Analytical Performances of the Novel i-STAT Alinity Point-of-Care Analyzer
被引:5
|作者:
Larcher, Romaric
[1
,2
]
Lottelier, Maxence
[3
]
Badiou, Stephanie
[1
]
Dupuy, Anne-Marie
[3
]
Bargnoux, Anne-Sophie
[1
]
Cristol, Jean-Paul
[1
]
机构:
[1] Univ Montpellier, Montpellier Univ Hosp, INSERM French Inst Hlth & Med Res, CNRS French Natl Ctr Sci Res,PhyMedExp Physiol & E, F-34000 Montpellier, France
[2] Nimes Univ Hosp, Infect & Trop Dis Dept, F-30000 Nimes, France
[3] Montpellier Univ Hosp, Biochem & Hormonol Dept, F-34000 Montpellier, France
来源:
关键词:
analytical performance;
Point-of-Care Analyzer;
i-STAT Alinity;
blood gas;
lactate;
chemistry;
urea;
creatinine;
hematology;
coagulation;
BLOOD;
COAGULATION;
AGREEMENT;
THERAPY;
D O I:
10.3390/diagnostics13020297
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
Many Point-of-Care devices have been released over the past decade. However, data regarding their analytical performances in real-world situations remains scarce. Herein, we aimed to assess the analytical performances of the i-STAT Alinity system. We conducted an analytical performances study with the i-STAT Alinity device using cartridges CG4+ (pH, Pco(2), Po-2, lactate, bicarbonate and base excess); CHEM8+ (Na, K, Cl, ionized Ca, urea, creatinine, glucose, hematocrit and hemoglobin) and PT/INR (prothrombin time and international normalized ratio). We assessed the imprecision and compared the results to those obtained on existing instruments in the central laboratory. We found that the within-lab coefficients of variation (CV) were very low (<2%) or low (2-5%), except for creatinine and PT (CV = 5.2% and CV = 6.3%, respectively). For almost all the parameters, the results were strongly (R-2 = 90-95%) or very strongly (R-2 > 95%) correlated with those of the existing laboratory instruments, and the biases were very low (<2%) or low (2-5%). However, correlations of the PT and INR measurements with existing instruments were lower (R-2 = 86.0% and 89.7%), and biases in the Po-2 (7.9%), creatinine (5.4%) and PT (-6.6%) measurements were higher. The i-STAT Alinity appeared as a convenient device for measurements of numerous parameters. However, clinicians should interpret Po-2, creatinine and PT results with caution.