Micro-CT Evaluation of Microgaps at Implant-Abutment Connection

被引:1
|
作者
Kowalski, Jakub [1 ]
Puszkarz, Adam K. [2 ]
Radwanski, Mateusz [3 ]
Sokolowski, Jerzy [1 ]
Cichomski, Michal [4 ]
Bourgi, Rim [5 ]
Hardan, Louis [5 ]
Sauro, Salvatore [6 ,7 ]
Lukomska-Szymanska, Monika [1 ]
机构
[1] Med Univ Lodz, Dept Gen Dent, PL-92213 Lodz, Poland
[2] Lodz Univ Technol, Inst Mat Sci Text & Polymer Composites, Fac Mat Technol & Text Design, 116 Zeromskiego St, PL-90924 Lodz, Poland
[3] Med Univ Lodz, Dept Endodont, PL-92213 Lodz, Poland
[4] Univ Lodz, Fac Chem, Dept Mat Technol & Chem, Pomorska 163, PL-90236 Lodz, Poland
[5] St Joseph Univ, Sch Dent, Dept Restorat Dent, Beirut 11072180, Lebanon
[6] Univ CEU Cardenal Herrera, Fac Cienciasde Salud, Dept Odontol, Dent Biomat & Minimally Invas Dent, C-Del Pozo SS-N, Valencia 46115, Spain
[7] I M Sechenov First Moscow State Med Univ, Dept Therapeut Dent, Moscow 119146, Russia
关键词
dental implant; microgap; micro-computed tomography; abutment; IN-VITRO ANALYSIS; DENTAL IMPLANTS; VERTICAL MISFIT; MARGINAL FIT; INTERFACE; TITANIUM; FATIGUE; LEAKAGE; DESIGN; ACCURACY;
D O I
10.3390/ma16124491
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The assessment of microgaps at the implant-abutment interface is an important factor that may influence clinical success. Thus, the aim of this study was to evaluate the size of microgaps between prefabricated and customised abutments (Astra Tech, Dentsply, York, PA, USA; Apollo Implants Components, Pabianice, Poland) mounted on a standard implant. The measurement of the microgap was performed using micro-computed tomography (MCT). Due to 15-degree rotation of samples, 24 microsections were obtained. Scans were performed at four levels established at the interface between the abutment and the implant neck. Moreover, the volume of the microgap was evaluated. The size of the microgap at all measured levels varied from 0.1 to 3.7 & mu;m for Astra and from 0.1 to 4.9 & mu;m for Apollo (p > 0.05). Moreover, 90% of the Astra specimens and 70% of the Apollo specimens did not exhibit any microgaps. The highest mean values of microgap size for both groups were detected at the lowest portion of the abutment (p > 0.05). Additionally, the average microgap volume was greater for Apollo than for Astra (p > 0.05). It can be concluded that most samples did not exhibit any microgaps. Furthermore, the linear and volumetric dimensions of microgaps observed at the interface between Apollo or Astra abutments and Astra implants were comparable. Additionally, all tested components presented microgaps (if any) that were clinically acceptable. However, the microgap size of the Apollo abutment was higher and more variable than that of the Astra one.
引用
收藏
页数:14
相关论文
共 50 条
  • [41] In Vitro Analysis of the Implant-Abutment Interface Connection and Bacterial Infiltration in Two Extraoral Implant Models
    de Souza, Ricardo Salgado
    Suffredini, Ivana Barbosa
    Cortizo, Daniela Lattuf
    Larsson, Anna
    Nannmark, Ulf
    Dib, Luciano Lauria
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2020, 35 (01) : 63 - 69
  • [42] COMMENTARY. radiographical evaluation of the gap at the implant-abutment interface
    Tarnow, Dennis
    JOURNAL OF ESTHETIC AND RESTORATIVE DENTISTRY, 2010, 22 (04) : 251 - 251
  • [43] Assessment of Leakage at the Implant-Abutment Connection Using a New Gas Flow Method
    Fauroux, Marie-Alix
    Levallois, Bernard
    Yachouh, Jacques
    Torres, Jacques-Henri
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2012, 27 (06) : 1409 - 1412
  • [44] Mechanics of the implant-abutment connection: An 8-degree taper compared to a butt joint connection
    Merz, BR
    Hunenbart, S
    Belser, UC
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2000, 15 (04) : 519 - 526
  • [45] Analysis of Implant Strength After Implantoplasty in Three Implant-Abutment Connection Designs: An In Vitro Study
    Alexandre Gehrke, Sergio
    Aramburu Junior, Jaime Sarda
    Dedavid, Berenice Anina
    Shibli, Jamil Awad
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2016, 31 (03) : E65 - E70
  • [46] Effect of Different Implant-Abutment Connection Materials on the Fracture Resistance of Zirconia Abutments
    AlAmar, Mohammed
    Alqahtani, Fawaz
    JOURNAL OF ORAL IMPLANTOLOGY, 2020, 46 (02) : 88 - 92
  • [47] The Mechanical Behavior of a Screwless Morse Taper Implant-Abutment Connection: An In Vitro Study
    Bagegni, Aimen
    Weihrauch, Vincent
    Vach, Kirstin
    Kohal, Ralf
    MATERIALS, 2022, 15 (09)
  • [48] Stability of Implant-Abutment Connection After Using the Rescue Kit: An In Vitro Study
    Sprenger, Achim
    Gutauskas, Kristupas
    Filippi, Andreas
    Rosentritt, Martin
    Kuehl, Sebastian
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2018, 33 (06) : 1274 - 1278
  • [49] Effect of Serrating Abutment-Implant Mating Surface on Torque Stability of Implant-Abutment Connection, Before and After Cyclic Loading
    Rezayani, Vida
    Alikhasi, Marzieh
    Monzavi, Abbas
    JOURNAL OF ORAL IMPLANTOLOGY, 2020, 46 (06) : 555 - 561
  • [50] Evaluation of different debridement strategies for implant-abutment connections: An in vitro
    Lang, Lilly K.
    Karl, Matthias
    Klaer, Virgilia
    JOURNAL OF PROSTHETIC DENTISTRY, 2023, 130 (05): : 738.e1 - 738.e8