Image quality and radiation dose assessment for the clinically applied 16-slice CT scanner using PMMA phantom and quality assurance phantom

被引:0
|
作者
Mekonin, Tadelech Sisay [1 ]
Deressu, Tilahun Tesfaye [1 ]
机构
[1] Addis Ababa Univ, Dept Phys, POB 1176, Addis Ababa, Ethiopia
来源
关键词
Computed tomography scanner; Image quality parameter; CT dose index; Quality assurance phantom; PMMA phantom; Medical radiation dose; COEFFICIENTS; ATTENUATION;
D O I
10.1186/s43055-023-01038-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The goal of this study was to assess the radiation dose delivered accuracy and the acceptable image quality parameter from the 16-slice Philips CT scanner in order to evaluate the current methodology for quality controls of computed tomography systems. The measured volume computed tomography dose index was 101 mGy and the displayed volume computed tomography dose index was 10.2 mGy and the dose length product is 680 mGy*cm for head scan, and the measured volume computed tomography dose index was 6.1 mGy and the displayed volume computed tomography dose index was 6.3 mGy and the dose length product was 587 mGy*cm for body scan, respectively. The image quality parameters were 4.4 HU, 2.3 HU, and 10 HU for uniformity, contrast-to-noise ratio, and computed tomography number accuracy (CT number accuracy) for quality assurance phantom, respectively. We propose that discussions and collaboration regarding patient and particular equipment adjustments be coordinated among radiologists, medical imaging technologists, and medical physicists. This can increase image quality, reduce absorbed radiation, and improve hospital medical care.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] A multi-scanner evaluation of PET image quality using phantom studies
    Surti, S
    Badawi, RD
    Holdsworth, CH
    El Fakhri, G
    Kinahan, PE
    Karp, JS
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2425 - 2427
  • [32] A multileaf collimator phantom for the quality assurance of radiation therapy planning systems and CT simulators
    McNiven, A
    Kron, T
    Van Dyk, J
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (03): : 994 - 1001
  • [33] A PHANTOM BASED STUDY OF CT IMAGE QUALITY AND PATIENT RADIATION DOSE FOR NUCLEAR MEDICINE ATTENUATION CORRECTION
    Midgley, S.
    INTERNAL MEDICINE JOURNAL, 2016, 46 : 12 - 12
  • [34] Effect of iterative reconstruction techniques on image quality in low radiation dose chest CT: a phantom study
    Xu, Yan
    Zhang, Ting-ting
    Hu, Zhi-hai
    Li, Juan
    Hou, Hong-jun
    Xu, Zu-shan
    He, Wen
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2019, 25 (06) : 442 - 450
  • [35] The impact of ASiR-V on abdominal CT radiation dose and image quality-A phantom study
    Morup, Svea Deppe
    Stowe, John
    Kusk, Martin Weber
    Precht, Helle
    Foley, Shane
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2022, 53 (03) : 453 - 459
  • [36] Evaluation of coronary CT scans radiation dose and image quality using different scanning protocol on a 256-slice CT scanner
    Fareed, A. M.
    Pernes, J. M.
    Gaux, J. C.
    Oraby, M.
    Nasr, G. M.
    Maklady, F.
    Dupouy, P.
    EUROPEAN HEART JOURNAL SUPPLEMENTS, 2010, 12 (0F) : F90 - F90
  • [37] Evaluation Of Coronary CT Scans Radiation Dose And Image Quality Using Different Scanning Protocols On a 256-slice CT Scanner
    Fareed, Ahmed M.
    Gaux, Jean Claude
    Pernes, Jean Marc
    Oraby, Mohamed
    Nasr, Gamila
    Maklady, Fathy
    Dupouy, Patrick
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 56 (13) : B89 - B89
  • [38] Quality assurance of 4D-CT using a programmable motorized phantom
    Simon, L
    Giraud, P
    Mazal, A
    Rosenwald, J
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S91 - S91
  • [39] Dose reduction and image quality in CT examinations using an iterative reconstruction algorithm: a phantom study
    Guariglia, S.
    Meliado, G.
    Zivelonghi, E.
    Pinali, L.
    Montemezzi, S.
    Cavedon, C.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2015, 1 (04):
  • [40] Evaluation of radiation dose and image quality of CT scan for whole-body pediatric PET/CT: A phantom study
    Yang, Ching-Ching
    Liu, Shu-Hsin
    Mok, Greta S. P.
    Wu, Tung-Hsin
    MEDICAL PHYSICS, 2014, 41 (09)