Analytical Performances of the Novel i-STAT Alinity Point-of-Care Analyzer

被引:5
|
作者
Larcher, Romaric [1 ,2 ]
Lottelier, Maxence [3 ]
Badiou, Stephanie [1 ]
Dupuy, Anne-Marie [3 ]
Bargnoux, Anne-Sophie [1 ]
Cristol, Jean-Paul [1 ]
机构
[1] Univ Montpellier, Montpellier Univ Hosp, INSERM French Inst Hlth & Med Res, CNRS French Natl Ctr Sci Res,PhyMedExp Physiol & E, F-34000 Montpellier, France
[2] Nimes Univ Hosp, Infect & Trop Dis Dept, F-30000 Nimes, France
[3] Montpellier Univ Hosp, Biochem & Hormonol Dept, F-34000 Montpellier, France
关键词
analytical performance; Point-of-Care Analyzer; i-STAT Alinity; blood gas; lactate; chemistry; urea; creatinine; hematology; coagulation; BLOOD; COAGULATION; AGREEMENT; THERAPY;
D O I
10.3390/diagnostics13020297
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Many Point-of-Care devices have been released over the past decade. However, data regarding their analytical performances in real-world situations remains scarce. Herein, we aimed to assess the analytical performances of the i-STAT Alinity system. We conducted an analytical performances study with the i-STAT Alinity device using cartridges CG4+ (pH, Pco(2), Po-2, lactate, bicarbonate and base excess); CHEM8+ (Na, K, Cl, ionized Ca, urea, creatinine, glucose, hematocrit and hemoglobin) and PT/INR (prothrombin time and international normalized ratio). We assessed the imprecision and compared the results to those obtained on existing instruments in the central laboratory. We found that the within-lab coefficients of variation (CV) were very low (<2%) or low (2-5%), except for creatinine and PT (CV = 5.2% and CV = 6.3%, respectively). For almost all the parameters, the results were strongly (R-2 = 90-95%) or very strongly (R-2 > 95%) correlated with those of the existing laboratory instruments, and the biases were very low (<2%) or low (2-5%). However, correlations of the PT and INR measurements with existing instruments were lower (R-2 = 86.0% and 89.7%), and biases in the Po-2 (7.9%), creatinine (5.4%) and PT (-6.6%) measurements were higher. The i-STAT Alinity appeared as a convenient device for measurements of numerous parameters. However, clinicians should interpret Po-2, creatinine and PT results with caution.
引用
收藏
页数:12
相关论文
共 50 条
  • [41] Analytical performance of the i-STAT cardiac troponin I assay
    Apple, FS
    Murakami, MM
    Christenson, RH
    Campbell, JL
    Miller, CJ
    Hock, KG
    Scott, MG
    CLINICA CHIMICA ACTA, 2004, 345 (1-2) : 123 - 127
  • [42] Evaluation of Safety of Warfarin Dose Adjustment Based on Correlation Between Critical Point-of-Care International Normalized Ratios Using the i-STAT Analyzer and Venipuncture International Normalized Ratios
    Janiga, X.
    Alfred, K.
    Menajovsky, L. B.
    JOURNAL OF THROMBOSIS AND THROMBOLYSIS, 2010, 29 (02) : 255 - 255
  • [43] Which point-of-care creatinine analyser for radiology: direct comparison of the i-Stat and StatStrip creatinine methods with different sample types
    Dimeski, Goce
    Tilley, Vera
    Jones, Brock W.
    Brown, Nigel N.
    ANNALS OF CLINICAL BIOCHEMISTRY, 2013, 50 (01) : 47 - 52
  • [44] Evaluation of the CoaguChek S, CoaguChek XS and i-STAT for Point of Care INR testing
    Beyer, L. K.
    Wockenfus, A.
    McBane, R.
    Chaudhry, R.
    Karon, B.
    Santrach, P.
    CLINICAL CHEMISTRY, 2008, 54 (06) : A198 - A199
  • [45] Analytical evaluation of i-STAT portable clinical analyzer for blood gas and electrolyte analyses in pediatric samples.
    Murthy, JN
    Hicks, JM
    Soldin, SJ
    CLINICAL CHEMISTRY, 1996, 42 (06) : 198 - 198
  • [46] Evaluation of the i-STAT portable clinical analyzer in chickens (Gallus gallus)
    Steinmetz, Hanspeter W.
    Vogt, Rainer
    Kastner, Sabine
    Riond, Barbara
    Hatt, Jean-Michel
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2007, 19 (04) : 382 - 388
  • [47] Analytical performance of the i-STAT cardiac troponin I assay.
    Apple, FS
    Murakami, MM
    Christenson, RH
    Campbell, JL
    Miller, CJ
    Hock, KG
    Scott, MG
    CLINICAL CHEMISTRY, 2004, 50 (06) : A9 - A9
  • [48] Oxygen Saturation Readouts Generated by the i-STAT Point-of-Care Analyzer Are More Accurate Than Those Generated by the Masimo Radical-7 Pulse Oximeter During Profound Alkalemia
    Demers, Bob
    Cosa, Joshua
    Welch, Melvin
    CHEST, 2016, 150 (04) : 329A - 329A
  • [49] Accuracy of capillary whole blood international normalized ratio on the CoaguChek S, CoaguChek XS, and i-STAT 1 point-of-care analyzers
    Karon, Brad S.
    McBane, Robert D., II
    Chaudhry, Rajeev
    Beyer, Lisa K.
    Santrach, Paula J.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2008, 130 (01) : 88 - 92
  • [50] Multicenter study on the analytical performance of a new point-of-care blood gas analyzer
    Goerlach-Graw, A
    Engel, B
    Halwachs-Baumann, G
    Kost, G
    Kraft, M
    Küster, H
    Miedema, K
    Nagel, D
    Normand, H
    Okorodudu, A
    Olander, B
    Rolinski, B
    Wiley, B
    Wu, J
    CLINICAL CHEMISTRY, 2004, 50 (06) : A171 - A171