Comparison of four commercial dose calculation algorithms in different evaluation tests

被引:3
|
作者
Rostami, Aram [1 ]
De Castro Neto, Aluisio Jose [1 ]
Paloor, Satheesh Prasad [1 ]
Khalid, Abdul Sattar [1 ]
Hammoud, Rabih [1 ]
机构
[1] Natl Ctr Canc Care & Res, Dept Radiat Oncol, Doha, Qatar
关键词
Treatment Planning System (TPS); Anisotropic Analytical Algorithm (AAA); Acuros (AXB); collapsed cone; convolution (CCC); Monte Carlo (MC); TREATMENT PLANNING SYSTEM; MONTE-CARLO; CLINICAL IMPLEMENTATION; DOSIMETRIC ACCURACY; ACUROS XB; PHOTON; VALIDATION; ENERGY; VERIFICATION; RADIOTHERAPY;
D O I
10.3233/XST-230079
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
BACKGROUND: Accurate and fast dose calculation is crucial in modern radiation therapy. Four dose calculation algorithms (AAA, AXB, CCC, and MC) are available in Varian Eclipse and RaySearch Laboratories RayStation Treatment Planning Systems (TPSs). OBJECTIVES: This study aims to evaluate and compare dosimetric accuracy of the four dose calculation algorithms applying to homogeneous and heterogeneous media, VMAT plans (based on AAPM TG-119 test cases), and the surface and buildup regions. METHODS: The four algorithms are assessed in homogeneous (IAEA-TECDOCE 1540) and heterogeneous (IAEATECDOC 1583) media. Dosimetric evaluation accuracy for VMAT plans is then analyzed, along with the evaluation of the accuracy of algorithms applying to the surface and buildup regions. RESULTS: Tests conducted in homogeneous media revealed that all algorithms exhibit dose deviations within 5% for various conditions, with pass rates exceeding 95% based on recommended tolerances. Additionally, the tests conducted in heterogeneous media demonstrate high pass rates for all algorithms, with a 100% pass rate observed for 6MV and mostly 100% pass rate for 15MV, except for CCC, which achieves a pass rate of 94%. The results of gamma index pass rate (GIPR) for dose calculation algorithms in IMRT fields show that GIPR (3%/3 mm) for all four algorithms in all evaluated tests based on TG119, are greater than 97%. The results of the algorithm testing for the accuracy of superficial dose reveal variations in dose differences, ranging from -11.9% to 7.03% for 15MV and -9.5% to 3.3% for 6MV, respectively. It is noteworthy that the AXB and MC algorithms demonstrate relatively lower discrepancies compared to the other algorithms. CONCLUSIONS: This study shows that generally, two dose calculation algorithms (AXB and MC) that calculate dose in medium have better accuracy than other two dose calculation algorithms (CCC and AAA) that calculate dose to water.
引用
收藏
页码:1013 / 1033
页数:21
相关论文
共 50 条
  • [31] The evaluation of several commercial IMRT treatment planning systems heterogeneity dose calculation algorithms using an anthropomorphic thorax phantom
    Davidson, S
    Followill, D
    Ibbott, G
    Prado, K
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1988 - 1988
  • [32] Dosimetric verification of four dose calculation algorithms for spine stereotactic body radiotherapy
    Hirashima, Hideaki
    Nakamura, Mitsuhiro
    Nakamura, Kiyonao
    Matsuo, Yukinori
    Mizowaki, Takashi
    [J]. JOURNAL OF RADIATION RESEARCH, 2024, 65 (01) : 109 - 118
  • [33] Comparative evaluation of four commercial tests for presumptive identification of Candida albicans
    Cárdenes, CD
    Carrillo-Muñoz, AJ
    Arias, A
    Rodríguez-Alvarez, C
    Torres-Lana, A
    Sierra, A
    Arévaloa, MP
    [J]. JOURNAL OF MICROBIOLOGICAL METHODS, 2004, 59 (02) : 293 - 297
  • [34] Evaluation of heterogeneity dose distributions for Stereotactic Radiotherapy (SRT): comparison of commercially available Monte Carlo dose calculation with other algorithms
    Takahashi, Wataru
    Yamashita, Hideomi
    Saotome, Naoya
    Iwai, Yoshio
    Sakumi, Akira
    Haga, Akihiro
    Nakagawa, Keiichi
    [J]. RADIATION ONCOLOGY, 2012, 7
  • [35] Evaluation of heterogeneity dose distributions for Stereotactic Radiotherapy (SRT): comparison of commercially available Monte Carlo dose calculation with other algorithms
    Wataru Takahashi
    Hideomi Yamashita
    Naoya Saotome
    Yoshio Iwai
    Akira Sakumi
    Akihiro Haga
    Keiichi Nakagawa
    [J]. Radiation Oncology, 7
  • [36] Dosimetric Comparison and Plan Evaluation of Different Dose Computing Algorithms for Different Radiotherapy Techniques in Head and Neck Tumors
    Pandu B.
    Khanna D.
    Mohandass P.
    Elavarasan R.
    Vivek T.R.
    Jacob S.
    Sunny G.
    Ninan H.
    Maddipati T.
    [J]. Iranian Journal of Medical Physics, 2022, 19 (06) : 346 - 355
  • [37] Dosimetric Comparison of Different Dose Calculation Algorithms in Postmastectomy Breast Cancer Patients Using Conformal Planning Techniques
    Gaur, Garima
    Dangwal, Vinod Kumar
    Banipal, Raja Paramjeet Singh
    Singh, Ranjit
    Kaur, Gurpreet
    Grover, Romikant
    Sachdeva, Sheetal
    Kang, Manraj Singh
    Singh, Simrandeep
    Garg, Pardeep
    Singh, Baltej
    [J]. JOURNAL OF MEDICAL PHYSICS, 2023, 48 (02) : 136 - 145
  • [38] A Comparison of Surface Dose Values Using Radiochromic Film Measurements and Treatment Plans Calculated with Different Calculation Algorithms
    Kaptan, I.
    Akdeniz, Y.
    Ispir, B.
    Ugurluer, G.
    Serin, M.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 3074 - 3074
  • [39] Dosimetric comparison between VMAT with different dose calculation algorithms and protons for soft-tissue sarcoma radiotherapy
    Fogliata, Antonella
    Scorsetti, Marta
    Navarria, Piera
    Catalano, Maddalena
    Clivio, Alessandro
    Cozzi, Luca
    Lobefalo, Francesca
    Nicolini, Giorgia
    Palumbo, Valentina
    Pellegrini, Chiara
    Reggiori, Giacomo
    Roggio, Antonella
    Vanetti, Eugenio
    Alongi, Filippo
    Pentimalli, Sara
    Mancosu, Pietro
    [J]. ACTA ONCOLOGICA, 2013, 52 (03) : 545 - 552
  • [40] Influence of different dose calculation algorithms on the estimate of NTCP for lung complications
    Hedin, Emma
    Back, Anna
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (05): : 127 - 139