Survival Analysis and Failure Modes of Total Hip Arthroplasty Using a Cemented Semi-Retentive Acetabular Cup

被引:0
|
作者
Stan, Gabriel [1 ]
Roman, Mihai Dan [2 ]
Orban, Horia [1 ]
Georgeanu, Vlad Alexandru [1 ]
Deculescu, Rares Stefan [1 ]
Brinduse, Lacramioara Aurelia [1 ]
Gheorghiu, Nicolae [1 ]
机构
[1] Carol Davila Univ Med & Pharm, Fac Med, Bucharest 050474, Romania
[2] Univ Lucian Blaga Sibiu, Fac Med, Sibiu 550024, Romania
关键词
constrained; dislocation; failure; survival; total hip arthroplasty; FEMORAL-HEAD SIZE; RETROSPECTIVE ANALYSIS; DISLOCATION; REVISION; SURVIVORSHIP; COMPONENTS; LINER; RISK;
D O I
10.3390/jcm12247506
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To reduce the incidence of total hip revisions, there have been continuous efforts to enhance prosthetic materials and designs to optimize implant survival. A primary implant with a constrained acetabular component is often used to minimize the risk of dislocations even though this approach has some drawbacks as reported in the literature. To address these concerns, this study aimed to assess the survivorship and dislocation rate of a semi-retentive cemented acetabular cup when used as a primary implant. The specific cemented cup that we studied was not present in any study that we consulted, so to fill this gap, we conducted a retrospective examination of 527 cemented hip prostheses that utilized the semi-retentive cup between the years 2005 and 2012. We employed Cox multiple regression models for our statistical analysis. The revision due to dislocation occurred in 12.8% of all cases, with a lower incidence of 5% (14 cases) in age groups >70 years than in age groups <70 years (14%-32 cases) (p < 0.001). The survival rates of the semi-retentive cemented acetabular cup were 98.6% (520 cases) at 5 years and 92.2% (487 cases) at 10 years. The survival rates were significantly lower in women than men, with 1.9% (7 cases) toward 0% at 5 years and 8.1% (30 cases) toward 5% (7 cases) at 10 years (p = 0.002). The difference in failure rates between age groups over 70 years (2.3%-10 cases) and age groups under 70 years (11.5%-34 cases) was also statistically significant (p < 0.001). Our study indicates that the semi-constrained design may cause frequent damage to the polyethylene liner due to impingement and wear, which are the primary factors for failure. Also, this implant has a similar risk of revision due to dislocation as reported in studies and may be beneficial as a primary implant in elderly patients with low-demanding lifestyles, muscular insufficiency, and low compliance regarding hip prosthetic behavior, without a major effect on survivorship.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Effect of acetabular cup position and orientation in cemented total hip arthroplasty
    Hirakawa, K
    Mitsugi, N
    Koshino, T
    Saito, T
    Hirasawa, Y
    Kubo, T
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2001, (388) : 135 - 142
  • [2] INCIDENCE AND MECHANISM OF FAILURE OF CEMENTED ACETABULAR COMPONENT IN TOTAL HIP ARTHROPLASTY
    EFTEKHAR, NS
    NERCESSIAN, O
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 1988, 19 (03) : 557 - 566
  • [3] Total hip arthroplasty: survival and modes of failure
    Karachalios, Theofilos
    Komnos, George
    Koutalos, Antonios
    EFORT OPEN REVIEWS, 2018, 3 (05): : 232 - 239
  • [4] Cemented total hip arthroplasty following acetabular fracture
    Scott, C. E. H.
    MacDonald, D.
    Moran, M.
    White, T. O.
    Patton, J. T.
    Keating, J. F.
    BONE & JOINT JOURNAL, 2017, 99B (10): : 1399 - 1408
  • [5] Acetabular cup size trends in total hip arthroplasty
    Mckenna, Daniel Patrick
    Price, Alex
    Mcaleese, Timothy
    Dahly, Darren
    Mckenna, Paul
    Cleary, May
    WORLD JOURNAL OF ORTHOPEDICS, 2024, 15 (01): : 39 - 44
  • [6] Should the acetabular cup be medialised in total hip arthroplasty?
    Bonnin, Michel P.
    Archbold, Pooler H. A.
    Basiglini, Luca
    Selmi, Tarik A.
    Beverland, David E.
    HIP INTERNATIONAL, 2011, 21 (04) : 428 - 435
  • [7] ACETABULAR CUP WEAR IN TOTAL HIP-ARTHROPLASTY
    RIMNAC, CM
    WILSON, PD
    FUCHS, MD
    WRIGHT, TM
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 1988, 19 (03) : 631 - 636
  • [8] Cementless total hip arthroplasty with a threaded acetabular cup
    J. Fernández-González
    R. Canosa Sevillano
    C. García-Araujo
    International Orthopaedics, 1997, 21 : 327 - 331
  • [9] Cementless total hip arthroplasty with a threaded acetabular cup
    Fernandez-Gonzalez, J
    Sevillano, RC
    Garcia-Araujo, C
    INTERNATIONAL ORTHOPAEDICS, 1997, 21 (05) : 327 - 331
  • [10] Improving Acetabular Cup Orientation in Total Hip Arthroplasty by Using Smartphone Technology
    Meermans, Geert
    Kats, Jan Jaap
    JOURNAL OF ARTHROPLASTY, 2013, 28 (07): : 1246 - 1246