Effect of syringe underfilling on the quality of venous blood gas analysis

被引:2
|
作者
Lippi, Giuseppe [1 ]
Pighi, Laura [2 ,3 ]
Tosi, Marco [2 ,3 ]
Vettori, Marco [2 ,3 ]
Celegon, Giovanni [2 ,3 ]
Favaloro, Emmanuel J. [4 ,5 ,6 ]
Salvagno, Gian Luca [2 ,3 ]
机构
[1] Univ Hosp Verona, Sect Clin Biochem, Piazzale LA Scuro 10, I-37134 Verona, Italy
[2] Univ Verona, Sect Clin Biochem, Verona, Italy
[3] Univ Verona, Sch Med, Verona, Italy
[4] Westmead Hosp, Inst Clin Pathol & Med Res ICPMR, Sydney Ctr Thrombosis & Haemostasis, Dept Haematol, Westmead, NSW, Australia
[5] Univ Sydney, Westmead Hosp, Fac Med & Hlth, Sch Med Sci, Westmead, NSW, Australia
[6] Charles Sturt Univ, Fac Sci & Hlth, Wagga Wagga, NSW, Australia
关键词
blood gas analysis; syringe; underfilling; errors; preanalytical variability; HEPARIN; VOLUME;
D O I
10.1515/dx-2023-0096
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives There is limited information on the influence of collecting small amounts of blood on the quality of blood gas analysis. Therefore, the purpose of this study was to investigate the effects of different degrees of underfilling of syringes on test results of venous blood gas analysis.Methods Venous blood was collected by venipuncture from 19 healthcare workers in three 1.0 mL syringes for blood gas analysis, by manually aspirating different volumes of blood (i.e., 1.0, 0.5 and 0.25 mL). Routine blood gas analysis was then immediately performed with GEM Premier 5,000. The results of the two underfilled syringes were compared with those of the reference syringe filled with appropriate blood volume.Results The values of most assayed parameters did not differ significantly in the two underfilled syringes. Statistically significant variations were found for lactate, hematocrit and total hemoglobin, the values of which gradually increased as the fill volume diminished, as well as for sodium concentration, which decreased in both insufficiently filled blood gas syringes. The bias was clinically meaningful for lactate in syringe filled with 0.25 mL of blood, and for hematocrit, total hemoglobin and sodium in both syringes containing 0.5 and 0.25 mL of blood.Conclusions Collection of smaller volumes of venous blood than the specified filling volume in blood gas syringes may have an effect on the quality of some test results, namely lactate, hematocrit, total hemoglobin and sodium. Specific indications must be given for standardizing the volume of blood to be collected within these syringes.
引用
收藏
页码:91 / 96
页数:6
相关论文
共 50 条
  • [31] QUALITY ASSURANCE IN CLINICAL BLOOD-GAS ANALYSIS
    DELANEY, CJ
    LEARY, ET
    KENNY, MA
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1976, : 7 - 7
  • [32] Peripheral venous blood gas analysis versus arterial blood gas analysis for the diagnosis of respiratory failure and metabolic disturbance in adults.
    Byrne, Anthony
    Bennett, Michael
    Symons, Rebecca
    Chaterji, Robindro
    Pace, Nathan
    Thomas, Paul
    EUROPEAN RESPIRATORY JOURNAL, 2018, 52
  • [33] VENOUS OR ARTERIAL BLOOD GAS MEASUREMENT
    LONG, AP
    JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1971, 217 (12): : 1706 - &
  • [34] Arterial and Venous Blood Gas Analyses
    Rieser, Teresa M.
    TOPICS IN COMPANION ANIMAL MEDICINE, 2013, 28 (03) : 86 - 90
  • [35] AN EVALUATION OF THE EFFECT OF VASODILATATION ON OXYGEN SATURATIONS MEASURED BY PULSE OXIMETRY AND VENOUS-BLOOD GAS-ANALYSIS
    BROOME, IJ
    MILLS, GH
    SPIERS, P
    REILLY, CS
    ANAESTHESIA, 1993, 48 (05) : 415 - 416
  • [36] Comparison of blood gas values in arterial and venous blood
    Chellam Kirubakaran
    J. Ebor Jacob Gnananayagam
    E. Kalaranjini Sundaravalli
    The Indian Journal of Pediatrics, 2003, 70 (10) : 781 - 785
  • [37] Effects of syringe type and storage conditions on results of equine blood gas and acid-base analysis
    Kennedy, Sarah A.
    Constable, Peter D.
    Sen, Ismail
    Couetil, Laurent
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2012, 73 (07) : 979 - 987
  • [38] EFFECT OF SYRINGE MATERIAL ON OXYGEN TENSION IN STORED BLOOD
    HILTY, H
    KARENDAL, B
    ACTA SOCIETATIS MEDICORUM UPSALIENSIS, 1969, 74 (3-4) : 193 - &
  • [39] EFFECT OF BLOOD COLLECTION ON ARTERIAL BLOOD-GAS ANALYSIS
    THELIN, OP
    POURCYROUS, M
    COOKE, RJ
    PEDIATRIC RESEARCH, 1990, 27 (04) : A227 - A227
  • [40] HEPARINS EFFECT ON BLOOD-GAS ANALYSIS
    ZUERLEIN, TJ
    AMERICAN JOURNAL OF DISEASES OF CHILDREN, 1990, 144 (12): : 1287 - 1288