Analytical Performances of the Novel i-STAT Alinity Point-of-Care Analyzer

被引:5
|
作者
Larcher, Romaric [1 ,2 ]
Lottelier, Maxence [3 ]
Badiou, Stephanie [1 ]
Dupuy, Anne-Marie [3 ]
Bargnoux, Anne-Sophie [1 ]
Cristol, Jean-Paul [1 ]
机构
[1] Univ Montpellier, Montpellier Univ Hosp, INSERM French Inst Hlth & Med Res, CNRS French Natl Ctr Sci Res,PhyMedExp Physiol & E, F-34000 Montpellier, France
[2] Nimes Univ Hosp, Infect & Trop Dis Dept, F-30000 Nimes, France
[3] Montpellier Univ Hosp, Biochem & Hormonol Dept, F-34000 Montpellier, France
关键词
analytical performance; Point-of-Care Analyzer; i-STAT Alinity; blood gas; lactate; chemistry; urea; creatinine; hematology; coagulation; BLOOD; COAGULATION; AGREEMENT; THERAPY;
D O I
10.3390/diagnostics13020297
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Many Point-of-Care devices have been released over the past decade. However, data regarding their analytical performances in real-world situations remains scarce. Herein, we aimed to assess the analytical performances of the i-STAT Alinity system. We conducted an analytical performances study with the i-STAT Alinity device using cartridges CG4+ (pH, Pco(2), Po-2, lactate, bicarbonate and base excess); CHEM8+ (Na, K, Cl, ionized Ca, urea, creatinine, glucose, hematocrit and hemoglobin) and PT/INR (prothrombin time and international normalized ratio). We assessed the imprecision and compared the results to those obtained on existing instruments in the central laboratory. We found that the within-lab coefficients of variation (CV) were very low (<2%) or low (2-5%), except for creatinine and PT (CV = 5.2% and CV = 6.3%, respectively). For almost all the parameters, the results were strongly (R-2 = 90-95%) or very strongly (R-2 > 95%) correlated with those of the existing laboratory instruments, and the biases were very low (<2%) or low (2-5%). However, correlations of the PT and INR measurements with existing instruments were lower (R-2 = 86.0% and 89.7%), and biases in the Po-2 (7.9%), creatinine (5.4%) and PT (-6.6%) measurements were higher. The i-STAT Alinity appeared as a convenient device for measurements of numerous parameters. However, clinicians should interpret Po-2, creatinine and PT results with caution.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] Assessment of using fingerstick blood sample with i-STAT point-of-care device for cardiac troponin I assay
    Loewenstein, Devin
    Stake, Christine
    Cichon, Mark
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2013, 31 (08): : 1236 - 1239
  • [22] QUALITY-CONTROL FOR THE I-STAT ANALYZER
    UTZ, C
    GIBSON, C
    HORTIN, GL
    CLINICAL CHEMISTRY, 1995, 41 (06) : S157 - S157
  • [23] Analysis of bloodgas, electrolytes and glucose from intraosseous samples using an i-STAT® point-of-care analyser
    Veldhoen, Esther S.
    de Vooght, Karen M. K.
    Slieker, Martijn G.
    Versluys, Anne B.
    Turner, Nigel Mcb.
    RESUSCITATION, 2014, 85 (03) : 359 - 363
  • [24] Assessing Performance of i-STAT at the Point of Care in the Emergency Room
    Yeo, Chin Pin
    Ngo, Adeline
    Ng, Wai Yoong
    Lim, Swee Han
    Jacob, Edward
    PROCEEDINGS OF SINGAPORE HEALTHCARE, 2011, 20 (03) : 157 - 161
  • [25] Measurement of ionized calcium in canine blood samples collected in prefilled and self-filled heparinized syringes using the i-STAT point-of-care analyzer
    Tappin, Simon
    Rizzo, Francesca
    Dodkin, Steve
    Papasouliotis, Kostas
    Tasker, Severine
    Murphy, Kate
    VETERINARY CLINICAL PATHOLOGY, 2008, 37 (01) : 66 - 72
  • [26] Use of the i-STAT portable clinical analyzer in mice
    Tinkey, Peggy
    Lembo, Thomas
    Craig, Suzanne
    West, Cheri
    Van Pelt, Carolyn
    LAB ANIMAL, 2006, 35 (02) : 45 - 50
  • [27] Analytical performance characteristics of the i-SmartCare 10 analyzer for point-of-care testing
    Kim, S.
    Oh, J.
    Park, H.
    CLINICA CHIMICA ACTA, 2022, 530 : S101 - S101
  • [28] Evaluation of the i-STAT Portable Clinical Analyzer in dogs
    Matburger, C
    Henke, J
    Hirschberger, J
    Matburger, S
    Erhardt, W
    TIERARZTLICHE PRAXIS AUSGABE KLEINTIERE HEIMTIERE, 2000, 28 (02): : 132 - 137
  • [29] Use of the i-STAT portable clinical analyzer in mice
    Peggy Tinkey
    Thomas Lembo
    Suzanne Craig
    Cheri West
    Carolyn Van Pelt
    Lab Animal, 2006, 35 : 45 - 50
  • [30] Evaluation of the i-STAT point-of-care capillary whole blood prothrombin time and international normalized ratio: Comparison to the Tcoag MDAII coagulation analyzer in the central laboratory
    Pena, J. A.
    Lewandrowski, K. B.
    Lewandrowski, E. L.
    Gregory, K.
    Baron, J. M.
    Van Cott, E. M.
    CLINICA CHIMICA ACTA, 2012, 413 (11-12) : 955 - 959