprandi, S.
    Miani, A.
    Zoglauer, A.
    Lauber, K.
    Thirolf, P. G.
    Parodi, K.
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [32] Studies of Prototype Dual Head Compton Camera for Applications in Medical Imaging
    Peterson, S.
    Leigh, M.
    Mackin, D.
    Chen, H.
    Maggi, P.
    Beddar, S.
    Polf, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E491 - E491
  • [33] Development and performance analysis of double-layer separated Compton camera
    Wang W.
    Li X.
    Wu J.
    Pang H.
    Luo Z.
    He Jishu/Nuclear Techniques, 2023, 46 (03):
  • [34] Performance Study of An Electron-Tracking Compton Camera for Medical Imaging
    Kabuki, S.
    Kimura, H.
    Kubo, H.
    Ogawa, K.
    Kunieda, E.
    Tanimori, T.
    MEDICAL PHYSICS, 2012, 39 (06) : 3639 - 3639
  • [35] Experimental performance of double-scattering Compton camera with anthropomorphic phantom
    Seo, H.
    Park, J. H.
    Ushakov, A.
    Kim, C. H.
    Kim, J. K.
    Lee, J. H.
    Lee, C. S.
    Lee, J. S.
    JOURNAL OF INSTRUMENTATION, 2011, 6
  • [36] Monte Carlo simulations on performance of double-scattering Compton camera
    Park, J. H.
    Seo, H.
    Kim, Y. S.
    Kim, C. H.
    Lee, J. H.
    Lee, C. S.
    Kim, S. M.
    Lee, J. S.
    JOURNAL OF INSTRUMENTATION, 2012, 7
  • [37] Gamma-ray imaging performance of the GRI plus Compton camera
    Caffrey, A.
    Harkness-Brennan, L. J.
    Alshammari, H.
    Boston, A. J.
    Boston, H. C.
    Griffiths, O.
    Judson, D. S.
    Kalantan, S.
    Le Crom, B.
    Nolan, P. J.
    Powell, A.
    Platt, J.
    Randall, G.
    Reid, C.
    Rintoul, E.
    Unsworth, C.
    Woodroof, T. F.
    Burrows, I.
    Lazarus, I. H.
    Simpson, J.
    Atkinson, K. D.
    JOURNAL OF INSTRUMENTATION, 2021, 16 (09)
  • [38] Development of a radioactive substance detection system integrating a Compton camera and a LiDAR camera with a hexapod robot
    Sato, Yuki
    Kakuto, Takeshi
    Tanaka, Takayuki
    Shimano, Hiroyuki
    Morohashi, Yuko
    Hatakeyama, Tomoyoshi
    Nakajima, Junsaku
    Ishiyama, Masahiro
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2024, 1063
  • [39] A performance study of an electron-tracking Compton camera with a compact system for environmental gamma-ray observation
    Mizumoto, T.
    Tomono, D.
    Takada, A.
    Tanimori, T.
    Komura, S.
    Kubo, H.
    Matsuoka, Y.
    Mizumura, Y.
    Nakamura, K.
    Nakamura, S.
    Oda, M.
    Parker, J. D.
    Sawano, T.
    Bando, N.
    Nabetani, A.
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [40] Multi head compton camera system for nuclear medicine: Simulation study
    Shirahata, T.
    Kohara, R.
    Nakazawa, T.
    Tanimori, T.
    Ogawa, K.
    Kubo, H.
    Kabuki, S.
    Miyazaki, O.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3423 - +