Single-test Ritchey-Common interferometry

被引:0
|
作者
Liu, Yiming [1 ]
Xu, Chuheng [1 ]
Hu, Yao [1 ]
Lu, Huiqin [2 ]
Li, Jinpeng [2 ]
Wu, Tiancheng [2 ]
Zhang, Shaohui [1 ]
Chang, Xu [3 ]
Hao, Qun [1 ]
机构
[1] Beijing Inst Technol, Sch Opt & Photon, Beijing Key Lab Precis Optoelect Measurement Instr, Beijing 100081, Peoples R China
[2] Chinese Acad Sci, Nanjing Astron Instruments Co Ltd, Nanjing 210042, Jiangsu, Peoples R China
[3] Natl Inst Metrol, Beijing, Peoples R China
基金
中国国家自然科学基金;
关键词
LARGE OPTICS; TELESCOPE;
D O I
10.1364/OE.505632
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The Ritchey-Common test is widely adopted to measure large optical flats. The traditional Ritchey-Common test eliminates the defocus error with multiple tests by changing the position of the mirrors, which suffers from cumbersome steps, poor repeatability, coupled system error, extra mirror deformation, and potential overturning. The above problems increase the test time, decrease the reliability and accuracy, increase the test cost, and threaten manufacturing safety. We propose a single-test Ritchey-Common interferometry to avoid the obligatory position change in the traditional method. A sub-aperture of test flat is directly measured by a small-aperture interferometer before the test, which is easy to implement, to replace the extra system wavefront measurement in different positions. The defocus is calculated in sub-aperture at exactly the same position as the full-field measurement without the position change, then the surface form under test can be obtained with accurate optical path modeling. Measurement experiments for 100 mm and 2050 mm aperture flats were performed to demonstrate the feasibility of this method. Compared with a direct test in a standard Zygo interferometer, the peak to valley (PV) and root mean square (RMS) errors were 0.0889 ? and 0.0126 ? (?=632.8 nm), respectively, which reaches the upper limit of accuracy of the interferometer. To the best of our knowledge, this is the first proposal of the Ritchey-Common test that can eliminate the defocus error and realize high accuracy measurement in a single test. Our work paves the way for reliable and practical optical metrology for large optical flats.
引用
收藏
页码:62 / 78
页数:17
相关论文
共 50 条
  • [1] Study on the Ritchey-Common interferometry for large piano optics
    National Astronomical Observatories/Nanjing Institute of Astronomical Optics and Technology, Chinese Academy of Sciences, Nanjing 210042, China
    不详
    Guangxue Jishu, 2007, 5 (737-740+744): : 737 - 740
  • [2] Virtual-real combination Ritchey-Common interferometry
    Liu, Yiming
    Hu, Yao
    Zhang, Shaohui
    Huang, Shen
    Li, Jinpeng
    Yan, Limin
    Lv, Jiahang
    Wang, Zhen
    Chang, Xu
    Hao, Qun
    OPTICS EXPRESS, 2022, 30 (09) : 15777 - 15795
  • [3] Ritchey-common test analyzes uncoated optics
    Zecchino, M. (mikez@veeco.com), 1600, Laurin Publishing Co. Inc. (37):
  • [4] Eliminating alignment error and analyzing Ritchey angle accuracy in Ritchey-Common test
    Zhu, Shuo
    Zhang, Xiaohui
    OPTICS COMMUNICATIONS, 2014, 311 : 368 - 374
  • [5] Application of Ritchey-Common test in large flat measurements
    Han, S
    Novak, E
    Schurig, M
    OPTICAL MEASUREMENT SYSTEMS FOR INDUSTRIAL INSPECTION II: APPLICATIONS IN PRODUCTION ENGINEERING, 2001, 4399 : 131 - 136
  • [6] Ritchey-Common test analyzes uncoated optics
    Zecchino, M
    PHOTONICS SPECTRA, 2003, 37 (07) : 34 - 34
  • [7] EXACT RAY TRACE ANALYSIS OF THE RITCHEY-COMMON TEST
    SHU, KL
    PARKS, RE
    SHANNON, RR
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1981, 71 (12) : 1587 - 1587
  • [8] Ritchey-Common Test used for Measurement of Astronomical Optic
    Han, S
    Novak, E
    Schurig, M
    SPECIALIZED OPTICAL DEVELOPMENTS IN ASTRONOMY, 2003, 4842 : 270 - 273
  • [9] Influence of reference sphere on test accuracy of Ritchey-Common method
    Zhu, Shuo
    Zhang, Xiaohui
    IET OPTOELECTRONICS, 2024, 18 (1-2) : 1 - 10
  • [10] USING RITCHEY-COMMON TEST FOR LARGE PLANE MIRRORS (FLATS)
    HOUSTON, JB