Development and validation of an optimal GATE model for proton pencil-beam scanning delivery

被引:1
|
作者
Asadi, Ali [1 ]
Akhavanallaf, Azadeh [2 ]
Hosseini, Seyed Abolfazl [1 ]
Vosoughi, Naser [1 ]
Zaidi, Habib [2 ,3 ,4 ,5 ,6 ]
机构
[1] Sharif Univ Technol, Dept Energy Engn, Tehran, Iran
[2] Geneva Univ Hosp, Div Nucl Med & Mol Imaging, Geneva, Switzerland
[3] Univ Geneva, Neuroctr, Geneva, Switzerland
[4] Univ Groningen, Univ Med Ctr Groningen, Dept Nucl Med & Mol Imaging, Groningen, Netherlands
[5] Univ Southern Denmark, Dept Nucl Med, Odense, Denmark
[6] Geneva Univ Hosp, Div Nucl Med & Mol Imaging, CH-1211 Geneva, Switzerland
来源
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK | 2023年 / 33卷 / 04期
基金
瑞士国家科学基金会;
关键词
Proton therapy; Monte Carlo simulation; active scanning; AAPM TG-119; prostate; MONTE-CARLO SIMULATIONS; CARBON-ION THERAPY; DOSE-CALCULATION; OPTIMIZATION; CODE; DISTRIBUTIONS; PARAMETERS; EQUIVALENT; PLATFORM; TOPAS;
D O I
10.1016/j.zemedi.2022.10.008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To develop and validate a versatile Monte Carlo (MC)-based dose calculation engine to support MC-based dose verification of treatment planning systems (TPSs) and quality assurance (QA) workflows in proton therapy. Methods: The GATE MC toolkit was used to simulate a fixed horizontal active scan-based proton beam delivery (SIEMENS IONTRIS). Within the nozzle, two primary and secondary dose monitors have been designed to enable the comparison of the accuracy of dose estimation from MC simulations with respect to physical QA measurements. The developed beam model was validated against a series of commissioning measurements using pinpoint chambers and 2D array ionization chambers (IC) in terms of lateral profiles and depth dose distributions. Furthermore, beam delivery module and treatment planning has been validated against the literature deploying various clinical test cases of the AAPM TG-119 (c-shape phantom) and a prostate patient. Results: MC simulations showed excellent agreement with measurements in the lateral depth-dose parameters and spread-out Bragg peak (SOBP) characteristics within a maximum relative error of 0.95 mm in range, 1.83% in entrance to peak ratio, 0.27% in mean point-to-point dose difference, and 0.32% in peak location. The mean relative absolute difference between MC simulations and measurements in terms of absorbed dose in the SOBP region was 0.93% +/- 0.88%. Clinical phantom studies showed a good agreement compared to research TPS (relative error for TG-119 planning target volume PTV-D95 = 1.8%; and for prostate PTV-D95 = -0.6%). Conclusion: We successfully developed a MC model for the pencil beam scanning system, which appears reliable for dose verification of the TPS in combination with QA information, prior to patient treatment.
引用
收藏
页码:591 / 600
页数:10
相关论文
共 50 条
  • [1] Development and Validation of an Optimal GATE Model for Pencil Beam Scanning (PBS) Proton Beam Delivery
    Asadi, Ali
    Varnaseri, Seraj
    Moshiri, Shirin
    Akhavanallaf, Azadeh
    Arabi, Hossein
    Zaidi, Habib
    [J]. 2023 30TH NATIONAL AND 8TH INTERNATIONAL IRANIAN CONFERENCE ON BIOMEDICAL ENGINEERING, ICBME, 2023, : 471 - 474
  • [2] A CASE STUDY IN PROTON PENCIL-BEAM SCANNING DELIVERY
    Kooy, Hanne M.
    Clasie, Benjamin M.
    Lu, Hsiao-Ming
    Madden, Thomas M.
    Bentefour, Hassan
    Depauw, Nicolas
    Adams, Judy A.
    Trofimov, Alexei V.
    Demaret, Denis
    Delaney, Thomas F.
    Flanz, Jacob B.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (02): : 624 - 630
  • [3] Golden beam data for proton pencil-beam scanning
    Clasie, Benjamin
    Depauw, Nicolas
    Fransen, Maurice
    Goma, Carles
    Panahandeh, Hamid Reza
    Seco, Joao
    Flanz, Jacob B.
    Kooy, Hanne M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (05): : 1147 - 1158
  • [4] Development and validation of an optimal GATE model for double scattering proton beam delivery
    Piruzan, E.
    Vosoughi, N.
    Mahani, H.
    [J]. JOURNAL OF INSTRUMENTATION, 2021, 16 (02)
  • [5] Dosimetric Impact of CT Metal Artifacts on Proton Pencil-Beam Scanning Delivery
    Verburg, J.
    Joshi, M.
    Bazalova, M.
    Madden, T.
    Kooy, H.
    Seco, J.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [6] Measurements of Proton Pencil-Beam Scanning Performance Characteristics
    Clasie, B.
    Bentefour, H.
    Kooy, H.
    Flanz, J.
    [J]. MEDICAL PHYSICS, 2008, 35 (06)
  • [7] Optimal minimum MU for intensity-modulated proton therapy with pencil-beam scanning proton beams
    Yi, Byongyong
    Mossahebi, Sina
    Jatczak, Jenna
    Mundis, Michelle
    Houser, Thomas
    Alicia, David
    Han, Dong
    Gonzalez, Rosette
    Olis, Stephen
    Zakhary, Mark
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024,
  • [8] Single Imager Proton Radiography with a Pencil-Beam Scanning System
    Harms, J.
    Maloney, L.
    Lin, Y.
    Liu, T.
    Erickson, A.
    Zhang, R.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E659 - E660
  • [9] Clinical Implementation of Proton Therapy Using Pencil-Beam Scanning Delivery Combined With Static Apertures
    Baeumer, Christian
    Plaude, Sandija
    Khalil, Dalia Ahmad
    Geismar, Dirk
    Kramer, Paul-Heinz
    Kroninger, Kevin
    Nitsch, Christian
    Wulff, Joerg
    Timmermann, Beate
    [J]. FRONTIERS IN ONCOLOGY, 2021, 11
  • [10] A simulator of proton pencil beam scanning delivery
    Poulsen, P. R.
    Nystrom, H.
    Skyt, P. S.
    Jensen, M. F.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S124 - S125