Development and validation of an optimal GATE model for proton pencil-beam scanning delivery

被引:2
|
作者
Asadi, Ali [1 ]
Akhavanallaf, Azadeh [2 ]
Hosseini, Seyed Abolfazl [1 ]
Vosoughi, Naser [1 ]
Zaidi, Habib [2 ,3 ,4 ,5 ,6 ]
机构
[1] Sharif Univ Technol, Dept Energy Engn, Tehran, Iran
[2] Geneva Univ Hosp, Div Nucl Med & Mol Imaging, Geneva, Switzerland
[3] Univ Geneva, Neuroctr, Geneva, Switzerland
[4] Univ Groningen, Univ Med Ctr Groningen, Dept Nucl Med & Mol Imaging, Groningen, Netherlands
[5] Univ Southern Denmark, Dept Nucl Med, Odense, Denmark
[6] Geneva Univ Hosp, Div Nucl Med & Mol Imaging, CH-1211 Geneva, Switzerland
来源
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK | 2023年 / 33卷 / 04期
基金
瑞士国家科学基金会;
关键词
Proton therapy; Monte Carlo simulation; active scanning; AAPM TG-119; prostate; MONTE-CARLO SIMULATIONS; CARBON-ION THERAPY; DOSE-CALCULATION; OPTIMIZATION; CODE; DISTRIBUTIONS; PARAMETERS; EQUIVALENT; PLATFORM; TOPAS;
D O I
10.1016/j.zemedi.2022.10.008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To develop and validate a versatile Monte Carlo (MC)-based dose calculation engine to support MC-based dose verification of treatment planning systems (TPSs) and quality assurance (QA) workflows in proton therapy. Methods: The GATE MC toolkit was used to simulate a fixed horizontal active scan-based proton beam delivery (SIEMENS IONTRIS). Within the nozzle, two primary and secondary dose monitors have been designed to enable the comparison of the accuracy of dose estimation from MC simulations with respect to physical QA measurements. The developed beam model was validated against a series of commissioning measurements using pinpoint chambers and 2D array ionization chambers (IC) in terms of lateral profiles and depth dose distributions. Furthermore, beam delivery module and treatment planning has been validated against the literature deploying various clinical test cases of the AAPM TG-119 (c-shape phantom) and a prostate patient. Results: MC simulations showed excellent agreement with measurements in the lateral depth-dose parameters and spread-out Bragg peak (SOBP) characteristics within a maximum relative error of 0.95 mm in range, 1.83% in entrance to peak ratio, 0.27% in mean point-to-point dose difference, and 0.32% in peak location. The mean relative absolute difference between MC simulations and measurements in terms of absorbed dose in the SOBP region was 0.93% +/- 0.88%. Clinical phantom studies showed a good agreement compared to research TPS (relative error for TG-119 planning target volume PTV-D95 = 1.8%; and for prostate PTV-D95 = -0.6%). Conclusion: We successfully developed a MC model for the pencil beam scanning system, which appears reliable for dose verification of the TPS in combination with QA information, prior to patient treatment.
引用
收藏
页码:591 / 600
页数:10
相关论文
共 50 条
  • [41] Pencil-beam scanning catheter for intracoronary optical coherence tomography
    Jiqiang Kang
    Rui Zhu
    Yunxu Sun
    Jianan Li
    Kenneth K.Y.Wong
    Opto-Electronic Advances, 2022, 5 (03) : 82 - 88
  • [42] Pencil-beam scanning proton therapy for anal cancer: a dosimetric comparison with intensity-modulated radiotherapy
    Ojerholm, Eric
    Kirk, Maura L.
    Thompson, Reid F.
    Zhai, Huifang
    Metz, James M.
    Both, Stefan
    Ben-Josef, Edgar
    Plastaras, John P.
    ACTA ONCOLOGICA, 2015, 54 (08) : 1209 - 1217
  • [43] Pencil-beam scanning catheter for intravascular optical coherence tomography
    Kang, Jiqiang
    Zhu, Rui
    Sun, Yunxu
    Li, Jianan
    Wong, Kenneth K. Y.
    2020 ASIA COMMUNICATIONS AND PHOTONICS CONFERENCE (ACP) AND INTERNATIONAL CONFERENCE ON INFORMATION PHOTONICS AND OPTICAL COMMUNICATIONS (IPOC), 2020,
  • [44] Quantifying Lateral Penumbra with a Prototype Aperture for Pencil Beam Scanning Proton Beam Delivery
    Ewart, S.
    Zimmerman, S.
    Lee, E.
    Speth, J.
    Mascia, A.
    MEDICAL PHYSICS, 2020, 47 (06) : E837 - E837
  • [45] Incorporation of the aperture thickness in proton pencil-beam dose calculations
    Slopsema, Roelf L.
    Kooy, Hanne M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (21): : 5441 - 5453
  • [46] Monte Carlo model of a gantry for proton pencil beam scanning
    Winterhalter, Carla
    Safai, Sairos
    Grassberger, Clemens
    Oxley, David
    Weber, Damien C.
    Lomax, Tony
    STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 (11) : 837 - 837
  • [47] Pencil-beam scanning catheter for intracoronary optical coherence tomography
    Kang, Jiqiang
    Zhu, Rui
    Sun, Yunxu
    Li, Jianan
    Wong, Kenneth K. Y.
    OPTO-ELECTRONIC ADVANCES, 2022, 5 (03)
  • [48] A Knowledge-Based Planning Approach for Dosimetric Prediction of Prostate Proton Pencil-Beam Scanning Plans
    Zou, W.
    Teo, B.
    O'Reilly, S.
    Ainsley, C.
    Lin, A.
    Xiao, Y.
    Dong, L.
    Yin, L.
    MEDICAL PHYSICS, 2018, 45 (06) : E658 - E658
  • [49] Comment on 'Collimated proton pencil-beam scanning for superficial targets: impact of the order of range shifter and aperture'
    Winterhalter, Carla
    Lomax, Antony J.
    Oxley, David
    Weber, Damien C.
    Safai, Sairos
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (20):
  • [50] Commissioning of a Mevion S250i with Hyperscan Pencil-Beam Scanning Proton Therapy System
    Jin, H.
    Chen, Y.
    Kendall, E.
    Ahmad, S.
    MEDICAL PHYSICS, 2019, 46 (06) : E325 - E325