Biomechanical Effect of Hybrid Dynamic Stabilization Implant on the Segmental Motion and Intradiscal Pressure in Human Lumbar Spine

被引:5
|
作者
Hsiao, Chih-Kun [1 ]
Tsai, Yi-Jung [1 ]
Yen, Cheng-Yo [2 ]
Li, Yi-Chen [2 ]
Hsiao, Hao-Yuan [2 ]
Tu, Yuan-Kun [2 ]
机构
[1] E Da Hosp, Dept Med Res, Kaohsiung 82445, Taiwan
[2] E Da Hosp, Dept Orthopaed, Kaohsiung 82445, Taiwan
来源
BIOENGINEERING-BASEL | 2023年 / 10卷 / 01期
关键词
dynamic stabilization systems; Dynesys-Transition-Optima; range of motion; intradiscal pressure;
D O I
10.3390/bioengineering10010031
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
The hybrid dynamic stabilization system, Dynesys-Transition-Optima, represents a novel pedicle-based construct for the treatment of lumbar degenerative disease. The theoretical advantage of this system is to stabilize the treated segment and preserve the range of motion within the adjacent segment while potentially decreasing the risk of adjacent segment disease following lumbar arthrodesis. Satisfactory short-term outcomes were previously demonstrated in the Dynesys-Transition-Optima system. However, long-term follow-up reported accelerated degeneration of adjacent segments and segmental instability above the fusion level. This study investigated the biomechanical effects of the Dynesys-Transition-Optima system on segment motion and intradiscal pressure at adjacent and implanted levels. Segmental range of motion and intradiscal pressure were evaluated under the conditions of the intact spine, with a static fixator at L4-5, and implanted with DTO at L3-4 (Dynesys fixator) and L4-5 (static fixator) by applying the loading conditions of flexion/extension (+/- 7.5 Nm) and lateral bending (+/- 7.5 Nm), with/without a follower preload of 500 N. Our results showed that the hybrid Dynesys-Transition-Optima system can significantly reduce the ROM at the fusion level (L4-L5), whereas the range of motion at the adjacent level (L3-4) significantly increased. The increase in physiological loading could be an important factor in the increment of IDP at the intervertebral discs at the lumbar spine. The Dynesys-Transition-Optima system can preserve the mobility of the stabilized segments with a lesser range of motion on the transition segment; it may help to prevent the occurrence of adjacent segment degeneration. However, the current study cannot cover all the issues of adjacent segmental diseases. Future investigations of large-scale and long-term follow-ups are needed.
引用
收藏
页数:14
相关论文
共 50 条
  • [41] THE EFFECT OF MULTIPLANAR LOADING ON THE INTRADISCAL PRESSURE OF THE WHOLE HUMAN SPINE:A SYSTEMATIC REVIEW AND META-ANALYSIS
    Liebsch, C.
    Wilke, H. J.
    EUROPEAN CELLS & MATERIALS, 2021, 41 : 388 - 400
  • [42] Biomechanical Study of Lumbar Spine (L2-L4) Using Hybrid Stabilization Device - A Finite Element Analysis
    Jain, Pushpdant
    Khan, Mohammed Rajik
    INTERNATIONAL JOURNAL OF MANUFACTURING MATERIALS AND MECHANICAL ENGINEERING, 2020, 10 (01) : 20 - 32
  • [43] Sagittal plane motion in the human lumbar spine: Comparison of the in vitro quasistatic neutral zone and dynamic motion parameters
    Gay, Ralph E.
    Ilharreborde, Brice
    Zhao, Kristin
    Zhao, Chunfeng
    An, Kai-Nan
    CLINICAL BIOMECHANICS, 2006, 21 (09) : 914 - 919
  • [44] The effect of follower load on the range of motion, facet joint force, and intradiscal pressure of the cervical spine: a finite element study
    Xin-Yi Cai
    Chen-Xi YuChi
    Cheng-Fei Du
    Zhong-Jun Mo
    Medical & Biological Engineering & Computing, 2020, 58 : 1695 - 1705
  • [45] The effect of follower load on the range of motion, facet joint force, and intradiscal pressure of the cervical spine: a finite element study
    Cai, Xin-Yi
    Yuchi, Chen-Xi
    Du, Cheng-Fei
    Mo, Zhong-Jun
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2020, 58 (08) : 1695 - 1705
  • [46] Dynamic stabilization adjacent to single-level fusion: Part I. Biomechanical effects on lumbar spinal motion
    Strube, Patrick
    Tohtz, Stephan
    Hoff, Eike
    Gross, Christian
    Perka, Carsten
    Putzier, Michael
    EUROPEAN SPINE JOURNAL, 2010, 19 (12) : 2171 - 2180
  • [47] Dynamic stabilization adjacent to single-level fusion: Part I. Biomechanical effects on lumbar spinal motion
    Patrick Strube
    Stephan Tohtz
    Eike Hoff
    Christian Gross
    Carsten Perka
    Michael Putzier
    European Spine Journal, 2010, 19 : 2171 - 2180
  • [48] Effect of Spacer Diameter of the Dynesys Dynamic Stabilization System on the Biomechanics of the Lumbar Spine: A Finite Element Analysis
    Shih, Shih-Liang
    Chen, Chen-Sheng
    Lin, Hung-Ming
    Huang, Li-Ying
    Liu, Chien-Lin
    Huang, Chang-Hung
    Cheng, Cheng-Kung
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2012, 25 (05): : E140 - E149
  • [49] Effect of the Cord Pretension of the Dynesys Dynamic Stabilization System on the Biomechanics of the Lumbar Spine: A Finite Element Analysis
    Zhong, Z. C.
    Liu, C. L.
    Hsu, H. W.
    Yang, B. S.
    Chen, C. S.
    6TH WORLD CONGRESS OF BIOMECHANICS (WCB 2010), PTS 1-3, 2010, 31 : 533 - +
  • [50] Clinical Outcomes and Complications After Pedicle-anchored Dynamic or Hybrid Lumbar Spine Stabilization A Systematic Literature Review
    Prud'homme, Marion
    Barrios, Carlos
    Rouch, Philippe
    Charles, Yann Philippe
    Steib, Jean-Paul
    Skalli, Wafa
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2015, 28 (08): : E439 - E448