Dosimetric verification of four dose calculation algorithms for spine stereotactic body radiotherapy

被引:1
|
作者
Hirashima, Hideaki [1 ]
Nakamura, Mitsuhiro [1 ,2 ]
Nakamura, Kiyonao [1 ]
Matsuo, Yukinori [1 ]
Mizowaki, Takashi [1 ]
机构
[1] Kyoto Univ, Grad Sch Med, Dept Radiat Oncol & Image Appl Therapy, 54 Kawahara Cho,Sakyo Ku, Kyoto 6068507, Japan
[2] Kyoto Univ, Grad Sch Med, Dept Adv Med Phys, 53 Kawahara Cho,Sakyo Ku, Kyoto 6068507, Japan
关键词
dosimetric verification; dose calculation algorithm; spine SBRT; HIGH-Z INHOMOGENEITY; RADIATION-THERAPY; MONTE-CARLO; HIP-PROSTHESIS; ACUROS XB; METASTASES; PROBABILITY; MULTICENTER; CONSENSUS;
D O I
10.1093/jrr/rrad086
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
The applications of Type B [anisotropic analytical algorithm (AAA) and collapsed cone (CC)] and Type C [Acuros XB (AXB) and photon Monte Carlo (PMC)] dose calculation algorithms in spine stereotactic body radiotherapy (SBRT) were evaluated. Water- and bone-equivalent phantoms were combined to evaluate the percentage depth dose and dose profile. Subsequently, 48 consecutive patients with clinical spine SBRT plans were evaluated. All treatment plans were created using AXB in Eclipse. The prescription dose was 24 Gy in two fractions at a 10 MV FFF on TrueBeam. The doses were then recalculated with AAA, CC and PMC while maintaining the AXB-calculated monitor units and beam arrangement. The dose index values obtained using the four dose calculation algorithms were then compared. The AXB and PMC dose distributions agreed with the bone-equivalent phantom measurements (within +/- 2.0%); the AAA and CC values were higher than those in the bone-equivalent phantom region. For the spine SBRT plans, PMC, AAA and CC were overestimated compared with AXB in terms of the near minimum and maximum doses of the target and organ at risk, respectively; the mean dose difference was within 4.2%, which is equivalent with within 1 Gy. The phantom study showed that the results from AXB and PMC agreed with the measurements within +/- 2.0%. However, the mean dose difference ranged from 0.5 to 1 Gy in the spine SBRT planning study when the dose calculation algorithms changed. Users should incorporate a clinical introduction that includes an awareness of these differences.
引用
收藏
页码:109 / 118
页数:10
相关论文
共 50 条
  • [1] Dosimetric Impact of Four Dose Calculation Algorithms On Stereotactic Body Radiation Therapy for Lung Cancer
    Kusunoki, T.
    Kurooka, M.
    Nonaka, T.
    Yoshino, S.
    Shioiri, K.
    Ide, S.
    Nakayama, Y.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2967 - 2967
  • [2] Dosimetric Comparison and Clinical Implications of Dose Calculation Algorithms for Stereotactic Body Radiation Therapy in Spine Metastases
    Rao, Arthi
    Reilly, Michael
    Shen, Zhilei
    Ye, Jason
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (02): : E62 - E62
  • [3] Dosimetric Comparison and Clinical Implications of Dose Calculation Algorithms for Stereotactic Body Radiation Therapy in Spine Metastases
    Rao, A. S.
    Reilly, M. P.
    Chang, E. L.
    Ye, J. C.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : E609 - E609
  • [4] Comparison of four techniques for spine stereotactic body radiotherapy: Dosimetric and efficiency analysis
    Aljabab, Saif
    Vellayappan, Balamurugan
    Vandervoort, Eric
    Bahm, Jamie
    Zohr, Robert
    Sinclair, John
    Caudrelier, Jean-Michel
    Szanto, Janos
    Malone, Shawn
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (02): : 160 - 167
  • [5] Analysis of Dosimetric Verification Results of Stereotactic Body Radiotherapy
    Zhang, X.
    Yang, R.
    Li, J.
    Li, J.
    Wang, J.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E610 - E610
  • [6] DOSIMETRIC COMPARISON OF THREE TECHNIQUES FOR SPINE STEREOTACTIC BODY RADIOTHERAPY
    Aljabab, Saif
    Vellayappan, Balamurugan
    Vandervoort, Eric
    Bahm, Jamie
    Zohr, Robert
    Sinclair, John
    Szanto, Janos
    Malone, Shawn
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 120 : S93 - S93
  • [7] Dosimetric Impact of Esophagus Motion in Spine Stereotactic Body Radiotherapy
    Yang, J.
    Wang, X.
    Zhao, Z.
    Yang, J.
    Zhang, Y.
    Court, L.
    Li, J.
    Brown, P.
    Ghia, A.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3436 - 3436
  • [8] Dosimetric Effect Of Rotation On Stereotactic Body Radiotherapy (SBRT) Of The Spine
    Djemil, T.
    Angelov, L.
    Chao, S.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3802 - 3802
  • [9] Accuracy of stereotactic body radiotherapy dose calculation algorithms outside treatment field
    Viren, T.
    Salomaa, S.
    Seppala, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S476 - S477
  • [10] Superficial dose verification of four dose calculation algorithms
    Cao, Y.
    Yang, Z.
    Yang, X.
    Qiu, X.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S166 - S166