Background: Compared to direct laryngoscopy, videolaryngoscopy (VL) can provide improved laryngeal and glottic view, higher intubation success rates in patients with a known or predicted difficult airway and reduced incidence of laryngeal/airway trauma. However, the cost and availability of these devices handicap its use in resource-restricted facilities. The objective was to design and manufacture a novel VL using additive manufacturing (AM) and evaluate its usability on an intubation manikin by comparing it to one of the most common video laryngoscopes used in clinical practice, the CMAC (R), by measuring the time to first pass of the endotracheal tube as the main outcome. Methods: A randomised cross-over study was performed with 36 anaesthetists attempting tracheal intubation of a manikin. The novel 3D-printed hyperangulated VL blade was compared to a CMAC (R) VL (D-blade). Participants had no prior experience or training with the novel device. The participants included consultants, registrars/trainees and medical officers in the Department of Anaesthesiology at the University of the Free State (UFS) in South Africa. Results: The CMAC (R) had a statistically shorter time to first pass (median 13.8 seconds) compared to the 3D-printed model (median 19.0 seconds) (95% confidence interval [CI] 1.0-6.2; P=0.0013). No failed attempts occurred with either device. Conclusion: Intubation times were faster with the CMAC (R) than with the novel device. However, with a comparable intubation success rate, 3D printing technology potentially can improve access to video laryngoscopy. Further design improvements, validation of materials and manufacturing processes are required before 3D-printed laryngoscope blades can be used in human subjects.
引用
收藏
页码:157 / 165
页数:9
相关论文
共 50 条
  • [21] Tracheal intubation in children with Morquio syndrome using the angulated video-intubation laryngoscope
    Dullenkopf, A
    Holzmann, D
    Feurer, R
    Gerber, A
    Weiss, M
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2002, 49 (02): : 198 - 202
  • [22] Tracheal intubation in daylight and in the dark: a randomised comparison of the Airway Scope®, Airtraq®, and Macintosh laryngoscope in a manikin
    Ueshima, H.
    Asai, T.
    ANAESTHESIA, 2010, 65 (07) : 684 - 687
  • [23] Video laryngoscope guided tracheal intubation. Comparison of tracheal tube introducer and stylet
    Duff, E.
    Junaidu, A.
    Hodzovic, I.
    ANAESTHESIA, 2008, 63 (08) : 898 - 899
  • [24] Efficacy of the Airtraq® laryngoscope with a fiberoptic bronchoscope compared with that of Airtraq® alone for tracheal intubation: a manikin study
    Kohki Nishikawa
    Eiko Hukuoka
    Toshiya Kawagishi
    Yuki Shimodate
    Michiaki Yamakage
    Journal of Anesthesia, 2011, 25 : 93 - 97
  • [25] Comparison of three types of laryngoscope for tracheal intubation during rhythmic chest compressions: A manikin study
    Koyama, J.
    Iwashita, T.
    Okamoto, K.
    RESUSCITATION, 2010, 81 (09) : 1172 - 1174
  • [26] Evaluation of new laryngoscope blade for tracheal intubation, Truview EVO2©:: a manikin study
    Miceli, L.
    Cecconi, M.
    Tripi, G.
    Zauli, M.
    Della Rocca, G.
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2008, 25 (06) : 446 - 449
  • [27] The usefulness of the GlideScope video laryngoscope in the education of conventional tracheal intubation for the novice
    You, J. S.
    Park, S.
    Chung, S. P.
    Park, Y. S.
    Park, J. W.
    EMERGENCY MEDICINE JOURNAL, 2009, 26 (02) : 109 - 111
  • [28] Efficacy of the Airtraq® laryngoscope with a fiberoptic bronchoscope compared with that of Airtraq® alone for tracheal intubation: a manikin study
    Nishikawa, Kohki
    Hukuoka, Eiko
    Kawagishi, Toshiya
    Shimodate, Yuki
    Yamakage, Michiaki
    JOURNAL OF ANESTHESIA, 2011, 25 (01) : 93 - 97
  • [29] A comparative study of Video laryngoscope vs Macintosh laryngoscope for prehospital tracheal intubation in Hiroshima, Japan
    Santou, N.
    Ueta, H.
    Nakagawa, K.
    Hata, K.
    Kusunoki, S.
    Sadamori, T.
    Takyu, H.
    Tanaka, H.
    RESUSCITATION PLUS, 2023, 13
  • [30] Comparison of Airtraq™ video-laryngoscope and Macintosh laryngoscope for tracheal intubation in adults—a randomised study
    Seema Meena
    Sujata Chaudhary
    Rashmi Salhotra
    Swati Bharti
    Bismanjeet Kaur Khurana
    Ain-Shams Journal of Anesthesiology, 15